We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Model |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
|
Rated Output Voltage |
V |
30 |
80 |
160 |
250 |
800 |
|
Rated Output Current |
A |
36 |
13.5 |
7.2 |
4.5 |
1.44 |
|
Rated Output Power |
W |
360 |
360 |
360 |
360 |
360 |
|
Power Ratio |
-- |
3 |
3 |
3.2 |
3.125 |
3.2 |
|
Constant Voltage Mode |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
|
Line Regulation (*1) |
mV |
18 |
43 |
83 |
128 |
403 |
|
Load Regulation (*2) |
mV |
20 |
45 |
85 |
130 |
405 |
|
Ripple and Noise (*3) |
|||||||
|
p-p (*4) |
mV |
60 |
60 |
60 |
80 |
150 |
|
r.m.s (*5) |
mV |
7 |
7 |
12 |
15 |
30 |
Temperature coefficient |
ppm/°C |
100ppm/°C of rated output voltage, after a 30 minute warm-up. |
|||||
Remote sense compensation voltage |
V/wire |
0.6 |
0.6 |
0.6 |
1 |
1 |
|
Rise Time (*6) |
|||||||
|
Rated Load |
ms |
50 |
50 |
100 |
100 |
150 |
|
No Load |
ms |
50 |
50 |
100 |
100 |
150 |
Fall Time (*7) |
|||||||
|
Rated Load |
ms |
50 |
50 |
100 |
150 |
300 |
|
No Load |
ms |
500 |
500 |
1000 |
1200 |
2000 |
Transient response time (*8) |
ms |
1 |
1 |
2 |
2 |
2 |
|
Constant Current Mode |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
|
Line regulation (*1) |
mA |
41 |
18.5 |
12.2 |
9.5 |
6.44 |
|
Load regulation (*9) |
mA |
41 |
18.5 |
12.2 |
9.5 |
6.44 |
|
Ripple and noise |
|||||||
|
r.m.s (*5 ) |
mA |
72 |
27 |
15 |
10 |
5 |
Temperature coefficient |
ppm/°C |
200ppm/°C of rated output current, after a 30 minute warm-up. |
|||||
Protection Function |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
|
Over voltage protection (OVP) |
|||||||
|
Setting range |
V |
3-33 |
8-88 |
16-176 |
20-275 |
20-880 |
|
Setting accuracy |
|
± (2% of rated output voltage) |
||||
Over current protection (OCP) |
|||||||
|
Setting range |
A |
3.6-39.6 |
1.35-14.85 |
0.72-7.92 |
0.45-4.95 |
0.144-1.584 |
|
Setting accuracy |
|
± (2% of rated output current) |
||||
Over temperature protection (OTP) |
|||||||
|
Operation |
|
Turn the output off. |
||||
Low AC input protection (AC-FAIL) |
|||||||
|
Operation |
|
Turn the output off. |
||||
Power limit (POWER LIMIT) |
|||||||
|
Operation |
|
Over power limit. |
||||
|
Value (fixed) |
|
Approx. 105% of rated output power |
||||
Front Panel |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
|
Display, 4 digits |
|||||||
|
Voltage accuracy 0.1% + |
mV |
20 |
20 |
100 |
200 |
400 |
|
Current accuracy 0.1% + |
mA |
40 |
20 |
5 |
5 |
2 |
Programming and Measurement (Interface) |
PSW |
30-36 |
80-13.5 |
160-7.2 |
250-4.5 |
800-1.44 |
|
Voltage programming accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
|
Current programming accuracy 0.1% + |
mA |
30 |
10 |
5 |
5 |
2 |
|
Voltage programming resolution |
mV |
1 |
2 |
3 |
5 |
14 |
|
Current programming resolution |
mA |
1 |
1 |
1 |
1 |
1 |
|
Voltage measurement accuracy 0.1% + |
mV |
10 |
10 |
100 |
200 |
400 |
|
Current measurement accuracy 0.1% + |
mA |
30 |
10 |
5 |
5 |
2 |
|
Voltage measurement resolution |
mV |
1 |
2 |
3 |
5 |
14 |
|
Current measurement resolution |
mA |
1 |
1 |
1 |
1 |
1 |
Common Specification |
|||||||
Input Characteristics |
PSW |
30V |
80V |
160V |
250V |
800V |
|
Nominal input rating |
|
100Vac to 240Vac, 50Hz to 60Hz, single phase |
|||||
Input voltage range |
|
85Vac ~ 265Vac |
|||||
Input voltage range |
|
47Hz ~ 63Hz |
|||||
Maximum input current |
|||||||
|
100Vac |
A |
360W: 5A, 720W: 10A, 1080W: 15A |
||||
|
200Vac |
A |
360W: 2.5A, 720W: 5A, 1080W: 7.5A |
||||
Inrush current |
A |
< 25A for 360W, < 50A for 720W, < 75A for 1080W |
|||||
Maximum input power |
VA |
360W: 500VA, 720W: 1000VA, 1080W: 1500VA |
|||||
Power factor |
|||||||
|
100Vac |
|
0.99 |
||||
|
200Vac |
|
