Technical Specifications |
---|
Display | 5.0 inch 480 x 800 TFT LCD capacitive screen | Size | 16.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm) | Weight | 1.5 lbs. (0.65 Kg) | Power Supply | 6600mAh, 23.8Wh lithium battery, charging time 4~6 hours, battery pack can be charged independent. Battery life typically 16 hours |
Environment | Specification guaranteed temperature range: (10~30) °C | Working Temperature: (-10~50) °C | Storage temperature: (-20~70) °C | Humidity: 0% ~ 95% RH, non-condensing | Warm-Up Time | 10 minutes | Ports Protection Voltage | 50V max | CE Certificate | TUV IEC61326,IEC61010 | Rohs Compliance | Rohs II Directive 2011/65/EU,EN50581:2012 | IP Protection Level | IP67, 1 meter drop test | Communication | Isolate USB-TYPEC (slave), Bluetooth BLE | Input Channels | CH1, CH2 analog channel, 6 pins smart lemo ports for RTD probe;MINI-TC ports for TC probe | Measurement Display | Single channel, dual channel, differential (e.g T1-T2) | Measuring Rates | CH1, CH2 analog channels alternately and cyclically measure | RTD measuring rate: 1.6S/single channel, 1.6S/dual channel | TC measuring rate: 0.8S/single channel, 0.8S/dual channel | Measurement Units | °C, °F, K | Statistics | Max, Min, Avg |
Measurement Specifications |
---|
PRT Measurement | RTD Types | ITS-90, CVD, Ohms, Pt100 (385), Pt10 (385), Pt25 (385), Pt50 (385), Pt100 (3916), Pt100 (3926), Pt100 (391), Cu100 (428), Cu50 (428), Cu10 (427), Ni100 (617), Ni100 (618), Ni120 (672), and custom RTD | Resistance Accuracy | 0~400Ω: ±0.5mΩ@(0~20Ω),±25ppm@(20~400Ω) | Measurement range | -200°C ~ 850°C | Resolution | ±0.1mΩ or 0.001°C | Connection Type | 4-wire smart connection | Excitation Current | 1 mA - alternating constant current | Temperature Coefficient | ±2ppm FS/°C (-10°C~10°C and 30°C~50°C) |
TC Measurement | TC Types | mV, S, R, B, K, N, E, J, T, C, D, G, L, U, LR, A, 10uV/°C, 1mV/°C, Standard TC | Electrical Measurement | -10~75mV: 50ppm RDG+2uV | TC measurement range | -270°C ~ 1800°C | Resolution | ±0.1uV or 0.001°C | Connection Type | Mini-TC | CJC compensation methods | Internal, external or manual entry | Temperature Coefficient | ±5ppm FS/°C (-10°C~10°C and 30°C~50°C) | Internal CJC Specification | ±0.15°C (-10°C~50°C) |
Accuracy (°C) |
---|
T, °C | Rx,Ω | ADT282 readout only (°C) | Readout with selected Probe Accuracy (oC)* |
---|
AM1760 | AM1751 | AM1730 | AM1640 | AM1660 | AM1710 | AM1612-2 | AM1612-1 |
---|
-200 | 18 | 0.005 | 0.013 | 0.021 | 0.021 | 0.053 | 0.053 | n/a | 0.072 | n/a | -40 | 84 | 0.005 | 0.013 | 0.018 | 0.018 | 0.042 | 0.042 | 0.018 | 0.051 | 0.051 | 0 | 100 | 0.006 | 0.009 | 0.014 | 0.014 | 0.036 | 0.036 | 0.014 | 0.051 | 0.051 | 100 | 140 | 0.009 | [1] | [1] | [1] | [1] | [1] | 0.019 | 0.051 | 0.051 | 160 | 163 | 0.011 | [1] | [1] | [1] | [1] | [1] | 0.023 | 0.052 | 0.052 | 232 | 190 | 0.013 | 0.019 | 0.024 | 0.024 | 0.059 | 0.059 | n/a | n/a | n/a | 420 | 257 | 0.018 | 0.027 | 0.033 | 0.033 | 0.077 | 0.077 | n/a | n/a | n/a | 660 | 338 | 0.026 | 0.04 | 0.046 | n/a | n/a | 0.109 | n/a | n/a | n/a | Note: [1] – These are non-standard calibration points, some probes are not calibrated at 100°C and/or 160°C. | * Includes readout accuracy, probe calibration, and probe drift. (K=2) |
Thermocouple Measurement (Environment Temperature: 20±10° C) |
---|
Type | Temperature Range (°C) | Accuracy (°C) External CJC compensation (1 year) | Accuracy (°C) Internal CJC compensation (1 year) |
---|
S | -50 to 1768 | -50~0 | 0.51 | 0.53 | 0~100 | 0.37 | 0.4 | 100~1768 | 0.28 | 0.32 | R | -50 to 1768 | -50~0 | 0.54 | 0.56 | 0~200 | 0.38 | 0.41 | 200~1768 | 0.25 | 0.29 |
B |
0 to 1820 | 200~300 | 1.01 | 1.02 | 300~500 | 0.66 | 0.68 | 500~800 | 0.41 | 0.44 | 800~1820 | 0.28 | 0.32 |
K |
