Standard Reference Transducers, Model CPR2550 | Accuracy(1) | 0.01%FS(3) | 0.01%IS-50(2) | Gauge Pressure | 0 ... 0.36 up to 10,000 psi 0 ... 25 mbar up to 700 bar | 0 ...14.5 up to 6000 psi 0 ... 1 up to 400 bar | Bi-Directional Pressure | -0.18. ... 0.18 to -14.5 ... 10,000 psi -12.5 ... 12.5 mbar to -1 ... 700 bar | -14.5 ... 145 to -14.5 ... 6000 psi -1 ... 10 to -1 ... 400 bar | Absolute Pressure(4) | 0 ... 7.5 up to 10,015 psi 0 ... 0.5 up to 701 bar | 0 ... 14.5 up to 6015 psi 0 ... 1 up to 401 bar | Precision(5) | 0.004%FS | 0.004%FS | Calibration Interval | 365 days(6) | 365 days | Media Compatibility | Metals in contact with media | 6000/7000 series Aluminum, 316 SS, brass | Non-metals in contact with media | PTFE (Teflon®), Urethane, Silicone, RTV, Silicone grease, PVC, Epoxy, Buna-N, fluoroelastomers (Viton®) | Sensor | Reading rate | 33 readings/second | Calibration adjustments | Internal zero adder and span multiplier, up to 11 point linearization for each sensor |
Premium Reference Transducers, Model CPR2580 | Accuracy(1) | 0.008% IS-33(7) | 0.008% IS-50(8) | 0.01% FS(3) | 0.014% FS(3) | Gauge Pressure | 0 ... 12 to 0 ...16.5 psig 0 ... 17.5 to 0 ... 33 psig 0 ... 80 to 0 ... 110 psig 0 ... 120 to 0 ... 220 psig | — | — | — | Absolute Pressure(4) | 0 ... 12 to 0 ...16.5 psia 0 ... 18.4 to 0 ... 33 psia 0 ... 36 to 0 ... 50 psia 0 ... 80 to 0 ... 110 psia 0 ... 160 to 0 ... 220 psia 0 ... 240 to 0 ... 500 psia | 0 ... 700 to 0 ... 1100 psia 0 ... 1400 to 0 ... 3300 psia 0 ... 4200 to 0 ... 6015 psia | 0 ... 8000 to 0 ... 11,000 psia | 0 ... 12,000 to 0 ... 22,000 psia 0 ... 24,000 to 0 ... 31,500 psia 0 ... 32,000 to 0 ... 42,000 psia | Precision(5) | 0.004% FS | 0.004%FS | 0.004%FS | 0.004%FS | Calibration Interval | 365 days | 365 days | 365 days | 365 days | Media Compatibility | Metals in contact with media | 6000/7000 series aluminum, 316 SS, brass, inconel | Non-metals in contact with media | PTFE (Teflon®), Urethane, Silicone, RTV, Silicone grease, PVC, Epoxy, Buna-N, fluoroelastomers (Viton®) | Sensor | Reading rate | 10 readings/second | Calibration adjustment | Internal Zero adder and Span multiplier, up to 11 point linearization for each sensor |
(1) It is defined by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range with recommended zero point adjustment every 30 days. (2) 0.01 % IS-50 accuracy: Between 0 ... 50 % of the full scale, the accuracy is 0.01% of half of the full scale value and between 50 ... 100 % of the full scale, the accuracy is 0.01 % of reading. (3) FS = full span. (4) The minimum calibrated range of absolute transducer(s) is 600 mTorr. (5) It is defined as the combined effects of linearity, repeatability and hysteresis throughout the stated compensated temperature range. (6) 180 days for pressure ranges below 1 bar (15 psi) and above 400 bar (6,000 psi) gauge or absolute, and -1...1 bar (-15 ...14.5 psi) bidirectional. 365 days for the remainder of the specified ranges. (7) 0.008 % IS-33 accuracy: Between 0 ... 33 % of the full scale, the accuracy is 0.008% of one third of the full scale value and between 33 ... 100 % of the full scale, the accuracy is 0.008 % of reading. (8) 0.008 % IS-50 accuracy: Between 0 ... 50 % of the full scale, the accuracy is 0.008% of half of the full scale value and between 50 ... 100 % of the full scale, the accuracy is 0.008 % of reading. Basic Instrument | Instrument | Instrument version | Standard: Table top with tilt feet Option: -19” rack-mounting with side panels incl. rack-mounting kit for single instrument mount. -19” rack-mounting with side panels incl. rack-mounting kit for dual instrument mount. | Dimensions | See Data Sheet | Weight | 12.5 lbs./ 5.7 kg (with all internal options) | Warm-up time | Approximately 15 minutes | Display | Screen | 7” color LCD | Resolution | Selectable from 4 to 7 digits, depending on range and units | Data entry | Touch screen keypad | Measurement Units | psi, psf, osi, atm, inH20@4C, inH20@20C, inH20@60F, mbar, bar, Dy/cm2, pascal, hPa, kPa, MPa, inHg@0C, inHg@60F, mTorr, Torr, mmHg@0C, cmHg@0C, mHg@0C, mmH20@4C, cmH20@4C, mH20@4C, mmH20@20C, cmH20@20C, mH20@20C, mSW, ftH20@4C, ftH20@20C, ftH20@60F, inSW, ftSW, tsi, tsf, g/cm2, kg/cm2, kg/m2, % of Range, + plus 2 user defined units (multiplier from psi, bar or pascal) | Rate Units | /sec., /min., /hr., /3-hr | Languages | English, German, Spanish, French, Italian, Portuguese, Polish, Russian, Chinese, Japanese, Korean | Measurement filters | Off, Low, Normal (default), High | Connections | Number of integrated transducer (selectable) | Standard: 1 reference transducer Optional: 2nd reference transducers, external transducer, internal barometric reference | Pressure connections | To 6000 psi: 7/16 - 20 female SAE/MS. 1/8” FNPT adapters provided. Ranges >6000 psi: Autoclave F250C/HIP HF4 | Pressure adaptors | Standard: None Optional: 1/4 inch Swagelok® tube fittings, 6 mm Swagelok® tube fittings, 1/8 in. female NPT fittings or 1/8 in. female BSP fittings. | Overpressure limits | 110 % FS typical, optional external relief valves are available | Voltage supply | Power input requirements | 100-120 or 200-240 VAC, 50-60Hz, 24VA max | Switching power supply | Output: 12 VDC, 1.67 A (includes 4 region specific plugs adapters) | Permissible ambient conditions | Storage temperature range | 0 to 70 deg C | Operating environment | 0 ... 95 % RH (relative humidity, non-condensing) | Operating temperature range | 15 ... 40 deg C | Communications | Remote interface | IEEE 488, RS--232, USB and Ethernet | Command sets | Mensor, WIKA SCPI |
CE conformity and certificates | CE compliance | EN61326-1:2013 electromagnetic compliance EN61010-1:2010 safety/CB scheme | Calibration | Calibration certificate per ISO/IEC 17025:2005. Accreditation is by the American Association for Laboratory Accreditation (A2LA). |
|