Rotronic HygroClip HC2-P05 Mini Insertion Probe

SKU
HC2-P05
Manufacturer:
Rotronic
Rotronic

Key Features:

  • Humidity measuring range: 0..100 %RH
  • Temperature limits: -40…85 °C (-40…185 °F)
  • Probe tip: diameter: 5 mm (0.2”),length: 200 mm (7.9”)
  • Probe cable length: 2 m (6.5 ft)
  • Handgrip: POM, black -, barrel: PPS, black
$1,491.00

Secured and trusted checkout with

The Rotronic HC2-P05 has an accuracy of ±1.5 %RH, ±0.3 K, at 23 °C ±5 K and the application range is from -40 to 85 °C / 0 to 100 %RH.
Dimensions∅ 5 x 200 mm
Weight160 g
For use withHygroLab C1, HP23-AW-A
Operating range electronics-40…85 °C / 0…100 %rh
Range of application-40…85 °C (-40…185 °F) / 0…100 %RH
Accuracy at 23 ±5 °C± 0.015 aw / 1.5 %rh / ±0.3 K
Power supply3.2…5 VDC, calibrated at 3.3 VDC, current: ~4.5 mA
Long-term stability sensor<1 %rh / year
Humidity sensorHYGROMER® WA-1 / Pt-100, 1/3 DIN Class B
Temperature dependence0.001 K/K, 0.0002 aw/K
Filter typeNo filter available
Response time t 63<15 sec.
Psychrometric calculationsDew or frost point
Start-up time1,5 s
Measurement interval1 s
MaterialStainless steel DIN 1.4305 (probe), POM (handle)
IP protection ratingIP30
FDA / GAMPFDA 21 CFR part 11 / GAMP5
CE / EMC conform2007 / 108 / EG



Dimensions

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account