The Fluke i30s current clamp is based on Hall Effect technology for use in measurement of both DC and AC current. The i30s may be used in conjunction with oscilloscopes and other suitable recording instruments for accurate non-intrusive current measurement.
Specifications
Electrical specifications
Specified current range
30 mA to 30 A DC, 30 mA to 20 A AC rms
Usable current range
5 mA to 30 A DC, 30 mA to 20 A AC rms
Crest factor
1.4 in
Output sensitivity
100 mV/A
Accuracy (at +25°C)
±1% of reading ±2 mA
Resolution
±1 mA
Load impedance
> 100 kΩ
Conductor position sensitivity
±1% relative to center reading
Frequency range
DC to 100 kHz (-0.5 dB)
Phase shift below 1 kHz
< 2°
Temperature coefficient
±0.01% of reading / °C
Power supply
9 V Alkaline, MN1604/PP3, 30 hours, low battery indicator
Working voltage (see Safety Standards section)
300 V AC rms or DC
General Specifications
Maximum conductor size
19 mm diameter
Output connection
Safety BNC connector, supplied with safety 4 mm adapter
Output zero
Manual adjust via thumbwheel
Cable length
2 m
Operating temperature range
0°C to +50°C
Storage temperature range (with battery removed)
-20°C to +85°C
Operating humidity
15% to 85% (non-condensing)
Weight
250 g
Safety standards
BS EN 61010-1: 2001
BS EN 61010-2-032: 2002
BS EN 61010-031: 2002
300 Vrms, Category III, Pollution Degree 2
Use of the probe on uninsulated conductors is limited to 300 V AC rms or DC and frequencies below 1 kHz
EMC standards
EN 61326: 1998 +A1, A2, and A3