We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Hioki’s FT6380 Clamp On Ground Resistance Tester allows for effortless measurement of grounding resistance by simply clamping to multiple-grounded ground wires. The slim, compact sensor provides AC current measurement functionalities to measure currents ranging from leakage current on the order of several mA to load currents of up to 60A.
Measurement principle | Instrument has two cores for voltage injection and current measurement. From the defined voltage and measured current, the total circuit loop resistance is calculated |
Earthing resistance | 0.20 Ω (0.01 Ω resolution) to 1600 Ω (20 Ω resolution), 10 ranges, Zero suppression: Less than 0.02 Ω, |
AC Current measurement | 20.00 mA (0.01 mA resolution) to 60.0 A (0.1 A resolution), 5 ranges, Zero suppression: Less than 0.05 mA, |
Max. allowable input | 100 A AC continuous, AC 200 A for 2 minutes or shorter (at 50/60 Hz, requires derating at frequency) |
Max. rated voltage to earth | 600 V AC CAT IV |
Memory function | 2000 data |
Alarm function | For resistance measurement and current measurement, Beeps when measured value is less than or greater than threshold |
Other functions | Data hold, Backlight, Filter, Auto power save |
Display | Digital LCD, max. 2000 dgt., Display refresh rate: 500 ms (2 times/s) |
Waterproof and dustproof | IP40 (EN60529) with clamp sensor closed |
Communication interface | N/A |
Power supply | LR6 (AA) alkaline battery ×2, Max. rated power: 450 mVA, Continuous use : 35 hr (in-house testing conditions) |
Core jaw dia. | φ 32 mm (1.26 in) |
Dimensions and mass | 73 mm (2.87 in) W × 218 mm (8.58 in) H × 43 mm (1.69 in) D, 620 g (21.9 oz) |
Accessories | Carrying case ×1, Resistance check loop (1 Ω, 25 Ω) ×1, Strap ×1, LR06 (AA) alkaline battery ×2, Instruction manual ×1 |
- Transmission Towers
- Swimming Pools
- Storage Tanks
- Cell Towers