We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Model | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Rated Output Voltage | V | 30 | 80 | 160 | 250 | 800 | |
Rated Output Current | A | 36 | 13.5 | 7.2 | 4.5 | 1.44 | |
Rated Output Power | W | 360 | 360 | 360 | 360 | 360 | |
Power Ratio | -- | 3 | 3 | 3.2 | 3.125 | 3.2 | |
Constant Voltage Mode | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Line Regulation (*1) | mV | 18 | 43 | 83 | 128 | 403 | |
Load Regulation (*2) | mV | 20 | 45 | 85 | 130 | 405 | |
Ripple and Noise (*3) | |||||||
| p-p (*4) | mV | 60 | 60 | 60 | 80 | 150 |
| r.m.s (*5) | mV | 7 | 7 | 12 | 15 | 30 |
Temperature coefficient | ppm/°C | 100ppm/°C of rated output voltage, after a 30 minute warm-up. | |||||
Remote sense compensation voltage | V/wire | 0.6 | 0.6 | 0.6 | 1 | 1 | |
Rise Time (*6) | |||||||
| Rated Load | ms | 50 | 50 | 100 | 100 | 150 |
| No Load | ms | 50 | 50 | 100 | 100 | 150 |
Fall Time (*7) | |||||||
| Rated Load | ms | 50 | 50 | 100 | 150 | 300 |
| No Load | ms | 500 | 500 | 1000 | 1200 | 2000 |
Transient response time (*8) | ms | 1 | 1 | 2 | 2 | 2 | |
Constant Current Mode | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Line regulation (*1) | mA | 41 | 18.5 | 12.2 | 9.5 | 6.44 | |
Load regulation (*9) | mA | 41 | 18.5 | 12.2 | 9.5 | 6.44 | |
Ripple and noise | |||||||
| r.m.s (*5 ) | mA | 72 | 27 | 15 | 10 | 5 |
Temperature coefficient | ppm/°C | 200ppm/°C of rated output current, after a 30 minute warm-up. | |||||
Protection Function | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Over voltage protection (OVP) | |||||||
| Setting range | V | 3-33 | 8-88 | 16-176 | 20-275 | 20-880 |
| Setting accuracy |
| ± (2% of rated output voltage) | ||||
Over current protection (OCP) | |||||||
| Setting range | A | 3.6-39.6 | 1.35-14.85 | 0.72-7.92 | 0.45-4.95 | 0.144-1.584 |
| Setting accuracy |
| ± (2% of rated output current) | ||||
Over temperature protection (OTP) | |||||||
| Operation |
| Turn the output off. | ||||
Low AC input protection (AC-FAIL) | |||||||
| Operation |
| Turn the output off. | ||||
Power limit (POWER LIMIT) | |||||||
| Operation |
| Over power limit. | ||||
| Value (fixed) |
| Approx. 105% of rated output power | ||||
Front Panel | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Display, 4 digits | |||||||
| Voltage accuracy 0.1% + | mV | 20 | 20 | 100 | 200 | 400 |
| Current accuracy 0.1% + | mA | 40 | 20 | 5 | 5 | 2 |
Programming and Measurement (Interface) | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Voltage programming accuracy 0.1% + | mV | 10 | 10 | 100 | 200 | 400 | |
Current programming accuracy 0.1% + | mA | 30 | 10 | 5 | 5 | 2 | |
Voltage programming resolution | mV | 1 | 2 | 3 | 5 | 14 | |
Current programming resolution | mA | 1 | 1 | 1 | 1 | 1 | |
Voltage measurement accuracy 0.1% + | mV | 10 | 10 | 100 | 200 | 400 | |
Current measurement accuracy 0.1% + | mA | 30 | 10 | 5 | 5 | 2 | |
Voltage measurement resolution | mV | 1 | 2 | 3 | 5 | 14 | |
Current measurement resolution | mA | 1 | 1 | 1 | 1 | 1 |
Common Specification | |||||||
Input Characteristics | PSW | 30V | 80V | 160V | 250V | 800V | |
Nominal input rating |
| 100Vac to 240Vac, 50Hz to 60Hz, single phase | |||||
Input voltage range |
| 85Vac ~ 265Vac | |||||
Input voltage range |
| 47Hz ~ 63Hz | |||||
Maximum input current | |||||||
| 100Vac | A | 360W: 5A, 720W: 10A, 1080W: 15A | ||||
| 200Vac | A | 360W: 2.5A, 720W: 5A, 1080W: 7.5A | ||||
Inrush current | A | < 25A for 360W, < 50A for 720W, < 75A for 1080W | |||||
Maximum input power | VA | 360W: 500VA, 720W: 1000VA, 1080W: 1500VA | |||||
Power factor | |||||||
| 100Vac |
| 0.99 | ||||
| 200Vac |
| 0.97 | ||||
