The Crystal Engineering m1M 3000 PSIG digital pressure gauge provides users with a 0.25% of reading accuracy over a wide temperature range.
Specifications
ACCURACY – 3 YEAR RECOMMENDED CALIBRATION INTERVAL
0 to 20% of Range:
±(0.05% of Full Scale)
Includes all effects of linearity, hysteresis, repeatability,temperature, and stability for three years.
20% to 110% of Range:
±(0.25% of Reading)
All models indicate vacuum, but vacuum specification appliesto 300 psi / 20 bar / 2 MPa models only.
Vacuum:
±(0.5% of Full Scale*)
* Full Scale = -14.5 psig / -1 bar / -0.1 MPa
Not recommended for continuous use below -14.5 psi. Refer toXP2i-DP data sheet for gauges that are intended for continuoushigh vacuum use.
OPERATING TEMPERATURE
Temperature Range:
-10 to 50° C (14 to 122° F)
Non-condensing. No change in accuracy over operatingtemperature range. Gauge must be zeroed to achieve ratedspecification.
DISPLAY
Screen:
5 digits, 16.8 mm (0.66") high
LCD readable in sunlight with bright backlight.
Display Rate:
4 readings/second
POWER
2 x AA:
900 hours, typical
Uses 2 alkaline AA (LR6) batteries. Use of backlight reducesoperating time. Extended battery life is keypad selectable. Donot overtighten rear cover when replacing batteries.
Water resistant housing incorporating o-ring seals [IEC 60529].
Housing:
Polyester/Polycarbonate Blend
LCD protected from impact damage by polycarbonate lens underpolyester window.
Keypad and Labels:
UV Resistant Polyester
Skydrol® compatible.
SENSOR
Wetted Materials:
(wrench tight) 316 stainless steel
(finger tight) 316 stainless steel and Viton® (internal o-ring)
All welded, with a permanently filled diaphragm seal.
Metal to metal cone seal; O-ring can be removed if necessary.
1/4" medium pressure tube system compatible with HiP LM4 andLF4 Series, Autoclave Engr SF250CX Male and Female Series.
Diaphragm Seal Fluid:
Silicone Oil
1/4" male NPT adapter included unless BSP or M20 is specified.
Connection:
Crystal CPF* Female
*U.S. Patent No. 8,794,677
STORAGE TEMPERATURE
Temperature Range:
-40 to 75° C (-40 to 167° F)
Batteries should be removed if stored for more than one month.
SPECIAL FEATURES
Averaging:
Smooths unsteady readings (2 to 20 samples)
Vertical Bar Graph:
Percent pressure indication for quick reference
Capture:
Minimum (LO) and Maximum (HI)