The Kanomax tracer gas diffuser is for use in performing the tracer gas test in accordance with ANSI/ASHRAE Standard 110-1995. The diffuser is made to the design and specifications of Standard drawing #110-83M.
The diffuser is placed in the fume hood and sulfur hexafloride gas is injected at at supply pressure of 30 psig. The internal critical orifice ensures a flow rate of 4 liters per minute. Other orifice sizes can be inserted.
|
||||||||||||||
|
Vapor Analyzer Specifications |
|
Type |
Single Beam Infrared Spectrophotometer |
Sample Flow |
10 liter per min (21.2 ft3/hr) |
Sample Cell Volume |
0.45 liters (0.0159 ft3) |
Gas Purge |
5 Cell Volumes for 99% purge of cell |
Concentration Alarm |
High & Low (selectable by user) |
Interface |
USB, RS232 |
Power Supply |
NiMH battery or 120-220V AC |
Dimensions |
15 x 7.3 x 7.5 inches (381 x 185 x 191 mm) |
Weight |
18 lbs (8.2 kg) |
Fog Machine Specifications |
|
Particle Size |
0.25 to 60 microns |
Max Consumption |
1 liter per hour |
Power Requirements |
120 volts, 50/60 Hz |
Dimensions |
14.3” x 5.6” x 7.5” |