Efficiency | |||||||
| 100Vac | % | 77 | 78 | 79 | 79 | 80 |
| 200Vac | % | 79 | 80 | 81 | 81 | 82 |
Hold-up time |
| 20ms or greater | |||||
Analog Programming and Monitoring | PSW | 30V | 80V | 160V | 250V | 800V | |
External voltage control output voltage |
| Accuracy and linearity: ±0.5% of rated output voltage. | |||||
External voltage control output current |
| Accuracy and linearity: ±1% of rated output current. | |||||
External resistor control output voltage |
| Accuracy and linearity: ±1.5% of rated output voltage. | |||||
External resistor control output current |
| Accuracy and linearity: ±1.5% of rated output current. | |||||
Output voltage monitor | |||||||
| Accuracy | % | ±1 | ±1 | ±1 | ±2 | ±2 |
Output current monitor | |||||||
| Accuracy | % | ±1 | ±1 | ±1 | ±2 | ±2 |
Shutdown control |
| Turns the output off with a LOW (0V to 0.5V) or short-circuit. | |||||
Output on/off control |
| Possible logic selections: | |||||
CV/CC/ALM/PWR ON/OUT ON indicator |
| Photocoupler open collector output; Maximum voltage 30V, maximum sink current 8mA. | |||||
Series and Parallel Capability | PSW | 30V | 80V | 160V | 250V | 800V | |
Parallel number | Units | 3 | 3 | 3 | 3 | 3 | |
Series Number | Units | 2 | 2 | 2 | None | None | |
Front Panel | |||||||
Indications |
| GREEN LED's: | |||||
Buttons |
| Function, OVP/OCP, Set, Test, Lock/Local, PWR DSPL, Output | |||||
Knobs |
| Voltage, Current | |||||
USB port |
| Type A USB connector | |||||
Interface Capabilities | |||||||
USB |
| TypeA: Host, TypeB: Slave, Speed: 1.1/2.0, USB Class: CDC(Communications Device Class) | |||||
LAN |
| MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask | |||||
GPIB |
| Optional: GUG-001 (GPIB to USB Adapter) | |||||
RS-232 |
| Optional: GUR-001 (RS-232 to USB Cable) | |||||
Environmental Conditions | |||||||
Operating temperature |
| 0˚C to 50˚C | |||||
Storage temperature |
| -25˚C to 70˚C | |||||
Operating humidity |
| 20% to 85% RH; No condensation | |||||
Storage humidity |
| 90% RH or less; No condensation | |||||
Altitude |
| Maximum 2000m | |||||
General Specifications | |||||||
Weight (main unit only) | kg | Approx. 3kg for 360W, | |||||
Dimensions (WxHxD) | mm3 | 360W: 71×124×350mm, | |||||
Cooling |
| Forced air cooling by internal fan. | |||||
EMC |
| Complies with the European EMC directive 2004/108 /EC for Class A test and measurement products. | |||||
Safety |
| Complies with the European Low Voltage Directive 2006 /95/EC and carries the CE-marking. | |||||
Withstand voltage |
| Between input and chassis: No abnormalities at 1500 Vac for 1 minute. | |||||
Between input and output: No abnormalities at 3000 Vac for 1 minute. | |||||||
Between output and chassis: | |||||||
No abnormalities at 1500 Vdc for 1 minute for 250V, 800V models. | |||||||
Insulation resistance |
| Between input and chassis: 500 Vdc, 100MΩ or more | |||||
Between input and output: 500 Vdc, 100MΩ or more | |||||||
Between output and chassis: | |||||||
1000Vdc, 100MΩ or more for 800V models. |
Notes
*1: At 85 ~ 132Vac or 170 ~ 265Vac, constant load.
*2: From No-load to Full-load, constant input voltage. Measured at the sensing point in Remote Sense.
*3: Measure with JEITA RC-9131B (1:1) probe
*4: Measurement frequency bandwidth is 10Hz to 20MHz.
*5: Measurement frequency bandwidth is 5Hz to 1MHz.
*6: From 10% to 90% of rated output voltage, with rated resistive load.
*7: From 90% to 10% of rated output voltage, with rated resistive load.
*8: Time for output voltage to recover within 0.1% + 10mV of its rated output for a load change from 50 to 100% of its rated output current.
*9: For load voltage change, equal to the unit voltage rating, constant input voltage.