The 830C and 890C are 11,000-count handheld capacitance meters that measure capacitance up to 200 mF (50 mF for 890C). Straightforward to operate, these meters make measurements quickly and simplify the capacitance sorting process.Unlike other handheld instruments such as LCR meters that make measurements using an AC signal with a specific test frequency, they measures capacitance by applying a constant current that briefly charges, then discharges the connected capacitor.These meters include many handy functions. The compare mode which supports 25 sets of High/Low limits, provides a convenient means to sort capacitors without needing to set up the meter each time. Data logging, Hold/Min/Max Average recording, tolerance, and relative mode features offer additional component troubleshooting tools.
Specifications
Electricalspecifications
Accuracy is given as (% of reading + counts of least significant digit) at 23°C ± 5°C, with relative humidity less than 80%.
Capacitance
Range
Resolution
Accuracy*
Measuring rate as full scale
(approx.)
1000 pF
0.1 pF
1% + 10
5 times /sec
10 nF
0.001 nF
1% + 5
5 times /sec
100 nF
0.01 nF
0.5% + 3
5 times /sec
1000 nF
0.1 nF
5 times /sec
10 µF
0.001 µF
5 times /sec
100 µF
0.01 µF
5 times /sec
1000 µF
0.1 µF
0.86 times /sec
10 mF
0.001 mF
1% + 5
0.13 times /sec
50 mF (890C)
0.01 mF
2% + 5
0.026 times /sec
199.99 mF (830C)
0.006 times /sec
* The accuracy is specified to measure film capacitor or better, and use relative mode to zero residual first.
General
Model
830C, 890C
Display
41/2-digit liquid crystal display (LCD) with maximum reading of 11000 counts
automatic polarity indication
Measuring rate (approx.)
5 times /sec for capacitance <100ºF
Low battery indicator
" " appears when the battery voltage drops below 6V (approx.)
Operating temperature
Full accuracy at 0°C to 40°C (32ºF~113ºF)
Storage temperature
-20°C to 50°C (-4ºF~122ºF) without battery
Operating humidity (R.H.)
Maximum 80% R.H. for temperature up to 31°C decreasing linearly to 50% R.H. at 40°C
Temperature coefficient
0.1 * (specified accuracy) / °C (from 0°C to 18°C or 28°C to 40°C)
Power supply
Single standard 9V battery or AC power adapter
Power consumption
6.5 mA for battery operation, 2 uA standby current (power is off)
Battery life
70 hours without backlight (approx.) (Based on new alkaline 9V battery)
Standard Accessories: Alligator clip leads, 9V battery, mini USB cable, and instruction manual. Additionally, an AC power adapter is included with the 830C, optional for model 890C, 830C EXD ships with 230 VAC adapter.