The 267 multirange low differential pressure transducer from Setra offers accuracies of ±0.25%, 0.5% and ±1% FS across an expansive pressure range of 0.1” W.C. up to 100” W.C.
Unlike other competing transducers, the 267 is built on the principle of customization, offering a myriad of electrical and mechanical options that can be applied to the standing design. 0.25” diameter pressure probe options is crafted from extruded aluminum and, in combination with baffles, prevents velocity pressure errors providing for faster and more cost effective work methods.
The physical construction of the Setra 267 makes it ideal for a variety of demanding applications. The NEMA 4 rated enclosure is built for the toughest environments. Available in both a wall and duct mount, the 267 can be installed anywhere with the wall mount and can maximize space efficiency with the duct probe configuration.
The Setra 267 uses the advanced technology of an all stainless steel capacitive sensing element. With all the sensing elements designed and manufactured by Setra, each sensor comes with the promise of upmost quality. Setra’s technology has been applied in over 8 million installations and proudly carries the highest field acceptance rate in the industry.
The combination of top quality technology, versatility and customization all in a resilient physical construction, make the Setra 267 one of the leading multirange low differential pressure transducers available.
Specifications
PerformanceData
PhysicalDescription
Standard
Optional
Case
IP65/NEMA 4 Plastic Glass-Filled
Polycarbonate UL94V-O Case
Accuracy RSS1
(at constant temp)
±1.0% FS
±0.4% FS
±0.25% FS
Electrical Connection
Screw Terminal Strip Inside of Case
Non-Linearity, BFSL
±0.98% FS
±0.38% FS
±0.22% FS
Electrical Terminations
PG-9/PG13.5 Strain Relief, 1/2” Conduit
Opening, or 9 Pin D-Sub Connector*
Hysteresis
±0.10% FS
±0.10% FS
±0.10% FS
Zero and Span Adjustments
Accessible Inside of Case
Non-Repeatability
±0.05% FS
±0.05% FS
±0.05% FS
Weight (approx.)
9.0 Ounces (255 grams)
9.5 Ounces (Duct Probe Assembly)
Position Effect: Consult factory
Electrical Data(Current)
PressureMedia
Circuit
2-Wire, Protected from Miswiring
Thermal Effects2,3
Output7
4 to 20 mA4
Compensated Range °F (°C)
+40 to +150 (+5 to +65)
Bidirectional Output at Zero
12 mA
Zero/Span Shift %FS/°F (°C)
±0.033 (±0.06)
Min. Loop Supply Voltage (VDC)
9 + 0.02 x (Resistance of Receiver plus line)
Maximum Line Pressure
10 PSI
Max. Loop Supply Voltage (VDC)
30 + 0.004 x (Resistance of Receiver plus line)
Overpressure
Up to 10 PSI (Range Dependant)
Electrical Data(Voltage)
Long-Term Stability
0.1% FS Total
Circuit
3-Wire (Exc, Gnd, Sig), Protected from Miswiring
EnvironmentalData
Excitation (for 0-5 VDC Output)
9 to 30 VAC /12 to 40 VDC
Operating6 Temperature °F (°C)
0 to +150 (-18 to +65)
Excitation (for 0-10 VDC Output)
11 to 30 VAC /13 to 40 VDC
Storage Temperature °F (°C)
-65 to +180 (-54 to +82)
Output3
0 to 5 VDC5 / 0 to 10 VDC5
1 RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
2 Units calibrated at nominal 70˚F. Maximum thermal error computed from this datum.
3 Calibrated into a 50K ohm load, operable into a 5000 ohm load or greater.
4 Zero output factory set to within ±0.16 mA (±0.08 mA for optional accuracies). Span
(Full Scale) output factory set to within ±0.16 mA (±0.08mA for optional accuracy).
5 Zero output factory set to within ±50mV (±25 mV for optional accuracies).
Span (Full Scale) output factory set to within ±50mV (±25 mV for optional accuracies)
6 Operating temperature limits of the electronics only. Pressure media temperatures may be considerably higher.
7 Calibrated at factory with a 24 VDC loop supply voltage and a 250 ohm load.
The store will not work correctly when cookies are disabled.
We use cookies to make your experience better.To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies.Learn more.