General | DSP Measurement Technique | Transit time | Flow Velocity Range | 0.1 m/s - 20 m/s | Accuracy | Pipe ID >75 mm | ±0.5% to ±2% of flow reading for flow rate >0.2 m/s | | Pipe ID 13 mm - 75 mm | ±3% of flow reading for flow rate >0.2 m/s | All pipe ID's | ±6% of flow reading for flow rate <0.2 m/s | Repeatability | ±0.5% of measured value or ±0.02 m/s whichever is the greater | Response Time | < 500 ms depending on pipe diameter | Selectable Flow Units | Velocity | m/sec, ft/sec. | | Volume | "l/s, l/min, l/h, gal/min, gal/h, USgals/min, USgals/h, Barrel/h, Barrel/day, m³/s, m³/min, m³/h" | Selectable Total Volume Units | liter, gallon, US gallons, Barrel, m³ | Total Volume | 12 digits | Menu Languages | EN, DE, FR, RU, SWE, IT, SP, P, NO, DEN | Environmental | Operating Temperature | -20 °C to 50 °C | -4 °F to 122 °F | Storage Temperature | -25 °C to 65 °C | -13 °F to 149 °F | Pipe Wall Temperature | -20 °C to 135 °C | -4 °F to 275 °F | Operating Humidity | Max. 90% relative humidity @ 50°C (122 °F) | Applicable Pipe Types | Pipe Materials | PVDF-SYGEF, PP-PROGEF, PE-ELGEF, PB-INSTAFLEX, ABS, PVC-U/PVC-C, Mild Steel, Ductile Iron, Stainless Steel 316, Copper | Pipe Dimension (OD) | Type PF220 | 13 mm to 1000 mm | 0.5 in. to 39 in. | | Type PF330 | 13 mm to 2000 mm | 0.5 in. to 78 in. | Pipe Wall Thickness | 1 mm to 75 mm | 0.04 in. to 3 in. | Pipe Lining | Applicable pipe linings include Rubber, Glass, Concrete, Epoxy, Steel | Pipe Lining Thickness | 0 mm to 10 mm | 0 in. to 0.4 in. | Electrical | Supply Voltage | 9 to 24 V DC | Power Consumption | Max. 10.5 W | Battery | | Technology | 5-cell NiMH | Capacity | 3.8 Ah | Operating Time (typical) | Typically 20 hours continuous with backlight and 4-20mA output OFF | Recharge Time | 6.5 h | Service Life | >500 charge/discharge cycles | Power supply | Input Voltage | 90 to 264 V AC (47 to 63 Hz) | Output Voltage | 12 V DC | Output Current Max. | 1.5 A | Approvals | UL, CUL, TUV, CB, CE | Outputs | Analog Output | Range | 4 to 20 mA, 0 to 20 mA, 0 to 16 mA | | Resolution | 0.1% of full scale | Load Max. | 620 Ω | Isolation | 1500 V Opto-isolated | Alarm Current | Adjustable between 0 to 26mA | Pulse Output | Type | Digital MOSFET relay | | Pulse Repetition | Max. 500 pps, user programmable | Pulse Width | 5 - 500 ms, user programmable | Voltage Max. | 48 V | Current Max. | 500 mA | Isolation | 1500 V opto isolated | USB Interface (PF330 only) | Protocol | Supports full speed (12Mbits/sec) data connection | | Software | USB driver software is provided with the package | Connector | Proprietary industrial connector | RS-232 Interface (PF330 only) | Protocol | Serial RS-232 communication including handshaking | | Connector | Proprietary industrial connector (GND, RxD, TxD, DTR, DSR) |
Data Logger (PF330 only) | Data Logged | Log application details, flow rate, total flow, unit, time stamp | Number of Data Points | 198 k | Number of Data Sites | 20 | Number of Data Points per Site | No limit (max. 198k) | Programmable Logging Interval | 5 s - 1 h | Start / Stop | Manually or timer controlled | Data Download | Via RS-232 / USB interface | Transducer Sets | Type A | Type PF220 & PF330 | 13 - 114 mm pipe O.D. (2MHz) | Type B | Type PF220 | 115 - 1000 mm pipe O.D. (1MHz) | | Type PF330 | 115 - 2000 mm pipe O.D. (1MHz) | Enclosure and Display | Material | ABS | Dimensions | 264 x 168 x 50 mm | 10.4 x 6.6 x 2.0 inch | Weight | 1.1 kg (incl. battery) | 2.45 lb | Keypad | 16 key tactile feedback membrane keypad | Display | Type | 240 x 64 pixel graphic display, high contrast black-on-white, with backlight | | Viewing angle | Min. 30°, typically 40° | Active area | 127 x 34 mm | 5 x 1.3 inch | IP Rating | IP 54 | Shipping Weight | | PF330 | PF220 | Box dimensions | 420 x 390 x 220 mm | 16.5 x 15.4 x 8.7 inch | 510 x 140 x 440 mm | 20 x 5.5 x 17.3 inch | Weight | 7.5 kg | 16.5 lb | 6 kg | 13.2 lb | Volumetric Weight | 5.7 kg | 12.5 lb | 5.2 kg | 11.5 lb | Standards and Approvals | | CE, RoHS compliant | Safety | BS EN 61010 | EMC | BS EN 61326 - 1:2006 | BS EN 61326-2-3:2006 | Power Supply | EN61204 - 3 | UL, CUL, TUV, CB, CE |
|