GW Instek PSU 6-200 6 Volt 200 Amp DC Power Supply

SKU
PSU 6-200
Manufacturer:
GW Instek
GW Instek

Key Features:

  • Voltage Rating: 6V
  • Output Power Rating: 1200W
  • CV / CC Priority ; Particularly Suitable for the Battery & LED industry
  • Adjustable Slew Rate
  • Series and Parallel operation (2 Units in Series or 4 units in parallel)
  • High Efficiency and High Power Density
  • 19 Inch Rack Mount Size Design
  • Standard Interface: LAN, RS-232 & RS-485, USB(Device / Host), Analog Control Interface
  • Optional Interface:GPIB, Analog Control Interface(Isolated Voltage Control), Analog Control Interface(Isolated Current Control)
Product Options:
$2,310.00

Secured and trusted checkout with

     The PSU 6-200 has a power range up to 1200W. The PSU 6-200 can be joined series for maximum 2 units or in parallel for maximum 4 units. A very useful feature for DUT protection are the CV and CC selections. CV mode works when the power supply is turned on and when the CC mode is activated the current becomes stable. To limit the current spike occurred at the threshold voltage and protect the DUT from the inrush current damage the PSU-Series is able to run under CC priority.

     Another way the PSU prevents DUT damage is with the OVP and OCP protections. Both the OVP and OCP levels can select the range of 10% to 110%, with the default level set at 110%, of the rated voltage / current of the power supply. The PSU has available at the rear panel the LabView driver. The driver is for external control of the power On/Off and the external monitoring of the power output voltage and current.

Output       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Rated Output Voltage (*1)

 

V

6

12.5

20

40

60

Rated Output Current (*2)

 

A

200

120

76

38

25

 Rated Output Power

 

W

1200

1500

1520

1520

1500

Constant Voltage Mode       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Line regulation (*3)

 

mV

2.6

3.25

4

6

8

Load regulation (*4)

 

mV

2.6

3.25

4

6

8

Ripple and noise (*5)

p-p (*6)

mV

60

60

60

60

60

 

r.m.s. (*7)

mV

8

8

8

8

8

Temperature coefficient

 

ppm/ °C

100ppm/°C of rated  output voltage, after a 30 minute warm-up.

Remote sense compensation voltage (single wire)

 

V

1

1

1

2

3

Rise time (*8)

Rated load

ms

80

80

80

80

80

 

No load

ms

80

80

80

80

80

Fall time (*9)

Rated load 

ms

10

50

50

80

80

 

No load

ms

500

700

800

1000

1100

Transient response time (*10)

 

ms

1.5

1

1

1

1

Constant Current Mode       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

 Line regulation (*3)

 

mA

22

14

9.6

5.8

4.5

Load regulation (*11)

 

mA

45

29

20.2

12.6

10

Ripple and noise (*12)

r.m.s.

mA

400

240

152

95

75

Temperature coefficient

 

ppm/ °C

100ppm/°C of rated  output voltage, after a 30 minute warm-up.

Protection Function       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Over voltage protection (OVP)

Setting range

V

0.6-6.6

1.25-13.75

2-22 

4-44

5-66

 

Setting accuracy

V

0.06

0.125

0.2

0.4

0.6

Over current protection (OCP)

Setting range

A

5-220

5-132

5-83.6

3.8-41.8

2.5-27.5

 

Setting accuracy

A

4

2.4

1.52

0.76

0.5

Under voltage limit (UVL)

Setting range

 

0-6.3

0-13.12

0-21

0-42

0-63

Over temperature protection (OHP)

Operation

 

Turn the output off.

Incorrect sensing connection protection (SENSE)

Operation

 

Turn the output off.

Low AC input protection (AC-FAIL)

Operation

 

Turn the output off.

Shutdown (SD)

Operation

 

Turn the output off.

Power limit (POWER LIMIT)

Operation

 

Over power limit.

 

Value (fixed)

 

Approx. 105% of rated output power    

Analog Programming and Monitoring       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

External voltage control output voltage

 

 

Turn the output off.

External voltage control output current

 

 

Turn the output off.

External resistor control output voltage

 

 

Turn the output off.

External resistor control output current

 

 

Turn the output off.

Output voltage monitor

 

 

Turn the output off.

Output current monitor

 

 

Turn the output off.

Shutdown control

 

 

Turns the output off with a LOW (0V to 0.5V) or short-circuit.

Output on/off control

 

 

Possible logic selections:
Turn the output on using a LOW (0V to 0.5V) or short-circuit, turn the output off using a HIGH (4.5V to 5V) or open-circuit
Turn the output on using a HIGH (4.5V to 5V) or open-circuit, turn the output off using a LOW (0V to 0.5V) or short-circuit.

Alarm clear control

 

 

Clear alarms with a LOW (0V to 0.5V) or short-circuit.

CV/CC/ALM/PWR ON/OUT ON indicator

 

 

Photocoupler open collector output; Maximum voltage 30V, maximum sink current 8mA.

