MadgeTech PRTC110 Differential Pressure and Thermocouple Temperature Data Logger - DISCONTINUED

SKU
PRTC110
Manufacturer:
MadgeTech
MadgeTech

Key Features:

This Product is DISCONTINUED

  • 5 year battery life
  • Wide temperature range
  • Accepts thermocouple types
    J, K, T, E, R, S, B, N
  • High speed download
  • Real time operation
  • Low cost
  • Programmable start time
  • Miniature size
Call For Price

Secured and trusted checkout with

MadgeTech’s PRTC110 Differential Pressure and Thermocouple Temperature Data Logger is available in a 0 to 2 inH2O or 0 to 5 inH2O model, suiting a variety of factory and laboratory applications.

 

Capable of measuring and recording data for up 10,922 measurements per channels, the PRTC110 has a non-volatile solid state memory storage medium that offers the utmost data security, even if the battery becomes discharged. The real time clock module lengthens the battery life for up to 5 years and facilitates high speed downloads.

 

The PRTC110 can be started and stopped directly from a PC and its compact design allows for portability and user flexibility. The data logger can be plugged into an empty COM or USB port and, along with the easy to use software, data retrieval becomes virtually effortless.

 

*Thermocouple probe sold separately.

TEMPERATURE

Remote Channel Temperature SensorThermocouple Types J, K, T, E, R, S, B, or N
Remote Channel Temperature RangeThermocouple Dependent
Remote Channel Temperature ResolutionThermocouple Dependent
Remote Channel Temperature Calibrated AccuracyThermocouple Dependent

PRESSURE

Pressure SensorSemiconductor
Pressure Range0 to 2 inH2O or 0 to 5 inH2O (differential)
Pressure Resolution0.002 inH2O and 0.005 inH2O
Pressure Calibrated Accuracy0.5 %FSR, 0.25 % at 25 °C (77 °F) typical

GENERAL

Memory10,922 readings per channel
Reading Rate1 reading every 2 seconds to 1 reading every 24 hours
Baud Rate57,600
Cold Junction CompensationAutomatic
LED IndicatorsRed
MaterialABS Plastic
Dimensions2.7 in x 1.6 in x 0.8 in (68 mm x 41 mm x 20 mm) plus connective tubing
Operating Environment-20 °C to +80 °C (-4 °F to +176 °F), 0 %RH to 95 %RH non-condensing
ApprovalsCE

 

ThermocoupleRange (°C)ResolutionAccuracy
J-210 to +7600.1 °C±.5 °C
K-270 to +13700.1 °C±.5 °C
T-270 to +4000.1 °C±.5 °C
E-270 to +9800.1 °C±.5 °C
R-50 to +17600.5 °C±2 °C
S-50 to +17600.5 °C±2 °C
B+50 to +18200.5 °C±2 °C
N-270 to +13000.1 °C±.5 °C
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account
  • R&D
  • Diagnostics
  • Measure Pressure and Temperature Pipelines
  • Clean Rooms
  • Atmospheric pressure gradient monitoring