Pressure Specification
|
|
760-LLP |
760-D |
760-MA |
Max Pressure Range |
±30 inH2O (75 mbar) |
-12.5 to 35 psi (-0.86 to 2.5 bar) |
-12.5 to 300 psig (-0.86 to 20 bar.g) |
Accuracy |
0.05%FS[1] |
0.02%FS[1] |
0.02%FS[1][3] |
Stability |
0.005%FS[2] |
0.005%FS[2] |
0.005%FS[2] |
Pressure Type |
Differential, Gauge |
Differential, Gauge |
Gauge, Absolute |
Over Range Indication |
120% |
Resolution |
6 digits |
Measurement Units |
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, mH2O@68°F, inH2O@20°C, kgf/cm2, mtorr, torr, lb/ft2, tsi, custom |
Barometric Accuracy |
N/A |
N/A |
55 Pa[4] |
Connection |
Barb fitting |
Hose, 5 ft (1.5 m), with built-in filter to 1/4BSPF, 1/4NPTF, and M20F adapters |
Hose, 5 ft (1.5 m), with built-in filter to 1/4BSPF, 1/4NPTF, and M20F adapters |
Pressure Output Rate |
<30 Seconds (30 inH2O / 100 ml) |
<10 Seconds (35 psi / 5 ml) |
<90 Seconds (300 psi / 5 ml) |
[1] FS specification applies to the span of the module range. [2] Stability based on FS of the internal pressure module. Stability is 0.005%FS or 0.05 pa whichever is greater. Internal module is switchable. [3] Specification based on gauge measurement. An additional 55 pa uncertainty will need to be included when measuring in absolute mode. [4] 55 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long-term stability (<30 Pa per year). Barometer range of 60 to 110 kPa.
|
Selection Guide - Differential Pressure |
Module |
Module Range[6] |
Media |
Accuracy (%FS)[1] |
Burst Pressure |
760-LLP |
760-D |
760-MA |
inH2O |
mbar |
DP025 |
±0.25 |
±0.62 |
G |
0.2[2] |
100x |
• |
|
|
DP050 |
±0.5 |
±1.25 |
G |
0.1[3] |
100x |
• |
|
|
DP1 |
±1 |
±2.5 |
G |
0.05[4] |
100x |
• |
|
|
DP2 |
±2 |
±5 |
G |
0.05[4] |
100x |
• |
|
|
DP5 |
±5 |
±10 |
G |
0.05[4] |
50x |
• |
|
|
DP10 |
±10 |
±25 |
G |
0.05[4] |
20x |
• |
• |
|
DP20 |
±20 |
±50 |
G |
0.05 |
20x |
• |
• |
|
DP30 |
±30 |
±75 |
G |
0.05 |
20x |
• |
• |
|
DP50 |
±50 |
±125 |
G |
0.05 |
3x |
|
• |
|
DP100 |
±100 |
±250 |
G |
0.02 |
3x |
|
• |
|
DP150 |
±150 |
±350 |
G |
0.02 |
3x |
|
• |
|
DP300 |
±300 |
±700 |
G |
0.02 |
3x |
|
• |
|
DP400 |
-380 to 400 |
-950 to 1K |
G |
0.02 |
3x |
|
• |
|
DP800 |
-380 to 800 |
-950 to 2K |
G |
0.02 |
3x |
|
• |
|
DP1K |
-380 to 1K |
-950 to 2.5K |
G |
0.02 |
3x |
|
• |
|
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules. |
[2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS. |
[3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS |
[4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS. |
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT760-MA |
[6] The low module pressure range may be outside the pressure range of the calibrator |
* ADT155 Pressure Modules are calibrated in psi and inH2O. |
Selection Guide - Gauge Pressure |
Module |
Module Range[6] |
Media |
Accuracy (%FS)[1] |
Burst Pressure |
760-LLP |
760-D |
760-MA |
psi |
bar |
CP10 |
±10 |
±0.7 |
G |
0.02[5] |
3x |
|
• |
• |
CP15 |
-13.5 to 15 |
-0.95 to 1 |
G |
0.02[5] |
3x |
|
• |
• |
CP30 |
-13.5 to 30 |
-0.95 to 2 |
G |
0.02[5] |
3x |
|
• |
• |
CP35 |
-13.5 to 35 |
-0.95 to 2.5 |
G |
0.02[5] |
3x |
|
• |
• |
CP50 |
-13.5 to 50 |
-0.95 to 3.5 |
G |
0.02[5] |
3x |
|
|
• |
CP100 |
-13.5 to 100 |
-0.95 to 7 |
G |
0.02[5] |
3x |
|
|
• |
CP150 |
-13.5 to 150 |
-0.95 to 10 |
G |
0.02[5] |
3x |
|
|
• |
CP200 |
-13.5 to 200 |
-0.95 to 14 |
G |
0.02[5] |
3x |
|
|
• |
CP300 |
-13.5 to 300 |
-0.95 to 20 |
G |
0.02[5] |
3x |
|
|
• |
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules. |
[2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS. |
[3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS |
[4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS. |
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT760-MA |
[6] The low module pressure range may be outside the pressure range of the calibrator |
* ADT155 Pressure Modules are calibrated in psi and inH2O. |
Electrical Specification |
|
Range |
Resolution |
Accuracy |
Note |
mA Measure |
±30 mA |
0.0001 mA |
0.01%RD+0.005%FS |
Impedance<10Ω |
V Measure |
±30 V |
0.0001 V |
0.01%RD+0.005%FS |
Impedance >1MΩ |
mA Source |
24 mA |
0.001 mA |
0.01%RD+0.005%FS |
20 mA @ 1K |
Loop Power Source |
24 V |
N/A |
±1 V |
50 mA (Max Loading) |
Pressure Switch |
Open, close. Support for mechanical switches and NPN/PNP digital switches. |
Temperature Compensation |
41°F to 95°F (5°C to 35°C) |
Temperature Coefficient |
< ± ( 0.001%RD + 0.001%FS ) / °C outside of 5°C to 35°C |
General Specifications |
Specification |
Description |
Channels |
Four total: one electrical, one internal pressure, two external pressure (measure only) |
Enclosure IP Rating |
IP52 water and dust proof |
Battery |
Rechargeable Li-Ion battery, typically 10 hours of operation, recharges in less than 4 hours |
Display |
Color 800 x 480 TFT 5-inch touch screen |
Communications |
USB, WiFi |
Weight |
<4 lbs (<1.8 kg) |
Size |
9.3 x 4.3 x 2.8 in (235 x 110 x 70 mm) |
Certification |
ISO17025 accredited certificate with data included |
HART Communications |
Optional (ADT760-X-DL model) |
Data Logging |
Optional (ADT760-X-DL model), up to 1,000,000 readings (date and time stamped) |
Task Documentation |
Optional (ADT760-X-DL model) up to 1000 tasks |
Automation Functions |
Switch test, auto step, leak test |
Misuse Protection |
Up to 30 V on any two sockets |
Multi Lingual Interface |
English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian |
Pump Life |
>500,000 cycles |
Power |
Rechargeable Li-Ion battery, external power: 110/220 V power adapter 10 V |
Environment Specifications |
Operation: 32°F to 122°F (0°C to 50°C), 0-90% RH, less than 3,000 meters Compensated temperature: 32°F to 122°F (0°C to 50°C) Storage temperature: -4°F to 158°F (-20°C to 70°C) |
Vibration and Shock |
Vibration: 4g (20 to 2,000 Hz) Shock: 8g, 1 meter drop test |
Compliance |
CE |
Software |
ACal, PCal, Land, LogII |
Warranty |
1 year |
|