The INOR IPAQ CT20 is a compact, analog, non-isolated, 2-wire temperature transmitter that is for the Pt100 sensors.
Input | 3-wire Pt100 IEC 60751, @=0.00385 |
Sensor current | ~ 0.5 mA |
Max. resistance/wire | 20 Ω/wire |
Output | |
Analog | 4…20 mA, temperature linear |
Response time (90 %) | < 10 ms |
Permissible load | 750 Ω @ 24 VDC |
Range configuration | |
Measuring range | PC-configurable |
Measuring range limits | -50 to +800°C |
Zero limits | -50 to +50°C |
Minimum span | 50°C or 20°C, If the zero point is set at -40 °C, -20 °C, 0 °C, +20 °C or +40°C the minimum span is 20 °C, otherwise it is 50 °C |
Sensor error compensation | Max. ±1 % of span (2 points compensation) |
Sensor monitoring | |
Sensor break | Selectable, Upscale (≥21.0 mA) or downscale (≤3.6 mA) action |
Sensor short-circiut | Fixed, downscale (≤3.6 mA) action |
Environmental Factors | |
Ambient temperature | -40 to +85 °C (storage and operation) |
Humidity | 0 to 100 %RH |
Vibrations | Acc. to IEC 60068-2-6, test Fc, 10-2000 Hz, 10g |
EMC | Acc. to IEC 61326-1 |
Accuracy and stability | |
Basic accuracy | Max. of ±0.2 °C or ±0.2 % of span |
Temperature influence | Max. of ±0.01 °C per °C or ±0.01 % of span per °C |
Long-term stability | ±0.1 % of span per year |
Design | |
Material | Thermoplastic * |
Weight | 16 g |
Protection class | IP67 |
Process connection in | M12x1, 4-pin, A-coded, Female |
Process connection out | M12x1, 4-pin, A-coded, Male |
General data | |
Isolation in – out | Unisolated |
Power supply | 7.5 to 32 VDC |
* Meets the hygienic requirements on materials and products intended to be in contact with food & beverage according to Regulation (EC) 1935/2004