GW Instek GPT-9804 5000V AC - 6000V DC Hipot Tester and Safety Analyzer

SKU
GPT-9804
Manufacturer:
GW Instek
GW Instek

Key Features:

  • 200VA AC Test Capacity
  • 240x64 Ice Blue Dot matrix LCD
  • Manual/Auto Mode
  • Function Key for quick selecting
  • High intensity flash for Caution & Status indication
  • Safety INTERLOCK function
  • Zero Crossing Turn-on Operation
  • Controllable Ramp-up Time
  • True RMS Current Measurement
  • High resolution: 1uA for measuring current, 2V for setting voltage
  • PWM amplifier to enhance the power efficiency and reliable testing
  • Max. 100 memory block for test condition (step) setting. And each step can be named individually.
  • Remote terminal on the front panel for “start” and “stop” control
  • Interface : RS-232C, USB device, Signal I/O and GPIB (optional)
Product Options:
$2,380.00

Secured and trusted checkout with

The GPT-9800 Series look to meet a variety of safety standards such as IEC, EN, UL, CSA, GB, JIS and other safety-related requirements for the safety tests of electronic products and components. The GPT-9804 is a 4-in-1 model capable of performing AC withstanding, DC withstanding, insulation resistance and ground bond tests.


 

 

AC WITHSTANDING
Output-Voltage Range0.100kV~ 5.000kV ac
Output-Voltage Resolution2V/step
Output-Voltage Accuracy± (1% of setting +5V) [no load]
Maximum Rated Load200 VA (5kV/40mA)
Maximum Rated Current40mA (0.5kV < V ≦ 5kV)10mA (0.1kV ≦ V≦ 0.5kV)
Output-Voltage WaveformSine wave
Output-Voltage Frequency50 Hz / 60 Hz selectable
Voltage regulation± (1% of reading +5V) [full load → no load]
Voltmeter Accuracy± (1% of reading+ 5V)
Current Measurement Range0.001mA~040.0mA
Current Best Resolution0.001mA / 0.01mA / 0.1mA
AC Current Measurement Accuracy±(1.5% of reading +30 counts) when HISET < 1.00mA±(1.5% of reading +3 counts) when HISET ≧ 1.00mA
Window Comparator MethodYes
ARC DetectYes
RAMP (Ramp-Up Time)0.1~999.9s
TIMER (Test Time)*OFF, 0.5s~999.9s
GNDON/OFF

DC WITHSTANDING

Output-Voltage Range0.100kV~ 6.000kV dc
Output-Voltage Resolution2V/step
Output-Voltage Accuracy± (1% of setting +5V) [no load]
Maximum Rated Load50W (5kV/10mA)
Maximum Rated Current10mA (0.5kV < V ≦ 6kV)2mA (0.1kV ≦ V≦ 0.5kV)
Voltage regulation± (1% of reading +5V) [full load → no load]
Voltmeter Accuracy± (1% of reading+ 5V)
Current Measurement Range0.001mA~010.0mA
Current Best Resolution0.001mA / 0.01mA / 0.1mA
DC Current Measurement Accuracy±(1.5% of reading +30 counts) when HISET < 1.00mA±(1.5% of reading +3 counts) when HISET ≧ 1.00mA
Window Comparator MethodYes
ARC DetectYes
RAMP (Ramp-Up Time)0.1~999.9s
TIMER (Test Time)*OFF, 0.5s~999.9s
GNDON/OFF

INSULATION RESISITANCE

Output-Voltage50V~1000V dc
Output-Voltage Resolution50V/step
Output-Voltage Accuracy± (1% of setting +5V) [no load]
Resistance Measurement Range1MW~ 9500MW
 Test voltageMeasurable RangeAccuracy
 50V≦V≦450V1MW~ 50MW51MW~ 2000MW±(5% of reading +1 count)±(10% of reading +1 counts)
 500V≦V≦1000V1MW~ 500MW501MW~ 9500MW±(5% of reading +1 count)±(10% of reading +1 count)
Window Comparator MethodYes
Output Impedance600kW
RAMP (Ramp-Up Time)0.1~999.9s
TIMER (Test Time)1s~999.9s
GNDOFF (fixed)

GROUND BOND

Output-Current Range03.00A~30.00A ac
Output-Current Resolution0.01A
Output-Current Accuracy3A≦I≦8A : ± (1% of setting +0.2A)8A<I≦30A : ± (1% of setting +0.05A)
Test-Voltage6Vac max. (open circuit)
Test-Voltage Frequency50 Hz / 60 Hz selectable
Resistance Measurement Range10mW~650.0mW
Resistance Measurement Resolution0.1mW
Resistance Measurement Accuracy± (1% of reading +2mW)
Window Comparator MethodYes
TIMER (Test Time)0.5s~999.9s
GNDOFF (fixed)
Test MethodFour Terminal

MEMORY

Single Step MemoryMANU : 100 blocks
Automatic Testing MemoryAUTO : 100 blocks, menu per auto : 16

INTERFACE

RS232Standard
USB (Device)Standard
GPIBOption
Remote Terminal (Front)Standard
SIGNAL IOStandard

General

DISPLY240 x 64 dot matrix LED back light LCD
POWER SOURCEAC100V/120V/220V/230V ±10%, 50/60Hz
POWER CONSUMPTION500VA max.
DIMENSIONS & WEIGHT330(W) x 148(H) x 452(D) mm (Max.),approx.19kg(Max)

* The “TIMER OFF” can only be performed when the tester is in the special MANU mode

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account
  • Safety Testing of Electrical Product in Manufacturing
  •    Power Cord

       Home Appliances

       Information Technology Equipment

       Medical Equipment

  • Quality Assurance Verification
  • Safety Standard Compliance
  • Pre-qualification in R&D
  • Products In This Series

    As low as

    Now $1,131.00