The GPT-9601 safety Hipot tester is capable of performing AC withstanding testing up to 5000 Volts AC. The AC test output voltage can be adjusted from 100 to 5000 VAC to meet UL, CSA, IEC and ETL testing requirements. The GPT-9601 also accurately monitors AC trip currents by using a true rms measurement. The GPT-9600 series is equipped with the high-efficiency PWM amplifier. The bright blue Dot Matrix LCD gives the test engineer all they need at their finger tips to meet IEC, EN, UL, ETL, CSA, GB, JIS and other safety standards. Custom calibration certificates are available to help the end user prove traceability and accuracy to their standards inspection body or organization.
The GW Instek GPT series is available with an optional ISO-17025 accredited calibration which is required by most testing agencies. If you are not sure about your hipot testing needs, feel free to talk to one of our testing experts.
Specifications
AC WITHSTANDING (All Models)
Output-Voltage Range
0.10kV~ 5.00kV ac
Output-Voltage Resolution
10V
Output-Voltage Accuracy
± (1.5% of setting + 2 counts) with no load
Maximum Rated Load
100 VA (5kV/20mA)
Maximum Rated Current
20mA (0.5kV < V ≦ 5kV)
5mA (0.1kV ≦ V≦ 0.5kV)
Output-Voltage Waveform
Sine wave
Output-Voltage Frequency
50 Hz / 60 Hz selectable
Voltage regulation
± (1.5% + 2 counts) [full load → no load]
Voltmeter Accuracy
± (1.5% of reading + 2 counts)
Current Measurement Range
0.01mA~20.0mA
Current Best Resolution
0.01mA / 0.1mA
Current Measurement Accuracy
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
Current Judgment Accuracy
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
Window Comparator Method
Yes
ARC Detect
Yes
RAMP (Ramp-Up Time)
0.1s fixed
TIMER (Test Time)
OFF, 1s~180s
GND
ON
DC WITHSTANDING (GPT-9602 & GPT-9603)
Output-Voltage Range
0.10kV~ 6.00kV dc
Output-Voltage Resolution
10V
Output-Voltage Accuracy
± (1.5% of setting + 2 counts) with no load
Maximum Rated Load
25W (5kV/5mA)
Maximum Rated Current
6mA (0.5kV < V ≦ 6kV)
2mA (0.1kV ≦ V≦ 0.5kV)
Voltage regulation
± (1.5% + 2 counts) [full load → no load]
Voltmeter Accuracy
± (1.5% of reading + 2 counts)
Current Measurement Range
0.01mA~6.00mA
Current Best Resolution
0.01mA
Current Measurement Accuracy
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
Current Judgment Accuracy
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
Window Comparator Method
Yes
ARC Detect
Yes
RAMP (Ramp-Up Time)
0.1s fixed
TIMER (Test Time)
OFF, 1s~180s
GND
ON
INSULATION RESISITANCE (GPT-9603 & GPT-9612)
Output-Voltage
50V, 100V, 250V, 500V, 1000V dc
Output-Voltage Accuracy
± (3.0% of setting + 1 count) [no load]
Resistance Measurement Range
1MW~ 2000MW
Test voltage
Measurable Range
Accuracy
50V/100V/250V
1~ 50MW
51~ 2000MW
±(5% of reading + 2MW)
±(10% of reading + 2MW)
500V/1000V
1~ 500MW
501~ 2000MW
±(5% of reading + 2MW)
±(10% of reading + 2MW)
Window Comparator Method
Yes
Output Impedance
600kW
RAMP (Ramp-Up Time)
0.1s fixed
TIMER (Test Time)
OFF, 1s~180s
GND
OFF (fixed)
TEST MODE*
SINGLE
ACW, DCW, IR
AUTO
AC-IR, IR-AC, DC-IR, IR-DC
INTERFACE
Remote Terminal (Front)
Standard
SIGNAL IO
Standard
General
DISPLY
240 x 48 Ice Blue dot matrix LCD
POWER SOURCE
AC 100V-120V/220V-240V ±10%, 50/60Hz
POWER COMSUMPTION
400VA max.
DIMENSIONS & WEIGHT
330(W) x 148(H) x 385(D) mm
approx.9kg(Max)
* The available “Test Mode” depends on selected model
The store will not work correctly when cookies are disabled.
We use cookies to make your experience better.To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies.Learn more.