Kanomax Model 3521 is a unique respirable aerosol monitor, providing direct mass concentration of particulates using Piezobalance technology. Impactor is 10 µm, 4 µm.
Principle
As air sample enters the system, it travels through the impactor, which captures and removes larger particulates away from the sample. Smaller particulates become electrically charged and deposited on the piezo-crystal. The total mass of the deposited particulates affects the piezo-crystal’s frequency. Since the change in frequency is proportional to the mass of the particulates, the actual weight of the particulates is obtained.
Specifications
Model
3521
3522
Measuring Object
Airborn Particle Matter < 10 µm
Airborn Particle Matter < 2.5 µm
Impactor
10 µm, 4µm
2.5 µm
Measuring Range
0.01 to 10 mg/m3
Sampling Flow Rate
1L/min
Accuracy
+/- 10% of reading +/- 1 digit (0 to 5 mg/m3) +/- 20% of reading +/- 1 digit (5 to 10 mg/m3)
Data Logging
500 data and time stamped measurements
Digital Output
RS-232C
Power Supply
AC or Ni-MH battery
Dimensions
H7.0” x D5.9” x W2.7” (180 x 150 x 65mm)
Weight (including batteries)
4.4 lbs. (2 kg)