Oscilloscope mode | Vertical | Frequency response - dc coupled | Without probes and test leads (with BB120) | dc to 40 MHz (-3 dB) | With STL120-IV 1:1 shielded test leads | DC to 12.5 MHz (-3 dB) / dc to 20 MHz (-6 dB) | With VP41 10:1 Probe | dc to 40 MHz (-3 dB) | Frequency response - ac coupled (lf roll off) | Without probes and test leads | <10 Hz (-3 dB) | With STL120-IV 1:1 shielded test leads | <10 Hz (-3 dB) | With VP41 10:1 Probe | <10 Hz (-3 dB) | Rise time, excluding probes, test leads | <8.75 ns | Input impedance | Without probes and test leads | 1 MΩ//20 pF | With BB120 | 1 MΩ//24 pF | With STL120-IV 1:1 shielded test leads | 1 MΩ//230 pF | With VP41 10:1 Probe | 5 MΩ//15.5 pF | Sensitivity | 5 mV to 200 V/div | Analog bandwidth limiter | 10 kHz | Display modes | A, -A, B, -B | Max. input voltage A and B | Direct, with test leads, or with VP41 Probe | 600 Vrms Cat IV, 750 Vrms maximum voltage. | With BB120 | 600 Vrms | Max. floating voltage, from any terminal to ground | 600 Vrms Cat IV, 750 Vrms up to 400Hz | Horizontal | Scope modes | Normal, Single, Roll | Ranges (normal) | Equivalent sampling | 10 ns to 500 ns/div | Real time sampling | 1 μs to 5 s/div | Single (real time) | 1 μs to 5 s/div | Roll (real time) | 1s to 60 s/div | Sampling rate (for both channels simultaneously) | Equivalent sampling (repetitive signals) | Up to 4 GS/s | Real time sampling 1 μs to 60 s/div | 40 MS/s |
Trigger | Screen update | Free run, on trigger | Source | A, B | Sensitivity A and B | @ DC to 5 MHz | 0.5 divisions or 5 mV | @ 40 MHz | 1.5 divisions | @ 60 MHz | 4 divisions | Slope | Positive, negative |
Advanced scope functions | Display modes | Normal | Captures up to 25 ns glitches and displays analog-like persistence waveform | Smooth | Suppresses noise from a waveform | Glitch off | Does not capture glitches between samples | Envelope | Records and displays the minimum and maximum of waveforms over time | Auto set (Connect-and-View™) | Continuous fully automatic adjustments of amplitude, time base, trigger levels, trigger gap, and hold-off. Manual override by user adjustment of amplitude, time base, or trigger level. |
Dual input meter | The accuracy of all measurements is within ±(% of reading + number of counts) from 18 °C to 28 °C. | Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe uncertainty +1%. More than one waveform period must be visible on the screen. |
Input A and input B | DC voltage (VDC) | Ranges | 500 mV, 5 V, 50 V, 500 V, 750 V | Accuracy | ±(0.5% +5 counts) | Common mode rejection (CMRR) | >100 dB @ dc, >60 dB @ 50, 60, or 400 Hz | Full scale reading | 5000 counts | True-rms voltages (V ac and V ac+dc) | Ranges | 500 mV, 5 V, 50 V, 500 V, 750 V | Accuracy for 5% to 100% of range (DC coupled) | DC to 60 Hz (V ac+dc) | ±(1% +10 counts) | 1 Hz to 60 Hz (V ac) | ±(1% +10 counts) | Accuracy for 5% to 100% of range (AC or dc coupled) | 60 Hz to 20 kHz | ±(2.5% +15 counts) | DC rejection (only VAC) | >50 dB | Common mode rejection (CMRR) | >100 dB @ dc | >60 dB @ 50, 60, or 400 Hz | Full scale reading | 5000 counts, reading is independent of any signal crest factor. | Peak | Modes | Max peak, Min peak, or pk-to-pk | Ranges | 500 mV, 5 V, 50 V, 500 V, 2200 V | Accuracy | Accuracy Max peak or Min peak | 5% of full scale | Accuracy Peak-to-Peak | 10% of full scale | Full scale reading | 500 counts | Frequency (Hz) | Ranges | 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz, 10 MHz, and 70 MHz | Frequency range | 15 Hz (1 Hz) to 50 MHz in continuous autoset | Accuracy @1 Hz to 1 MHz | ±(0.5% +2 counts) | Full scale reading | 10,000 counts | RPM | Max reading | 50.00 kRPM | Accuracy | ±(0.5% +2 counts) | Duty cycle (PULSE) | Range | 2% to 98% | Frequency range | 15 Hz (1 Hz) to 30 MHz in continuous autoset | Pulse width (PULSE) | Frequency range | 15 Hz (1 Hz) to 30 MHz in continuous autoset | Full scale reading | 1000 counts | Amperes (AMP) | With current clamp | Ranges | Same as V dc, V ac, V ac+dc, or PEAK | Scale factors | 0.1 mV/A, 1 mV/A, 10 mV/A, 100 mV/A, 400 mV/A, 1 V/A, 10 mV/mA | Accuracy | Same as V dc, V ac, V ac+dc, or PEAK (add current clamp uncertainty) | Temperature (TEMP) with optional temperature probe | Range | 200 °C/div (200 °F/div) | Scale factor | 1 mV/°C and 1 mV/°F | Accuracy | As V dc (add temp. probe uncertainty) | Decibel (dB) | 0 dBV | 1 V | 0 dBm (600 Ω /50 Ω) | 1 mW referenced to 600 Ω or 50 Ω | dB on | V dc, V ac, or Vac+dc | Full scale reading | 1000 counts | Crest factor (CREST) | Range | 1 to 10 | Full scale reading | 90 Counts | Phase | Modes | A to B, B to A | Range | 0 to 359 degrees | Resolution | 1 degree | Power | Configurations | 1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET mode only) | Power factor (PF) | Ratio between watts and VA range - 0.00 to 1.00 | Watt | RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes) | Full scale reading | 999 counts | VA | Vrms x Arms | Full scale reading | 999 counts | VA reactive (var) | √((VA)2-W2) | Full scale reading | 999 counts | Vpwm | Purpose | To measure on pulse width modulated signals, like motor drive inverter outputs | Principle | Readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency | Accuracy | As Vrms for sinewave signals |
Input A to common | Ohm (Ω) | Ranges | | 50 Ω, 500 Ω , 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ | Accuracy | ±(0.6% + 5 counts) 50 Ω ±(2% + 20 counts) | Full scale reading | 50 Ω to 5 MΩ - 5000 counts, 30 MΩ - 3000 counts | Measurement current | 0.5 mA to 50 nA, decreases with increasing ranges | Open circuit voltage | <4 V | Continuity (Cont) | Beep | <(30 Ω ±5 Ω) in 50 Ω range | Measurement current | 0.5 mA | Detection of shorts of | ≥1 ms | Diode | Measurement voltage | @0.5 mA | >2.8 V | @open circuit | <4 V | Measurement current | 0.5 mA | Polarity | + on input A, - on COM | Capacitance (CAP) | Ranges | 50 nF, 500 nF, 5 μF, 50 μF, 500 μF | Full scale reading | 5000 counts | Measurement current | 500 nA to 0.5 mA, increases with increasing ranges |
Advanced meter functions | Zero Set | Set actual value to reference | AutoHold (on A) | Captures and freezes a stable measurement result. Beeps when stable. AutoHold works on the main meter reading, with thresholds of 1 Vpp for AC signals and 100 mV for DC signals. | Fixed decimal point | Activated by using attenuation keys |
Cursor Readout | Sources | A, B | Single vertical line | Average, min and max readout | Average, min, max and time from start of readout (in ROLL mode; instrument in HOLD) | Min, max and time from start of readout (in RECORDER mode; instrument in HOLD) | Harmonics values in POWER QUALITY mode. | Dual vertical lines | Peak-peak, time distance and reciprocal time distance readout | Average, min, max and time distance readout (in ROLL mode; instrument in HOLD) | Dual horizontal lines | High, low and peak-peak readout | Rise or fall time | Transition time, 0%-level and 100%-level readout (manual or auto leveling; auto leveling only possible in single channel mode) | Accuracy | As oscilloscope accuracy |
Recorder | The recorder captures meter readings in Meter Recorder mode or continuously captures waveform samples in Scope Recorder mode. The information is stored on internal memory or on optional SD card | The results are displayed as Chart recorder display that plots a graph of min and max values of Meter measurements over time or as a waveform recorder display that plots all the captured samples. | Meter readings | Measurement Speed | Maximum 2 measurement/s | Record Size (min, max, average) | 2 M readings for 1 channel | Recorded Time Span | 2 weeks | Maximum number of events | 1024 | Waveform record | Maximum sample rate | 400 K sample/s | Size Internal memory | 400 M samples Recorded Time | Span internal memory | 15 minutes at 500 μs/div 11 hours at 20 ms/div | Record Size SD card | 1.5 G samples | Recorded Time Span SD card | 11 hours at 500 μs/div 14 days at 20 ms/div | Maximum number of events | 64 |
