The Wika Mensor CPG2400 Digital Barometer is a low cost digital benchtop Barometer with a range of 8 to 17 psia (Custom ranges available, please call).The CPG2400 has an accuracy of 0.03%of reading and is an excellent replacement for Mercury manometers. Alternative ranges can be specified in a variety of units including inches of Mercury (inHg A), millibar (mbar) and kilo pascals (kPa). Built-in peak and null functions provide a convenient way to analyze barometric trends. Applications The CPG2400 Digital Barometer is used as a desk top or rack mounted barometric reference for weather stations, laboratories, clean rooms, nuclear power plants, pharmaceutical processing, health care equipment manufacturing and Calibration of medical radiation equipment. The CPG2400 Digital Barometer has a clear LCD display with an easy menu structure for easy navigation within the setup screens. A membrane key pad provides the technician the capability to navigate within these menus to set the digital display features such as peak, null, pressure units. It also allows access to a password protected calibration section for re-calibration of the internal sensor. The CPG2400 Digital Barometer can be operated from a remote computer via a RS-232 (Standard) or RS-485( Optional) serial port connection. The serial port can be used to configure, read or calibrate the sensor. This is useful when remote indication of the digital barometric pressure is required. RS-232 is used for communication over short distances and the RS-485 for longer distances and multi-drop applications.
Specifications
Accuracy (1)
0.03% R
Pressure range
psia
8 to 17 psia (other unit ranges are available within the equivalent psi range)
Resolution
Digits
5
Overpressure limits
% FS
150
Compensated temperature
°C
15 to 45
Storage temperature
°C
-20 to 70
Warm-up time
Min.
<1
Reading rate
/sec
~4.6 sensor readings
Response time
mS
<252
Case size
in.
2.6 high x 4.2 wide x 4.9 deep
Weight
lbs.
<2
Orientation
Negligible, can be removed with re-zeroing
Communications
RS-232 or RS-485 (100 ft cable length recommended)
Media compatibility
Pressure port: Clean, dry, non-corrosive, non-combustible, non-oxidizing gases for all rated ranges. Ranges ≥ 7.5 psi: All other media compatible with aluminum, 316 stainless steel, brass, Buna N, Viton, sealant, silicone grease and RTV. Not designed for oxygen service. Cannot guarantee accuracy on media other than gases. Reference port: Clean, dry, non-corrosive gases.
(1) Accuracy is defined as the total measurement uncertainty, which is expressed with the coverage factor (k=2) and includes the following factors: the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range during a periodic zero point adjustment.