Altek 334A Loop Calibrator - DISCONTINUED

SKU
334A
Manufacturer:
Altek
Altek

Key Features:

This Product is DISCONTINUED
Please see the PIECAL 334 as a Recommended Replacement

  • All 4 to 20 mA Loop Functions— Source 0.00 to 24.00 mA or -25.0 to 125.0% read mA or percent to 52.00 mA, simulate 2-wire transmitters, measure -99.99 to 99.99 volts DC power & read 2-wire transmitters.
  • Quik-Chek Switch— Select 4 or 20 mA output or 2-wire simulation. Accurate to ±0.006 mA.
  • Full Range 10 Microamp Resolution— Adjust 4 digit display to ±0.05% without range change.
  • Long Battery Life— 3x9 volt alkaline batteries source 20 mA for over 20 hours, 12 mA for over 33 hours; 2-wire simulate and read for 150 hours. Much longer than NiCd batteries or calibrators with a single 9 V battery!
  • Drive Capability— 1200 Ohms with current limited, fuseless protection to 135 V AC/DC.
Call For Price

Secured and trusted checkout with

Altek 334A checks all loop signals, mA, % of Span and DC V. Saves time on every mA transmitter, recorder, alarm trip, I/P and computer/PLC you calibrate. Flip the QUIK-CHEK switch for instant outputs of 4 and 20 mA accurate to 0.006 mA, and trim your device in seconds.

Specification - Altek 334A

ACCURACY:
   Quik-Chek 4 & 20 mA ±(0.03% of Setting + 1 LSD)
   Milliamp & % Ranges ±(0.05% of Reading + 1 LSD)
   Read DC Volts ±(0.05% of Span + 1 LSD)
WARM UP TIME: 10 seconds to specified accuracy, 2 minutes to maximum accuracy
TEMPERATURE EFFECT:
   Milliamp Ranges: ±0.008% of Reading/°C
   Read DC Volts: ±0.012% of Span/°C
BATTERIES: Three “9V”, (1604) batteries (Alkaline supplied and recommended)
BATTERY LIFE:
    SOURCE & POWER TRANSMITTER: Nominal 33 hours at 12 mA, 20 hours at 20 mA with 250 Ohm load
   READ mA, DC V & 2-WIRE SIMULATE: Nominal 100 hours
LOW BATTERY INDICATION: “BAT” indication will first appear on the display when approximately 4 hours remain to
source 12 mA into 250 Ohms. This is typical use for a full shift.
OVERVOLTAGE PROTECTION: Protected to 135V AC/DC for 30 seconds without fuses
OPERATING TEMPERATURE RANGE: -5 to +140 °F (-20 to +60°C)
STORAGE TEMPERATURE RANGE: -22 to +160°F (-30 to +70°C)
RELATIVE HUMIDITY: 10 to 90%, non-condensing for 24 hours from 0 to 35°C
OVERALL SIZE: 121.9 x 66.0 x 35.8 mm (4.8 x 2.6 x 1.41 inches)
WEIGHT: 0.22 kg (7.8 oz)
MILLIAMP SOURCE
   RANGES:
   0.00 to 24.00 mA; -25.0 to 125.0 % of 4 to 20 mA
   ACCURACY:
   Quik-Chek 4mA ±(0.03% of 4mA + 0.01 mA) = 0.011mA
   Quik-Chek 20mA ±(0.03% of 20mA + 0.01 mA) = 0.016mA
   0 to 24mA ±(0.05% of Reading + 1 LSD)
   TYPICAL DRIVE CAPABILITY: 1200 Ohms @ 20.00 mA with fresh batteries, 800 Ohms @ 20.00 mA with low batteries
   COMPLIANCE: nominal 27VDC @ 20 mA with fresh batteries
   POWER & MEASURE 2-WIRE TRANSMITTERS
   RANGES & ACCURACY: Same as for MILLIAMP SOURCE
   OUTPUT CURRENT: up to 24.00 mA
   TYPICAL DRIVE CAPABILITY:1200 Ohms @ 20.00 mA with fresh batteries, 800 Ohms @ 20.00 mA with low batteries
   COMPLIANCE: nominal 27VDC @ 20 mA with fresh batteries
   2-WIRE TRANSMITTER SIMULATOR
   RANGES & ACCURACY: Same as for MILLIAMP SOURCE
   LOOP VOLTAGE LIMITS: Minimum, 2 VDC; Maximum 100 VDC
   OVERLOAD PROTECTION: Current limited to 25 mA nominal
MILLIAMP READ
   RANGES: 0.00 to 52.00 mA; -25.0 to 300.0 % of 4 to 20 mA
   ACCURACY: Same as for MILLIAMP SOURCE
   VOLTAGE BURDEN: 0.9V at 4 mA, 1.2V at 20 mA, 1.9V at 50 mA excitation current
READ DC VOLTS
   RANGE:-99.99 to +99.99V DC
   INPUT IMPEDANCE: >2 Meg Ohms

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account