The GW Instek GPT-15001 is the first safety analyzer in the world to comply with IEC 61010-2-034. The GPT-15001 executes AC withstanding voltage test and continuity test.
Includes:
• Quick Start Guide x1, CD x1(completed user manual) • Power cord x1, Interlock key x1 • Remote terminal Cable GHT-119 x1 • Test lead GHT-115 x1, GTL-215 x1
Specifications
AC WITHSTANDING
Output-Voltage Range
0.050kV~5.000kV
Output-Voltage Resolution
1V
Output-Voltage Accuracy
± (1% of setting + 5V) [no load]
Maximum Rated Load
500 VA (5kV/100mA)
Maximum Rated Current
100mA (0.5kV< V ≦5kV)
10mA (0.05kV≦ V ≦0.5kV)
Output-Voltage Waveform
Sine wave
Output-Voltage Frequency
50 Hz / 60 Hz selectable
Voltage Regulation
± (1% + 5V) [maximum rated load → no load]
Voltmeter Accuracy
± (1% of reading + 5V)
Current Measurement Range
1μA~100.0mA
Current Best Resolution
1μA / 10μA / 100μA
Current Measurement Accuracy
±(1.5% of reading + 30μA)
Current Offset
60μA Maximum
Window Comparator Method
Yes
ARC Detect
Yes
RAMP UP (Rise T ime)
0.1s~999.9s
RAMP DOWN (Fall Time)
0.0s~999.9s
TIMER (Test Time)*
OFF, 0.3s~999.9s
TIMER Accuracy
±(100ppm + 20ms)
WAIT TIME
0.0s~999.9s
GND
ON/OFF
DC WITHSTANDING
Output-Voltage Range
0.050kV~6.000kV
Output-Voltage Resolution
1V
Output-Voltage Accuracy
± (1% of setting + 5V) [no load]
Maximum Rated Load
100W (5kV/20mA)
Maximum Rated Current
20mA (0.5kV< V ≦6kV)
2mA (0.05kV≦ V ≦0.5kV)
Voltage Regulation
± (1% + 5V) [maximum rated load → no load]
Voltmeter Accuracy
± (1% of reading + 5V)
Current Measurement Range
1μA~20.00mA
Current Best Resolution
0.1μA /1μA /10μA
Current Measurement Accuracy
±(1.5% of reading + 3μA) when I Reading < 1mA
±(1.5% of reading + 30μA) when I Reading ≧1mA
Current Offset
5μA Maximum
Window Comparator Method
Yes
ARC Detect
Yes
RAMP UP (Rise T ime)
0.1s~999.9s
RAMP DOWN (Fall Time)
0.0s~999.9s
TIMER (Test Time)*
OFF, 0.3s~999.9s
TIMER Accuracy
±(100ppm + 20ms)
WAIT TIME
0.0s~999.9s
GND
ON/OFF
INSULATION RESISTANCE
Output Voltage
50V~1200V dc
Output-Voltage Resolution
50V
Output-Voltage Accuracy
± (1% of setting + 5V) [no load]
Resistance Measurement
Test Voltage
Display Range
Measurement Range / Accuracy
50V≦V≦100V
0.1MΩ~ 10.00GΩ
0.1MΩ~1MΩ:±(5% of reading + 3 count)
1 MΩ~50MΩ:±(5% of reading + 1 count)
51MΩ~2GΩ:±(10% of reading + 1 count)
150V≦V≦450V
0.1MΩ~ 20.00GΩ
500V≦V≦1200V
0.1MΩ~ 50.00GΩ
0.1MΩ~1MΩ:±(5% of reading + 3 count)
1 MΩ~500MΩ:±(5% of reading + 1 count)
501MΩ~9.999GΩ:±(10% of reading + 1 count)
10GΩ~50GΩ:±(20% of reading + 1 count)*
Voltage Regulation
± (1% + 5V) [maximum rated load → no load]
Voltmeter Accuracy
± (1% of reading + 5V)
Short-Circuit Current
10mA max.
Output Impedance
2kΩ
Window Comparator Method
Yes
RAMP UP (Rise T ime)
0.1s~999.9s
RAMP DOWN (Fall Time)
0.0s~999.9s
TIMER (Test Time)*
OFF, 0.3s~999.9s
TIMER Accuracy
±(100ppm + 20ms)
WAIT TIME
0.0s~999.9s
GND
ON/OFF
Ground Bond
Output-Current
03.00A~32.00A ac
Output-Current Resolution
0.01A
Output-Current Accuracy
3A≦ I≦ 8A:±(1% of reading + 0.2A)
8A<I≦ 32A:±(1% of reading + 0.05A)
Test-Voltage
8Vac max (open circuit)
Test-Voltage Frequency
50Hz/60Hz selectable
Ohmmeter Measurement Range
1mW~ 650mW
Ohmmeter Measurement Resolution
0.1mW
Ohmmeter Measurement Accuracy
±(1% of reading + 2 mW)
Window Comparator Method
Yes
TIMER (Test Time)
0.3s~999.9s
TIMER Accuracy
±(100ppm + 20ms)
Test Method
Four Terminal
GND
ON/OFF
Continuity Test
Output-Current
100mA dc (fixed)
Ohmmeter Measurement Range
0.10W~ 70.00W
Ohmmeter Measurement Resolution
0.01W
Ohmmeter Measurement Accuracy
±(10% of reading + 2 W)
Window Comparator Method
Yes
TIMER (Test Time)*
0.3s~999.9s
TIMER Accuracy
±(100ppm + 20ms)
MEMORY
Single Step Memory
MANU:100 blocks
Automatic Testing Memory
AUTO:100 blocks, manu per auto:10
INTERFACE
REMOTE (Front) terminal
Standard
USB host (Front)
Standard
Rear Output
Standard
RS-232C
Standard
USB device
Standard
Signal I/O
Standard
GPIB
Option
LAN
Option
DISPLAY
7” color LCD
POWER SOURCE
AC 100V~240V ± 10%, 50Hz/60Hz
DIMENSION & WEIGHT
GPT-15001/15002/15003
GPT-15004
380(W) x 148(H) x 492(D) mm; Approx. 17kg
380(W) x 148(H) x 546(D) mm; Approx. 21kg
* When Ground Mode is “ON”, the measurement range is 30GΩ max. and adding 10% error for accuracy.
The store will not work correctly when cookies are disabled.