0.97 |
||||
Efficiency |
|||||||
|
100Vac |
% |
77 |
78 |
79 |
79 |
80 |
|
200Vac |
% |
79 |
80 |
81 |
81 |
82 |
Hold-up time |
|
20ms or greater |
|||||
Analog Programming and Monitoring |
PSW |
30V |
80V |
160V |
250V |
800V |
|
External voltage control output voltage |
|
Accuracy and linearity: ±0.5% of rated output voltage. |
|||||
External voltage control output current |
|
Accuracy and linearity: ±1% of rated output current. |
|||||
External resistor control output voltage |
|
Accuracy and linearity: ±1.5% of rated output voltage. |
|||||
External resistor control output current |
|
Accuracy and linearity: ±1.5% of rated output current. |
|||||
Output voltage monitor |
|||||||
|
Accuracy |
% |
±1 |
±1 |
±1 |
±2 |
±2 |
Output current monitor |
|||||||
|
Accuracy |
% |
±1 |
±1 |
±1 |
±2 |
±2 |
Shutdown control |
|
Turns the output off with a LOW (0V to 0.5V) or short-circuit. |
|||||
Output on/off control |
|
Possible logic selections: |
|||||
CV/CC/ALM/PWR ON/OUT ON indicator |
|
Photocoupler open collector output; Maximum voltage 30V, maximum sink current 8mA. |
|||||
Series and Parallel Capability |
PSW |
30V |
80V |
160V |
250V |
800V |
|
Parallel number |
Units |
3 |
3 |
3 |
3 |
3 |
|
Series Number |
Units |
2 |
2 |
2 |
None |
None |
|
Front Panel |
|||||||
Indications |
|
GREEN LED's: |
|||||
Buttons |
|
Function, OVP/OCP, Set, Test, Lock/Local, PWR DSPL, Output |
|||||
Knobs |
|
Voltage, Current |
|||||
USB port |
|
Type A USB connector |
|||||
Interface Capabilities |
|||||||
USB |
|
TypeA: Host, TypeB: Slave, Speed: 1.1/2.0, USB Class: CDC(Communications Device Class) |
|||||
LAN |
|
MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask |
|||||
GPIB |
|
Optional: GUG-001 (GPIB to USB Adapter) |
|||||
RS-232 |
|
Optional: GUR-001 (RS-232 to USB Cable) |
|||||
Environmental Conditions |
|||||||
Operating temperature |
|
0˚C to 50˚C |
|||||
Storage temperature |
|
-25˚C to 70˚C |
|||||
Operating humidity |
|
20% to 85% RH; No condensation |
|||||
Storage humidity |
|
90% RH or less; No condensation |
|||||
Altitude |
|
Maximum 2000m |
|||||
General Specifications |
|||||||
Weight (main unit only) |
kg |
Approx. 3kg for 360W, |
|||||
Dimensions (WxHxD) |
mm3 |
360W: 71×124×350mm, |
|||||
Cooling |
|
Forced air cooling by internal fan. |
|||||
EMC |
|
Complies with the European EMC directive 2004/108 /EC for Class A test and measurement products. |
|||||
Safety |
|
Complies with the European Low Voltage Directive 2006 /95/EC and carries the CE-marking. |
|||||
Withstand voltage |
|
Between input and chassis: No abnormalities at 1500 Vac for 1 minute. |
|||||
Between input and output: No abnormalities at 3000 Vac for 1 minute. |
|||||||
Between output and chassis: |
|||||||
No abnormalities at 1500 Vdc for 1 minute for 250V, 800V models. |
|||||||
Insulation resistance |
|
Between input and chassis: 500 Vdc, 100MΩ or more |
|||||
Between input and output: 500 Vdc, 100MΩ or more |
|||||||
Between output and chassis: |
|||||||
1000Vdc, 100MΩ or more for 800V models. |
Notes
*1: At 85 ~ 132Vac or 170 ~ 265Vac, constant load.
*2: From No-load to Full-load, constant input voltage. Measured at the sensing point in Remote Sense.
*3: Measure with JEITA RC-9131B (1:1) probe
*4: Measurement frequency bandwidth is 10Hz to 20MHz.
*5: Measurement frequency bandwidth is 5Hz to 1MHz.
*6: From 10% to 90% of rated output voltage, with rated resistive load.
*7: From 90% to 10% of rated output voltage, with rated resistive load.
*8: Time for output voltage to recover within 0.1% + 10mV of its rated output for a load change from 50 to 100% of its rated output current.
*9: For load voltage change, equal to the unit voltage rating, constant input voltage.