Specifications : APAQ-HRF/-HRFX Input | Pt100 (α = 0.00385), 3-wire connection | Adjustable to specific ranges within: -50 to +550 °C / -60 to +1120 °F | Sensor current | ~1.1 mA | Max. sensor wire resistance | 15 Ω / wire | Monitoring | Sensor break detection, selectable | Upscale ~25 mA, downscale ~3 mA | Adjustments | Zero | -50 to +50 °C / -60 to +120 °F | Span, selectable | 50 to 500 °C / 100 to 1000 °F | Span, fine adjustment | ±10 % (± 5% for 600 / 800 / 1000 °F) | Output | Current | 4 - 20 mA | Linearity | Temperature linear | Current limitation | ~ 25 mA | Permissible load | APAQ-HRF | 700 Ω @ 24 VDC, 25 mA | | APAQ-HRFX | 620 Ω @ 24 VDC, 25 mA | Temperature | Ambient, storage | -40 to +100 °C / -40 to +212°F | Ambient, operating | APAQ-HRF | -40 to +85 °C / -40 to +185 °F | | APAQ-HRFX | ATEX:T4 /+85 °C, T5 /+55 °C, T6 /+40 °C; FM: T4/+80 °C | General data | Response time 10-90% | ≤ 0.2 s | Humidity (non-condensing) | 0 to 95 %RH | Intrinsic safety | APAQ-HRFX | ATEX: II 1 G EEx ia IIB T4, T5, T6 FM: Class I, Div.1, Group A-D | Power supply, polarity protected | Supply voltage | APAQ-HRF | 6.5 to 32 VDC | | APAQ-HRFX | 8.5 to 30 VDC | Permissible ripple | 4 Vp-p @ 50/60 Hz | Accuracy | Linearity | ±0.1 % of span | Calibration | ±0.1 % of span | Temperature influence | ±0.6 % of span/25 °C, ±0.7 % of span/50 °F | Sensor wire influence | ±0.005 °C/Ω / ±0.009 °F/Ω 1) | RFI influence,0.15-1000MHz, 10 V or V/m | ±0.2 % of span (typical) | Supply voltage influence | ±0.02 % of span/V | Supply ripple influence, 50/60 Hz, 4 Vp-p | ±0.05 % of span | Long term stability | ±0.1 % of span/year | Housing | Material / Flammability(UL) | Zinc alloy + ABS / V0 | Mounting | DIN B-head or larger | Connection, single/stranded wires | ≤2.5 mm2, AWG 14 | Weight | 40 g | Protection, housing with cover/terminals | IP 20 / IP 10 |
1) Per wire, with equal resistance Specifications : APAQ-HCF/-HCFX Input | Thermocouples | Selectable, type J, L, T, K and N with ranges within -5 to +55 mV | Input impedance | >5 MΩ | Max. sensor wire resistance | 500 Ω (total loop) | Monitoring | Sensor break detection, selectable | Upscale ~25 mA, downscale ~3 mA | Adjustments | Zero | ±10 % of span | Span, selectable | 10 to 50 mV | Span, fine adjustment | ±10 % | Output | Current | 4 - 20 mA | Linearity | Voltage linear | Current limitation | ~ 25 mA | Permissible load | APAQ-HCF | 700 Ω @ 24 VDC, 25 mA | | APAQ-HCFX | 620 Ω @ 24 VDC, 25 mA | Temperature | Ambient, storage | -40 to +100 °C / -40 to +212°F | Ambient, operating | APAQ-HCF | -40 to +85 °C / -40 to +185 °F | | APAQ-HCFX | ATEX:T4 /+85 °C, T5 /+55 °C, T6 /+40 °C; FM: T4/+80 °C | General data | Response time 10-90% | ≤ 0.2 s | Humidity (non-condensing) | 0 to 95 %RH | Intrinsic safety | APAQ-HCFX | ATEX: II 1 G Ex ia IIB T4, T5, T6 FM: Class I, Div.1, Group A-D | Power supply, polarity protected | Supply voltage | APAQ-HCF | 6.5 to 32 VDC | | APAQ-HCFX | 8.5 to 30 VDC | Permissible ripple | 4 Vp-p @ 50/60 Hz | Accuracy | Linearity (mA output to mV input) | ±0.1 % of mV span | Calibration | ±0.1 % of span | Cold Junction Compensation (CJC) | ±1.0 °C /±1.8 °F | Temperature influence | ±0.6 % of span/25 °C, ±0.7 % of span/50 °F | Temperature influence CJC | ±1.25 °C/25 °C, ±2.5 °F/50 °F 1) | Sensor wire influence | 0.4 μV/Ω | RFI influence,0.15-1000MHz, 10 V or V/m | ±0.2 % of span (typical) | Supply voltage influence | ±0.02 % of span/V | Supply ripple influence, 50/60 Hz, 4 Vp-p | ±0.05 % of span | Long term stability | ±0.1 % of span/year | Housing | Material / Flammability(UL) | Zinc alloy + ABS / V0 | Mounting | DIN B-head or larger | Connection, single/stranded wires | ≤2.5 mm2, AWG 14 | Weight | 40 g | Protection, housing with cover/terminals | IP 20 / IP 10 |
1) ±2.5 °C/25 °C, ±5.0 °F/50 °F for type T |