The Kanomax tracer gas diffuser is for use in performing the tracer gas test in accordance with ANSI/ASHRAE Standard 110-1995. The diffuser is made to the design and specifications of Standard drawing #110-83M.
The diffuser is placed in the fume hood and sulfur hexafloride gas is injected at at supply pressure of 30 psig. The internal critical orifice ensures a flow rate of 4 liters per minute. Other orifice sizes can be inserted.
Specifications
Model
Dif-Kit
Critical Orifice
4 or 8 Liter pre Minute (4 lpm Standard)
Regulator
Dual Stage, Specialty Gas
Flow Meter
Calibrated 150mm/200psi, Glass Tube
Pressure Gage
Large Dial, 0 to 60psi
Standard Accessories
Diffuser, Pressure Gauge, Shut-off Valve, In-line Flow Meter 25 ft of Tubing, Tank Regulator for Gas Cylinder
Option Accessories
Rebuild Kit (includes new Critical Orifice, Washers, Bronze Sinter Filter) 8 lpm Swappable Ejector Base
Model
Test-Manikin
Standard Accessories
Lab Coat and Stand, Breathing Zone Aspirator and Tubing to Leak Meter
Vapor Analyzer Specifications
Type
Single Beam Infrared Spectrophotometer
Sample Flow
10 liter per min (21.2 ft3/hr)
Sample Cell Volume
0.45 liters (0.0159 ft3)
Gas Purge
5 Cell Volumes for 99% purge of cell
Concentration Alarm
High & Low (selectable by user)
Interface
USB, RS232
Power Supply
NiMH battery or 120-220V AC
Dimensions
15 x 7.3 x 7.5 inches (381 x 185 x 191 mm)
Weight
18 lbs (8.2 kg)
Fog Machine Specifications
Particle Size
0.25 to 60 microns
Max Consumption
1 liter per hour
Power Requirements
120 volts, 50/60 Hz
Dimensions
14.3” x 5.6” x 7.5”