The APM CPF Series portable pressure modules offer the functionality and flexibility to perform pressure calibrations with a multifunction signal calibrator. Or, combine with a hand-help pressure calibrator to expand its range. The APM delivers the same high level accuracy whether in the lab or in the field.
Specifications
ACCURACY
Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.
All models indicate vacuum, but vacuum specification applies to 30, 100, and 300 psi models only.
Not recommended for continuous use below -14.5 psi. Refer to the XP2i-DP data sheet for gauges that are intended for continuous high vacuum use.
psi (Gauge Pressure)
bar (Gauge Pressure)
30, 100, and 300 psi modules
3, 10, and 30 bar modules
0 to 30% of Range:
± (0.0075% of Full Scale)
0 to 30% of Range:
± (0.0075% of Full Scale)
30 to 110% of Range:
± (0.025% of Reading)
30 to 110% of Range:
± (0.025% of Reading)
Vacuum (30 & 100 psi)
± (0.06 of Full Scale*)
Vacuum (3 & 10 bar)
± (0.06 of Full Scale*)
Vacuum (300 psi)
± (0.06 of Full Scale*) ± 1 LSD
Vacuum (30 bar)
± (0.06 of Full Scale*) ± 1 LSD
* Full Scale = -14.5 psi
* Full Scale = -14.5 psi
1000 and 3000 psi modules
100 and 300 bar modules
0 to 30% of Range:
± (0.015% of Full Scale)
0 to 30% of Range:
± (0.015% of Full Scale)
30 to 110% of Range:
± (0.05% of Reading)
30 to 110% of Range:
± (0.05% of Reading)
10 000 and 15 000 psi modules
700 and 1000 bar modules
0 to 30% of Range:
± (0.03% of Full Scale)
0 to 30% of Range:
± (0.03% of Full Scale)
30 to 110% of Range:
± (0.1% of Reading)
30 to 110% of Range:
± (0.1% of Reading)
psiA (Pressure with BARO module)
barA (Pressure with BARO module)
30 psi module
3 bar module
0.200 to 14.500:
± 0.011 psiA, typical
0.0138 to 1.0000:
± 0.0008 barA, typical
14.5000 to 44.500:
± (0.025% of Reading) + 0.003 psiA
1.0000 to 4.0000:
± (0.025% of Reading) + 0.0003 barA
100 psi module
10 bar module
0.200 to 14.500:
± 0.011 psiA, typical
0.0138 to 1.0000:
± 0.0008 barA, typical
14.500 to 44.500:
± 0.011 psiA
1.0000 to 4.0000:
± 0.0010 barA
44.500 to 114.500:
± (0.025% of Reading)
4.0000 to 11.0000:
± (0.025% of Reading)
300 psi module
30 bar module
0.20 to 14.50:
± 0.01 psiA, typical
0.014 to 1.000:
± 0.001 barA, typical
14.50 to 104.50:
± 0.03 psiA
1.000 to 10.000:
± 0.003 barA
104.50 to 314.50:
± (0.025% of Reading)
10.000 to 31.000:
± (0.025% of Reading)
1000 psi module
100 bar module
14.5 to 314.5:
± 0.2 psiA
1.000 to 31.000:
± 0.015 barA
314.5 to 1014.5:
± (0.05% of Reading)
31.000 to 101.000:
± (0.05% of Reading)
3000 psi module
300 bar module
14.5 to 914.5:
± 0.5 psiA
1.00 to 91.00:
± 0.05 barA
914.5 to 3014.5:
± (0.05% of Reading)
91.00 to 301.00:
± (0.05% of Reading)
10 000 psi module
700 bar module
15 to 3015:
± 3 psiA
1.00 to 211.00:
± 0.21 barA
3015 to 10 015:
± (0.1% of Reading)
211.00 to 701.00:
± (0.1% of Reading)
15 000 psi module
1000 bar module
15 to 4515:
± 5 psiA
1.00 to 301.00:
± 0.30 barA
4515 to 15 015:
± (0.1% of Reading)
301.00 to 1001.00:
± (0.1% of Reading)
SENSOR
Wetted Materials:
316 stainless steel*
*Wrench tight. Finger tight also includes Viton o-ring.
Diaphragm Seal Fluid:
Silicon Oil
All welded, with a permanently filled diaphragm seal.
Connection:
Crystal CPF Female
Metal/Metal cone seal; O-ring can be removed.
1/4" male NPT adapter included unless BSP, M20 or 15KPSI is specified.
1/4" medium pressure tube system compatible with HiP LM4 and LF4 Series, Autoclave Engr SF250CX Male and Female Series.
OPERATING TEMPERATURE
Temperature Range:
-20 to 50°C (-4 to 122°F)
<95% RH, non-condensing. No change in accuracy over operating temperature range.
Applies to all modules.
ENCLOSURE
Weight:
680 g (24.0 oz)
Weight includes one pressure module, one RTD module, 4AA battery module, and protective boot.
Housing:
Impact resistant injection molded
LCD protected from impact damage by 1.5 mm (0.06") thick polycarbonate lens.
Dimensions:
M4 x 0.7 [8 mm (0.31")] deep threaded insert mounting locations
STORAGE TEMPERATURE
Temperature Range:
-40 to 75°C (-40 to 167°F)
Batteries should be removed if stored for more than one month.
The store will not work correctly when cookies are disabled.
We use cookies to make your experience better.To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies.Learn more.