Martel BETA DPC300A is with full digital precision and accuracy, a dual pressure system and a built-in loop calibrator function, it’s a pressure calibration powerhouse in a portable format. The internal pump goes up to 300 psi with push-button ease. Exact values are set using the fine vernier control knob. Plant air, up to 100 psi is controlled with a precision regulator for low inches of water on up. DPC300A has a built-in loop calibrator with 24 VDC instrument loop power supply; high capacity rechargeable NiMH batteries for 50 hours of use or 125 cycles of pumping to 300 psi, Martel’s trademark 3 key function menus, ClearBrite™ LCD display and multimode operation to name a few. This high performance calibrator comes in a high performance housing. It’s a rugged IP67 rated blown resin case that can take a pounding when used in the field. IP67 means waterproof including immersion when the case is closed.
Specifications
Instrument Setup Recall
5 setups; last used setup displays on power-up
Environmental
Operating Temperature
-10 °C to +50 °C
Storage Temperature
-20 °C to +60 °C
Battery Type
9 cell NiMH rechargeable; 3700 mAh @ 10.8V
Battery Life (fully-charged)
Approx. 50 hours (only measurement or with external pressure supply)
125 pump cycles to 20 bar
300 pump cycles to 10 bar
1,000 pump cycles to 2 bar
Physical
Dimensions
15.25" x 12” x 7" (387.4 x 304.8 x 177.8 mm)
Weight
15.5 lbs. (approx. 7 kg)
EMI/RFI Conformance
EN50082-1: 1992 and EN55022: 1994 Class B
Safety
EN/IEC 61010-1:2010 3rd Edition
(Low Voltage Directive)
Connectors/Ports
1/8” NPT; BetaPort-P pressure module jack; power jack
Included Accessories
Reference manual, NIST-traceable calibration certificate, test leads and hose kit.
Ranges
Pressure
-12 to 300.00 psi / -0.8 to 20 bar / -80 to 2000 kPa
mA
0 to 24.000 mA
Volts
0 to 30.000 VDC
Engineering Units
psi, bar, mbar, kPa, MPa, kgcm2,
mmH2O @ 4°C, mmH2O @ 20°C,
cmH2O @ 4°C, cmH2O @ 20°C,
inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F,
mmHg @ 0°C, inHg @ 0°C
Accuracy
Pressure
6 months: ±0.025% of reading ±0.01% of full scale
1 year: ±0.035% of reading ±0.01% of full scale
mA
±0.015% of reading ±0.002mA
Volts
±0.015% of reading ±0.002V
Temperature Effect (all functions)
No effect on accuracy on all functions from 15°C to 35°C
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C