The Fuji Electric PHF is a 3 or 6 channel paperless recorder that features a wide variety of display modes to allow users to view data as trends (horizontally and vertically), bar graphs, analog meters, digital/numeric output, totalized output, event summaries and more. The PHF allows users to record thermocouples, RTDs, DC voltage, and current inputs on one unit.
Specifications
General specifications
Mounting method
Panel mount
Dimensions
160 (H) × 144 (W) × 185 (D) mm
Power supply voltage
100–240 V AC, 50/60 Hz
Power consumption
Approx. 42 VA (at 240 V AC)
Terminal
M3 screw terminal
Input section
Number of input
3 or 6 (as specified by order)
Measurement cycle
100 ms
Input signal
Thermocouple (type B, R, S, K, E, J, T, N, W, L, U, or PN) RTD (Pt100) DC voltage (50 mV, 500 mV, or 5 V) DC current (with shunt resister)
Burnout function
Provided as standard for thermocouple or RTD input
Display unit
Display
5.7" TFT color LCD (320 x 240 dots) with automatic display off function
Life of backlight
50,000 hours (at 20°C)
Display contents
Trend, bar graph, digital display, event summery, tag, scale Historical trend, parameter, Ethernet log
Recording function
Storage medium
CompactFlash
Recording capacity
2 GB max.
Data saving cycle
Links to refreshment cycle of the trend display
Trend data
Average, instantaneous value, or maximum and minimum value are saved.
Event data
Alarm data, message data, power on and off are saved.
Setting data
Parameter setting is saved.
Data format
ASCII (readable by Excel) Binary (not readable by Excel)
Storage capacity
About 4 years' worth at a display refresh cycle of 30 seconds (ASCII, 6 channels, 512 MB CompactFlash)
Memory consumption
The display unit displays how much memory has been used. A user can select whether to stop recording or to delete stored file.
Calculation function
Difference between channels, square root extraction
Alarm
Number of settings
Up to 4 alarms for each channel
Alarm type
High/low limits
Indication
Alarm status is displayed on digital display unit when an alarm occurs. Histories are displayed in the alarm summary.
Relay contact output
up to 10 (option)
Performance
Indication accuracy
±(0.15% + 1 digit) of input range
Indication resolution
0.1°C
Reference junction compensation error
±0.5°C
Others
Memory backup
Parameter settings are saved to the internal nonvolatile memory. The clock is backed up by a built-in lithium battery. Trend data is stored in CompactFlash with no back-up.
Option
Alarm output
Relay output, 10 points
Digital input
No-voltage contact input, 5 points
Communication
Ethernet
Accessories
CD-ROM (parameter loader software, instruction manual), panel mounting bracket, noise filter for power cable