Trigger out

 

 

Maximum low level output = 0.8V; minimum high level output = 2V; Maximum source current = 8mA.

Trigger in

 

 

Maximum low level input voltage = 0.8V; minimum high level input votage = 2.0V, Maximum sink current = 8mA.

Front Panel       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Display, 4 digits

 

 

 

 

 

 

 

Voltage accuracy

0.1% +

mV

12

25

40

80

120

Current accuracy

0.2% +

mA

600

360

228

114

75

Indications

 

 

GREEN LED's: CV, CC, V, A, VSR, ISR, DLY, RMT, LAN, M1, M2, M3, RUN, Output ON
RED LED's: ALM, ERR

Buttons

 

 

Lock/Local(Unlock), PROT(ALM_CLR), Function(M1), Test(M2), Set(M3), Shift, Output

Knobs

 

 

Voltage, Current

USB port

 

 

Type A USB connector

Programming and Measurement (RS-232/485, USB, LAN, GPIB)       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Output voltage programming accuracy

0.05% +

mV

3

6.25

10

20

30

Output current programming accuracy

0.2% +

mA

200

120

76

38

25

Output voltage programming resolution

 

mV

0.2

0.4

0.7

1.3

2

Output current programming resolution

 

mA

6

4

2.5

1.2

0.8

Output voltage measurement accuracy

0.1% +

mV

6

12.5

20

40

60

Output current measurement accuracy

0.2% +

mA

400

240

152

76

50

Output voltage measurement resolution

 

mV

0.2

0.4

0.7

1.3

2

Output current measurement resolution

 

mA

6

4

2.5

1.2

0.8

Input Characteristics       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Norminal input rating

 

 

100Vac to 240Vac, 50Hz to 60Hz, single phase

Input voltage range

 

 

85Vac ~ 265Vac

Input frequency range

 

 

47Hz ~ 63Hz

Maximum input current

100Vac

A

21

 

200Vac

A

11

Inrush current

 

 

Less than 50A.

Maximum input power

 

VA

2000

Power factor

100Vac

 

0.99

 

200Vac

 

0.98

Efficiency (*13)

100Vac

 

77

82

83

84

84

 

200Vac

 

79

85

86

87

87

Hold-up time

 

 

20ms or greater

Interface Capabilities       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

USB

 

 

TypeA: Host, TypeB: Slave, Speed: 1.1/2.0, USB Class: CDC(Communications Device Class)

LAN

 

 

MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask

GPIB

 

 

SCPI - 1993, IEEE 488.2 compliant interface

RS-232/RS-485

 

 

Complies with EIA232D / EIA485 Specifications

Environment Conditions       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Operaing temperature

 

 

0˚C to 50˚C

Storage temperature

 

 

-25˚C to 70˚C

Operating humidity

 

 

20% to 85% RH; No condensation

Storage humidity

 

 

90% RH or less; No condensation

Altitude

 

 

Maximum 2000m

Environment Conditions       

Model

 

PSU

6-200

12.5-120

20-76

40-38

60-25

Weight

main unit only

kg

Less than 8.7kg

Dimensions

(W×H×D)

mm3

423×43.6×447.2

Cooling

 

 

Forced air cooling by internal fan.

EMC

 

 

Complies with the European EMC directive 2004 /108/EC for Class A test and measurement products.

Safety

 

 

Complies with the European Low Voltage Directive 2006 /95/EC and carries the CE-marking.

Withstand voltage

 

 

AC to Chassis: 1500Vac/1min
AC to Output terminal: 3000Vac/1min
Output terminal to Chassis: 1000Vdc/1min

Insulation resistance

 

 

Chassis and output terminal; chassis and AC input; AC input and output terminal: 100MΩ or more (DC 500V)

 

 

 

 

 

 

 

 

Notes:
(*1) Minimum voltage is guaranteed to maximum 0.2% of the rated output voltage.
(*2) Minimum current is guaranteed to maximum 0.4% of the rated output current.
(*3) At 85 ~ 132Vac or 170 ~ 265Vac, constant load.
(*4) From No-load to Full-load, constant input voltage. Measured at the sensing point in Remote Sense.
(*5) Measure with JEITA RC-9131B (1:1) probe
(*6) Measurement frequency bandwidth is 10Hz to 20MHz.
(*7) Measurement frequency bandwidth is 5Hz to 1MHz.
(*8) From 10% to 90% of rated output voltage, with rated resistive load.
(*9) From 90% to 10% of rated output voltage, with rated resistive load.
(*10) Time for output voltage to recover within 0.5% of its rated output for a load change from 10 to 90% of its rated output current. Voltage set point from 10% to 100% of rated output.
(*11) For load voltage change, equal to the unit voltage rating, constant input voltage.
(*12) For 6V model the ripple is measured at 2 ~ 6V output voltage and full output current. For other models, the ripple is measured at 10 ~ 100% output voltage and full output current.
(*13) At rated output power.