The new 725Ex Intrinsically Safe Multi-function Process Calibrator is powerful yet easy-to-use. Combined with one of the (8) new Fluke-700PEx pressure modules, the 725Ex is able to calibrate almost any process instrument likely to need service in an area where explosive gasses may be present.
Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.
Do you need to calibrate transmitters? The 725Ex split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725Ex has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.
From the moment you pick it up, the Fluke 725Ex is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.
Specifications
Measurement Accuracy
Voltage DC
30.000 V
0.02% + 2 counts (upper display)
30.000 V
0.02% + 2 counts (lower display)
100.00 mV
0.02% + 2 counts
-10.00 mV to 75.00 mV
0.025 % + 1 count (via TC connector)
Current DC
24.000 mA
0.02 %+ 2 counts
Resistance
0.0 to 400.0 Ohm
0.1Ohm (4-wire), 0.15Ohm (2- and 3-wire)
401 to 1500 Ohm
0.5Ohm (4-wire), 1Ohm (2- and 3-wire)
1500 to 3200 Ohm
1Ohm (4-wire), 1.5Ohm (2- and 3-wire)
Frequency
2.0 to 1000.0 CPM
0.05 % + 1 count
1.0 to 1100.0 Hz
0.05 % + 1 count
1.00 to 10.00 kHz
0.05 % + 1 count
Sensitivity
1 V peak-to-peak-minimum
Pressure
Accuracy
from 0.025% of range using any of 8 intrinsic safe pressure modules.
(for detailed specifications refer to pressure modules in options and accessories)
Modules available for differential, gage, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC
100.00 mV
0.02% +2 counts
10.000 V
0.02% +2 counts
-10.00 mV to 75.00 mV
0.025 % + 1 count (via TC connector)
Current DC
24.000 mA (Source)
0.02% + 2 counts
24.000 mA (Simulate)
0.02% + 2 counts
Resistance
15.0 to 400.0 Ohm
0.15 (exc. current 0.15 to 0.5 mA), 0.1 Ohm (exc. current 0.5 to 2 mA)
401 to 1500 Ohm
0.5 Ohm(excitation current 0.05 to 0.8 mA)
1500 to 3200 Ohm
1 Ohm(excitation current 0.05 to 0.4 mA)
Frequency
2.0 to 1000.0 CPM
0.05%
1.0 to 1100.0 Hz
0.05%
1.00 to 10.00 kHz
0.25%
Waveform:
5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure Accuracy:
NI-120
0.2 Deg C
PT-100 (385)
0.33 Deg C
PT-100 (393)
0.3 Deg C
PT-100 (JIS)
0.3 Deg C
PT-200 (385)
0.2 Deg C
PT-500 (385)
0.3 Deg C
PT-1000 (385)
0.2 Deg C
Resolution:
0.1 Deg C
J:
0.7 Deg C
K:
0.8 Deg C
T:
0.8 Deg C
E:
0.7 Deg C
R:
1.8 Deg C
S:
1.5 Deg C
B:
1.4 Deg C
L:
0.7 Deg C
U:
0.75 Deg C
N:
0.9 Deg C
Resolution
J, K, T, E, L, N, U: 0.1 Deg C, 0.1 Deg F B, R, S: 1 Deg C, 1 Deg F
Source Accuracy:
NI-120
0.2 Deg C
PT-100 (385)
0.33 Deg C
PT-100 (393)
0.3 Deg C
PT-100 (JIS)
0.3 Deg C
PT-200 (385)
0.2 Deg C
PT-500 (385)
0.3 Deg C
PT-1000 (385)
0.2 Deg C
Resolution:
0.1 Deg C
Note:
Accuracy stated for 4-wire measurement.
J:
0.7 Deg C
K:
0.8 Deg C
T:
0.8 Deg C
E:
0.7 Deg C
R:
1.4 Deg C
S:
1.5 Deg C
B:
1.4 Deg C
L:
0.7 Deg C
U:
0.75 Deg C
N:
0.9 Deg C
Resolution
J, K, T, E, L, N, U: 0.1 Deg C, B, R, S: 1 Deg C
Specifications
Ramp Functions
Source functions:
Voltage, current, resistance, frequency, temperature
Ramps
Slow ramp, Fast ramp, 25% step-ramp
Loop Power Function
Voltage:
12 V
Accuracy:
10%
Maximum current:
22 mA, short circuit protected
Step Functions
Source functions:
Voltage, current, resistance, frequency, temperature
Steps
25% of range, 100% of range
Safety Specifications
Agency Approvals
ATEX II 1 G EEx ia IIB 171Deg C
I.S. Class I, Division 1 Groups
Mechanical & General Specifications
Size
130 x 236 x 61 mm
5.188" x 9.291" x 2.402"
Weight
0.85 kg (1.874 lbs.)
Batteries
4 AA alkaline batteries
Battery Replacement
Separate battery compartment, accessible without breaking calibration seal