We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Airy Technology H303 is the industry’s highest concentration of 15,000,000 particles/ft³ @ 10% coincidence loss and measures 0.3 μm to 25.0 μm. The H303 shows up to 3 user-selectable size channels as well as temperature and relative humidity. Users can analyze data on screen or via printer, real time through its versatile output options, or export to Instrument Management Software (IMS).
Models | H-303 | H-503 |
Size Range | 0.3 to 25 μm | 0.5 to 25 μm |
Size Channels | Factory calibrated at 0.3, 0.5, 1.0 μm | Factory calibrated at 0.5, 0.7, 1.0 μm |
Counting Efficiency | 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS | 50% @ 0.5 μm; 100% for particles >0.75 μm per JIS |
Flow rates | 0.1 CFM (2.83 LPM) | |
Concentration Limit | 15,000,000 Particles/ft³ @ 10% coincidence loss | |
Light Source | Long life laser diode | |
Battery Run Time | >10 hours continuous operation | |
Zero Count | <1 count / 5 minutes (<2 particles / ft³) (per ISO 21501-4 & JIS) | |
Count Alarms | 1 to 9,999,999 counts | |
Calibration | NIST traceable | |
Display | 4.3" (10.9 cm) WQVGA (480x272) color touch screen | |
Printer (Optional) | External thermal printer | |
Vacuum Source | Internal pump with automatic flow control | |
Filtered Exhaust | Internal HEPA filter | |
Number of Channels | 3 | |
Custom Size Channels | Calibration for custom size channels available | |
Audible Alarm | Adjustable built-in alarm | |
Battery | Removable Li-ion | |
Battery Recharge Time | 4 hours within instrument (<2 hours with external battery charger) | |
Communication Modes | USB | |
Optional Communication Mode | Ethernet, Wireless 802.11 b/g, RS485 or RS232 | |
Alarm | Alarms on counts for all particle sizes, low battery, sensor failure, environmental sensors and flow | |
Standards | ISO 21501-4 and JIS B9921 | |
Calibration | Recommended minimum once per year | |
External Surface | High impact injection molded plastic | |
Dimensions (L x W x H) | 10” x 5” x 4.5” (25.4 cm x 12.9 cm x 11.4 cm) includes handle and does not include probes | |
Weight | 2.2 lb (1.0 kg) | |
Accessories | Operating manual on USB flash drive, isokinetic probe, purge filter, battery, data download software, USB cable, power supply & cable | |
Optional Accessories | Printed manual, carrying case, spare battery, external battery charger, external printer, isokinetic probes, and temperature relative humidity sensor | |
Buffer Memory | 45,000 sample records (rotating buffer) including particle count data, environmental data, locations and times. Scrollable on screen or printout | |
Sample Locations | Up to 1,000 locations 20 characters long | |
Sample Time | 1 second to 99 hours | |
Power | 110 to 240 VAC 50/60 Hz universal in-line power supply | |
Operating Conditions | 41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing | |
Storage Conditions | 32º to 122ºF (0º to 50ºC) / Up to 98% non-condensing | |
Warranty | 1 Year. Extended warranties available |