The HOBO Water Level Titanium is recommended for saltwater deployment for recording water levels and temperatures in wetlands and tidal areas. This data logger features high accuracy at a great price and HOBO ease-of-use, with no cumbersome vent tubes or desiccants to maintain.
Specifications
Pressure and Water Level Measurements U20-001-02 and U20-001-02-Ti
Operation Range
0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
Burst Pressure
500 kPa (72.5 psia) or 40.8 m (134 ft) depth
Water Level Accuracy*
Typical error: ±0.05% FS, 1.5 cm (0.05 ft) water Maximum error: ±0.1% FS, 3 cm (0.1 ft) water
Raw Pressure Accuracy**
±0.3% FS, 1.20 kPa (0.17 psi) maximum error
Resolution
<0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water
Pressure Response Time (90%)***
<1 second; measurement accuracy also depends on temperature response time
Temperature Measurements (All Models)
Operation Range
-20° to 50°C (-4° to 122°F)
Accuracy
±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F), see Plot A
Resolution
0.10°C at 25°C (0.18°F at 77°F), see Plot A
Response Time (90%)
5 minutes in water (typical)
Stability (Drift)
0.1°C (0.18°F) per year
Logger
Real-time Clock
± 1 minute per month 0° to 50°C (32° to 122°F)
Battery
2/3 AA, 3.6 Volt lithium, factory-replaceable
Battery Life (Typical Use)
5 years with 1 minute or greater logging interval
Memory (Non-volatile)
64K bytes memory (approx. 21,700 pressure and temperature samples)
Weight
Stainless steel models: approximately 210 g (7.4 oz) Titanium models: approximately 140 g (4.8 oz)
Dimensions
2.46 cm (0.97 inches) diameter, 15 cm (5.9 inches) length; mounting hole 6.3 mm (0.25 inches) diameter
Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOware software manual.
Launch Modes
Immediate start and delayed start
Offload Modes
Offload while logging; stop and offload
Battery Indication
Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile.
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
* Water Level Accuracy: With accurate reference water level measurement, known water density, accurate Barometric Compensation Assistant data, and a stable temperature environment.
** Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.
*** Changes in Temperature: Allow 10 minutes in water to achieve full temperature compensation of the pressure sensor. Maximum error due to rapid thermal changes is approximately 0.5%.
The store will not work correctly when cookies are disabled.
We use cookies to make your experience better.To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies.Learn more.