Altek 211 RTD Calibrator - DISCONTINUED

SKU
Altek-211
Manufacturer:
Altek
Altek

Key Features:

This Product is DISCONTINUED
Please see the PIECAL 211 as a Recommended Replacement

  • 8 RTD TYPES & OHMS

Select Pt, Cu, or Ni

  • TEMPERATURE INPUT & OUTPUT

Directly in degrees

  • “QUIK CHEK® ” SWITCH

Three Points, HI, LO, & Set

  • 0.025% ACCURACY,1° OR 0.1° RESOLUTION

Field Selectable °F or °C

  • THREE YEAR WARRANTY

Toolbox tough

  • CUSTOM RANGES AVAILABLE
Call For Price

Secured and trusted checkout with

      Altek’s Model 211 RTD Calibrator lets you simulate and read RTDs over the entire industrial temperature range. Use with transmitters, single & dual channel recorders, controllers, alarms and indicators. Switch between four Pt 100, one Pt 1000, one Cu 10, two Ni RTD curves or ohms ranges from 0.00 to 410.00 and 0.0 to 2010.0. Field customize the Model 211 to lock-in 1° resolution, fixed ° C or ° F or front selectable ° F/° C operation. Read and simulate 2-wire and automatically compensate for 3-wire RTDs with built-in leads. A plug-in lead is supplied for 4-wire RTD measurements. RTD types may be selected each time the unit is turned on. If you always use one RTD curve, lock-in the selected RTD type with the internal DIP switch to prevent accidental change to an unwanted RTD type.

Altek 211 Technical Specifications:


Accuracy: ±0.025%; ±0.013% typical
Temperature drift: ±0.01 %/°C
Operating temperature range: -5 to +140° F (-20 to +60° C)
Storage temperature range: -22 to +175° F (-30 to +80° C)
Relative humidity: 10 to 90%, non-condensing
Warm up time: 30 seconds to maximum accuracy
Overload protection: Electronically protected to 40 VDC between leads 1 & 2, fuse protected to 125 VAC, 5A between leads 2 & 3
Battery life: 9 Volt Alkaline: 50 hours
Low battery: “BAT” indication on LCD at 7 Volts nominal, approximately 10 hours left
Overall size: 2 1/2 x 2 5/8 x 5 1/8 inches (63.5 x 66.7 x 130 mm)
Leads: Approximately 18" (0.46m) terminated with spade lugs
Weight: 11.5 oz. (0.33 kg)
Carrying case: Included, zippered with belt loop

Simulate RTD or resistance
Output resistance range: 0.00 to 2010.0 Ohms
Ohms: 0.00 Ω to 410.00 Ω (±0.10 Ω); 0.0 Ω to 2010.0 Ω (±0.50 Ω)
Allowable external excitation current: 0.090 to 5.800mA, DC
Intermittent excitation current mode: Accepts intermittent excitation currents
with minimum 125 msec fixed current at minimum repetition rate of 1/sec from 0.090 to 5.800mA
Measuring RTD sensors
Excitation current supplied: 1 mA, nominal
Normal mode rejection: 50/60 Hz, 50 db
Common mode rejection: 50/60 Hz, 100 db

RTD type

Alpha

Range °C

 Accuracy °C

Range °F

Accuracy °F

Pt 100 (DIN/IEC/JIS 1989)

1.3850

-200.0 to 850.0 °C

±0.15 °C

-328.0 to 1562.0 °F

±0.25 °F

Pt 100 (Old JIS 1981)

1.3916

-200.0 to 648.9 °C

±0.15 °C

-328.0 to 1200 °F

±0.25 °F

Pt 100 (Burns)

1.3902

-195.6 to 648.9 °C

±0.15 °C

-320.0 to 1200 °F

±0.25 °F

Pt 100 (US Lab)

1.3926

-200.0 to 862.0 °C

±0.15 °C

-328.0 to 1584 °F

±0.25 °F

Pt 1000 (HVAC)

1.3750

-184 to 275  °C

±0.10 °C

-300 to 527  °F

±0.15 °F

Cu 10 (Minco)

1.4274

-200 to 260 °C

±1.10 °C

-328 to 500  °F

±2.00 °F

Ni 120 (Pure)

1.6720

-80.0 to 273.0 °C

273 to 316  °C

±0.10 °C

-112.0 to 524.0 °F

524 to 600 °F

±0.15 °F

Ni 110 (Bristol 7 NA)

1.5801

-106.7 to 315.6 °C

±0.10 °C

-160.0 to 600 °F

±0.20 °F

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account