The Rotronic HC2-SH probe has an accuracy of ±0.5 %rh, ±0.1 K, at 23 °C ±5 K.
Fast response • By decoupling the Pt100 temperature sensor from the probe, thermal response is significantly improved
Innovative filter technology • The latest in filter technology provides a high level of protection and maintains the optimum measurement environment for the Hygromer humidity and Pt100 temperature sensors
AirChip3000 • Compensates humidity and temperature over 30,000 reference points • 2,000 data point memory • Dew/frost point calculation • Sensor auto-diagnostics and automatic correction • Active alarming and information • ASIC (Application Specifi c Integrated Circuit), micro-controller and EEPROM on one chip
Maximum flexibility • The UART 1 interface and user scalable analog outputs allow the probe to be used as a stand-alone device that can also be effortlessly integrated into an OEM application • The probes can be interchanged without adjustment 1 Universal Asynchronous Receiver Transmitter
Specifications
Probe type
Standard probe
Dimensions
Ø15 x 83 mm
Weight
10 g
Operating range electronics
-50…100 °C / 0…100 %rh
Accuracy at 23 ±5 °C
±0.5 %rh / ±0.1 K
Accuracy w. adjustment profile "High Precision"
at 23 °C / 10, 20, 30, 40, 50, 60, 70, 80, 90 %rh ± 0.5 %rh / 0.1 K
Power supply / Power consumption
3.2…5 VDC ±0 % / < 4.5 mA
Long-term stability sensor
<1 %rh / year
Humidity sensor
Hygromer IN-1
Temperature sensor
PT100 1/3 DIN Class B
Probe protection
Polycarbonate plastic cage, black
Filter cartridge
Polyethylene dust filter, gray, 20 um
Response time t 63
<15 sec.
Allowed wind speed
20 m/s, with filtre
Psychrometric calculations
Dew or frost point
Reproducibility Airchip
<0.02 %rh / 0.01 K
Exploitation software
ROTRONIC HW4.2
Probe adjustment by software
1 Point & multipoint %rh & °C
Firmware update
With internet and HW4 software
User information
from Software HW4 (devicemanager)
Data processing via HW4
with interface cables
Device Protection by password
Yes
Type of output signals
2 x 0…1 VDC
Analogue outputs scaleable by the user
Yes
Analog output signal (standard)
0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital output signal (standard)
UART
Service Interface
UART
Reverse voltage protection
Mechanically in the connector
Max. cable length
5 m
Enclosure material
Polycarbonate
FDA/GAMP compatibility
FDA 21 CFR Part 11 and GAMP compatible
Dimensions & Connections