AC WITHSTANDING |
Output-Voltage Range |
0.100kV~ 5.000kV ac |
Output-Voltage Resolution |
2V/step |
Output-Voltage Accuracy |
± (1% of setting +5V) [no load] |
Maximum Rated Load |
200 VA (5kV/40mA) |
Maximum Rated Current |
40mA (0.5kV < V ≦ 5kV)10mA (0.1kV ≦ V≦ 0.5kV) |
Output-Voltage Waveform |
Sine wave |
Output-Voltage Frequency |
50 Hz / 60 Hz selectable |
Voltage regulation |
± (1% of reading +5V) [full load → no load] |
Voltmeter Accuracy |
± (1% of reading+ 5V) |
Current Measurement Range |
0.001mA~040.0mA |
Current Best Resolution |
0.001mA / 0.01mA / 0.1mA |
AC Current Measurement Accuracy |
±(1.5% of reading +30 counts) when HISET < 1.00mA±(1.5% of reading +3 counts) when HISET ≧ 1.00mA |
Window Comparator Method |
Yes |
ARC Detect |
Yes |
RAMP (Ramp-Up Time) |
0.1~999.9s |
TIMER (Test Time)* |
OFF, 0.5s~999.9s |
GND |
ON/OFF |
DC WITHSTANDING
|
Output-Voltage Range |
0.100kV~ 6.000kV dc |
Output-Voltage Resolution |
2V/step |
Output-Voltage Accuracy |
± (1% of setting +5V) [no load] |
Maximum Rated Load |
50W (5kV/10mA) |
Maximum Rated Current |
10mA (0.5kV < V ≦ 6kV)2mA (0.1kV ≦ V≦ 0.5kV) |
Voltage regulation |
± (1% of reading +5V) [full load → no load] |
Voltmeter Accuracy |
± (1% of reading+ 5V) |
Current Measurement Range |
0.001mA~010.0mA |
Current Best Resolution |
0.001mA / 0.01mA / 0.1mA |
DC Current Measurement Accuracy |
±(1.5% of reading +30 counts) when HISET < 1.00mA±(1.5% of reading +3 counts) when HISET ≧ 1.00mA |
Window Comparator Method |
Yes |
ARC Detect |
Yes |
RAMP (Ramp-Up Time) |
0.1~999.9s |
TIMER (Test Time)* |
OFF, 0.5s~999.9s |
GND |
ON/OFF |
INSULATION RESISITANCE
|
Output-Voltage |
50V~1000V dc |
Output-Voltage Resolution |
50V/step |
Output-Voltage Accuracy |
± (1% of setting +5V) [no load] |
Resistance Measurement Range |
1MW~ 9500MW |
|
Test voltage |
Measurable Range |
Accuracy |
|
50V≦V≦450V |
1MW~ 50MW51MW~ 2000MW |
±(5% of reading +1 count)±(10% of reading +1 counts) |
|
500V≦V≦1000V |
1MW~ 500MW501MW~ 9500MW |
±(5% of reading +1 count)±(10% of reading +1 count) |
Window Comparator Method |
Yes |
Output Impedance |
600kW |
RAMP (Ramp-Up Time) |
0.1~999.9s |
TIMER (Test Time) |
1s~999.9s |
GND |
OFF (fixed) |
GROUND BOND
|
Output-Current Range |
03.00A~30.00A ac |
Output-Current Resolution |
0.01A |
Output-Current Accuracy |
3A≦I≦8A : ± (1% of setting +0.2A)8A<I≦30A : ± (1% of setting +0.05A) |
Test-Voltage |
6Vac max. (open circuit) |
Test-Voltage Frequency |
50 Hz / 60 Hz selectable |
Resistance Measurement Range |
10mW~650.0mW |
Resistance Measurement Resolution |
0.1mW |
Resistance Measurement Accuracy |
± (1% of reading +2mW) |
Window Comparator Method |
Yes |
TIMER (Test Time) |
0.5s~999.9s |
GND |
OFF (fixed) |
Test Method |
Four Terminal |
MEMORY
|
Single Step Memory |
MANU : 100 blocks |
Automatic Testing Memory |
AUTO : 100 blocks, menu per auto : 16 |
INTERFACE
|
RS232 |
Standard |
USB (Device) |
Standard |
GPIB |
Option |
Remote Terminal (Front) |
Standard |
SIGNAL IO |
Standard |
General
|
DISPLY |
240 x 64 dot matrix LED back light LCD |
POWER SOURCE |
AC100V/120V/220V/230V ±10%, 50/60Hz |
POWER CONSUMPTION |
500VA max. |
DIMENSIONS & WEIGHT |
330(W) x 148(H) x 452(D) mm (Max.),approx.19kg(Max) |