-270 to 1372 | -250 to -200 | 0.48 | 0.5 | -200 to -100 | 0.15 | 0.21 | -100 to 600 | 0.08 | 0.17 | 600 to 1372 | 0.14 | 0.21 |
N |
-270 to 1300 | -250 to -200 | 0.76 | 0.77 | -200 to -100 | 0.22 | 0.27 | -100 to 1300 | 0.12 | 0.19 |
E |
-270 to 1000 | -250~-200 | 0.26 | 0.3 | -200~-100 | 0.1 | 0.18 | -100~700 | 0.06 | 0.16 | 700~1000 | 0.08 | 0.17 | J | -210~1200 | -210~-100 | 0.13 | 0.2 | -100~700 | 0.06 | 0.16 | 700~1200 | 0.1 | 0.18 | T | -270 to 400 | -250~-100 | 0.36 | 0.39 | -100~0 | 0.08 | 0.17 | 0~400 | 0.05 | 0.16 | C | 0 to 2315 | 0 to 1000 | 0.16 | 0.22 | 1000 to 1800 | 0.26 | 0.3 | 1800 to 2315 | 0.42 | 0.45 | D | 0~2315 | 0~100 | 0.21 | 0.26 | 100~1200 | 0.16 | 0.22 | 1200~2000 | 0.27 | 0.31 | 2000~2315 | 0.42 | 0.45 |
G |
0 to 2315 | 50~100 | 0.6 | 0.62 | 100~200 | 0.38 | 0.41 | 200~400 | 0.24 | 0.28 | 400~1500 | 0.16 | 0.22 | 1500~2315 | 0.32 | 0.35 |
L |
-200 to 900 | -200 to -100 | 0.07 | 0.17 | -100 to 400 | 0.06 | 0.16 | 400 to 900 | 0.07 | 0.17 | U | -200 to 600 | -200 to 0 | 0.14 | 0.21 | 0 to 600 | 0.05 | 0.16 | LR | -200~800 | -200~0 | 0.09 | 0.17 | 0~800 | 0.06 | 0.16 | A | 0~2500 | 0~1200 | 0.2 | 0.25 | 1200~2000 | 0.33 | 0.36 | 2000~2500 | 0.48 | 0.5 |
Secondary PRT Information |
---|
Specification | AM1710 Series | AM1730 Series | AM1751 Series | AM1760 Series |
---|
Temperature Range [3] | -60°C to 160°C | -200°C to 420°C | -200°C to 670°C | -200°C to 670°C | Resistance at 0°C | Nominal 100Ω | Temperature Coefficient | 0.003925 Ω / Ω / °C | Calibrated Accuracy (k=2)[2][3] | ±0.025°C at -40°C | ±0.025°C at -40°C | ±0.025°C at -40°C | ±0.010°C at -196°C | ±0.015°C at 0.01°C | ±0.015°C at 0.01°C | ±0.015°C at 0.01°C | ±0.006°C at 0.01°C | ±0.025°C at 160°C | ±0.035°C at 420°C | ±0.035°C at 420°C | ±0.015°C at 420°C | | | ±0.05°C at 661°C | ±0.025°C at 661°C | Drift | ±0.01°C at TPW after 100 hours at 160°C | ±0.01°C at TPW after 100 hours at 420°C | ±0.01°C at TPW after 100 hours at 661°C | ±0.004°C at TPW after 100 hours at 661°C | Short Term Stability | ±0.007°C | ±0.002°C | Thermal Shock | ±0.005°C after (10) thermal cycles from minimum to maximum temperatures | ±0.002°C after (10) thermal cycles from minimum to maximum temperatures | Hysteresis | <=0.005°C | <=0.001°C | Self-heating | 50 mW/°C | 0.0015°C at 0.5mA | Response Time | 9 seconds for 63% response to step change in water moving at 3 feet per second | Measurement Current | 0.5 mA or 1 mA | Sensor Length | 32 mm | 42 mm | Sensor Location | 5 mm from tip | Insulation Resistance | >1000 MΩ at room temperature | Sheath Material | Stainless Steel | Inconeltm |
Dimension | AM1710-12-ADT | AM1730-12-ADT | AM1751-12-ADT | AM1760-12-ADT | 0.25 in dia X 12 in (6.35 mm X 305 mm) | 0.25 in dia X 12 in (6.35 mm X 305 mm) | 0.25 in dia X 12 in (6.35 mm X 305 mm) | 0.25 in dia X 12 in (6.35 mm X 305 mm) | AM1710-BEND-ADT | AM1730-BEND-ADT | AM1751-BEND-ADT | AM1760-BEND-ADT | 0.25 in dia X 12 in (6.35 mm X 305 mm), 90° bend at 7.4 inch (190 mm) from probe end | 0.25 in dia X 12 in (6.35 mm X 305 mm), 90° bend at 9.6 inch (245 mm) from probe end | 0.25 in dia X 12 in (6.35 mm X 305 mm), 90° bend at 9.6 inch (245 mm) from probe end | 0.25 in dia X 12 in (6.35 mm X 305 mm), 90° bend at 9.6 inch (245 mm) from probe end | External Leads | Teflontm –insulated copper wire, 4 leads, 0.8 meters | Handle Dimension | 15 mm (OD) x 65 mm (L) | Handle Temperature Range[1] | -50°C to 160°C | -50°C to 180°C | Calibration | NIST traceable calibration with data included. Accredited calibration available per request. | [1] Handle temperatures outside this range will cause damage to the probe. | [2] Includes calibration and 100 hour drift. | [3] Probe calibration ranges may differ from probe temperature ranges (see Calibrated Accuracy for calibration ranges). | *PRT Information from www.accumac.com |
|