Power Quality | Readings | Watt, VA, var, PF, DPF, Hz | Watt, VA, var ranges (auto) | 250 W to 250 MW, 625 MW, 1.56 GW | When selected: total (% r) | ±(2% + 6 counts) | When selected: fundamental (% f) | ±(4% + 4 counts) | DPF | 0.00 to 1.00 | PF | 0.00 to 1.00, ±0.04 | Frequency range | 10.0 Hz to 15.0 kHz 40.0 Hz to 70.0 Hz | Number of Harmonics | DC to 51 | Readings / Cursor readings (fundamental 40 Hz to 70 Hz) | Vrms / Arms /Watt | Each harmonic from fundamental maybe selected for individual readings | Includes frequency of fundamental, phase Angle and K-factor (in Amp and Watt) |
BusHealth tester | Type | Subtype | Protocol | AS-i | NEN-EN50295 | CAN | ISO-11898 | Interbus S | RS-422 | EIA-422 | Modbus | RS-232 | RS-232/EIA-232 | RS-485 | RS-485/EIA-485 | Foundation Fieldbus | H1 | 61158 type 1, 31.25 kBit | Profibus | DP | EIA-485 | PA | 61158 type 1 |
Miscellaneous | Display | Type | 5.7-inch color active matrix TFT | Resolution | 640 x 480 pixels | Waveform Display | Vertical | 10 div of 40 pixels | Horizontal | 12 div of 40 pixels | Power | External | Via Power Adapter BC430 | Input voltage | 10 V DC to 21 V DC | Power consumption | 5 W typical | Input connector | 5 mm jack | Internal | Via Battery Pack BP290 | Battery power | Rechargeable Li-Ion 10.8 V | Operating time | 7 hours with 50% backlight brightness | Charging time | 4 hours with test tool off, 7 hours with test tool on | Allowable ambient temp | 0 to 40 °C (32 to 104 °F) during charging | Memory | Internal memory can store 20 data sets (screen waveform and setup) | Micro SD card slot with optional SD card (max size of 32 GB) | Mechanical | Size | 259 x 132 x 55 mm (10.2 x 5.2 x 2.15 in) | Weight | 1.4 kg (3.2 lb) including battery pack | Interface | Optically isolated | Transfer screen copies (bitmaps), settings and data | USB to PC/laptop | OC4USB optically isolated USB adapter/cable, (optional), using FlukeView® software for Windows®. | Optional WiFi adapter | Fast transfer of screen copies (bitmaps), settings and data to PC/laptop, tablet, smartphone, etc. A USB port is provided for attaching the WiFi dongle. Do not use the USB port with a cable for safety reasons. |
Environmental | Environmental | MIL-PRF-28800F, Class 2 | Temperature | Battery Operation | 0 to 40 °C (32 to 104 °F) | Power Adapter Operation | 0 to 50 °C (32 to 122 °F) | Storage | -20 to 60 °C (-4 to 140 °F) | Humidity (Operating) | @ 0 to 10 °C (32 to 50 °F) | Non-condensing | @ 10 to 30 °C (50 to 86 °F) | 95% | @ 30 to 40 °C (86 to 104 °F) | 75% | @ 40 to 50 °C (104 to 122 °F) | 45% | Storage | @ -20 to 60 °C (-4 to 140 °F) | Non-condensing | Altitude | Operating at 3 km (10,000 feet) | CAT III 600 V | Operating at 2 km (6,600 feet) | CAT IV 600 V | Storage | 12 km (40,000 feet) | EMC electromagnetic compatibility | International | IEC 61326-1: Industrial, CISPR 11: Group 1, Class A | Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) | USA (FCC) | 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. | Wireless radio with adapter | Frequency range | 2412 MHz to 2462 MHz | Output power | <100 mW | Enclosure protection | IP51, ref: EN/IEC60529 | Safety | General | IEC 61010-1: Pollution Degree 2 | Measurement | IEC 61010-2-033: CAT IV 600 V/CAT III 750 V | Max. input voltage input A and B | Direct on input or with leads | 600 Vrms CAT IV for derating | With Banana-to-BNC Adapter BB120 | 600 Vrms for derating | Max. floating voltage from any terminal to ground | 600 Vrms Cat IV, 750 Vrms up to 400 Hz |
|