The GPT-9603 safety tester is a 3-in-1 instrument for voltage safety withstand testing (hipot) and insulation resistance testing. Perform AC hipot tests to 5000 volts AC or 6000 VDC. The GPT-9603 also performs insulation resistance testing to 2000 M-Ohms. The bright blue Dot Matrix LCD gives the test engineer all they need at their finger tips to meet IEC, EN, UL, ETL, CSA, GB, JIS and other safety standards. Custom calibration certificates are available to help the end user prove traceability and accuracy to their standards inspection body or organization.
Specifications
AC WITHSTANDING (All Models)
Output-Voltage Range
0.10kV~ 5.00kV ac
Output-Voltage Resolution
10V
Output-Voltage Accuracy
± (1.5% of setting + 2 counts) with no load
Maximum Rated Load
100 VA (5kV/20mA)
Maximum Rated Current
20mA (0.5kV < V ≦ 5kV)
5mA (0.1kV ≦ V≦ 0.5kV)
Output-Voltage Waveform
Sine wave
Output-Voltage Frequency
50 Hz / 60 Hz selectable
Voltage regulation
± (1.5% + 2 counts) [full load → no load]
Voltmeter Accuracy
± (1.5% of reading + 2 counts)
Current Measurement Range
0.01mA~20.0mA
Current Best Resolution
0.01mA / 0.1mA
Current Measurement Accuracy
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
Current Judgment Accuracy
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
Window Comparator Method
Yes
ARC Detect
Yes
RAMP (Ramp-Up Time)
0.1s fixed
TIMER (Test Time)
OFF, 1s~180s
GND
ON
DC WITHSTANDING (GPT-9602 & GPT-9603)
Output-Voltage Range
0.10kV~ 6.00kV dc
Output-Voltage Resolution
10V
Output-Voltage Accuracy
± (1.5% of setting + 2 counts) with no load
Maximum Rated Load
25W (5kV/5mA)
Maximum Rated Current
6mA (0.5kV < V ≦ 6kV)
2mA (0.1kV ≦ V≦ 0.5kV)
Voltage regulation
± (1.5% + 2 counts) [full load → no load]
Voltmeter Accuracy
± (1.5% of reading + 2 counts)
Current Measurement Range
0.01mA~6.00mA
Current Best Resolution
0.01mA
Current Measurement Accuracy
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
Current Judgment Accuracy
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
Window Comparator Method
Yes
ARC Detect
Yes
RAMP (Ramp-Up Time)
0.1s fixed
TIMER (Test Time)
OFF, 1s~180s
GND
ON
INSULATION RESISITANCE (GPT-9603 & GPT-9612)
Output-Voltage
50V, 100V, 250V, 500V, 1000V dc
Output-Voltage Accuracy
± (3.0% of setting + 1 count) [no load]
Resistance Measurement Range
1MW~ 2000MW
Test voltage
Measurable Range
Accuracy
50V/100V/250V
1~ 50MW
51~ 2000MW
±(5% of reading + 2MW)
±(10% of reading + 2MW)
500V/1000V
1~ 500MW
501~ 2000MW
±(5% of reading + 2MW)
±(10% of reading + 2MW)
Window Comparator Method
Yes
Output Impedance
600kW
RAMP (Ramp-Up Time)
0.1s fixed
TIMER (Test Time)
OFF, 1s~180s
GND
OFF (fixed)
TEST MODE*
SINGLE
ACW, DCW, IR
AUTO
AC-IR, IR-AC, DC-IR, IR-DC
INTERFACE
Remote Terminal (Front)
Standard
SIGNAL IO
Standard
General
DISPLY
240 x 48 Ice Blue dot matrix LCD
POWER SOURCE
AC 100V-120V/220V-240V ±10%, 50/60Hz
POWER COMSUMPTION
400VA max.
DIMENSIONS & WEIGHT
330(W) x 148(H) x 385(D) mm
approx.9kg(Max)
* The available “Test Mode” depends on selected model