The GPT-9602 hipot voltage breakdown tester is capable of performing AC voltage testing up to 5000 volts and and DC withstandi tests to 6000 volts. It has never been easirer to meet safety standards to IEC, EN, UL, ETL, CSA, GB, JIS and others. The GPT-9601, GPT-9602, GPT-9603, and GPT-9612 are all equipped with the high-efficiency PWM amplifier.
AC WITHSTANDING (All Models) | |||
Output-Voltage Range | 0.10kV~ 5.00kV ac | ||
Output-Voltage Resolution | 10V | ||
Output-Voltage Accuracy | ± (1.5% of setting + 2 counts) with no load | ||
Maximum Rated Load | 100 VA (5kV/20mA) | ||
Maximum Rated Current | 20mA (0.5kV < V ≦ 5kV) 5mA (0.1kV ≦ V≦ 0.5kV) | ||
Output-Voltage Waveform | Sine wave | ||
Output-Voltage Frequency | 50 Hz / 60 Hz selectable | ||
Voltage regulation | ± (1.5% + 2 counts) [full load → no load] | ||
Voltmeter Accuracy | ± (1.5% of reading + 2 counts) | ||
Current Measurement Range | 0.01mA~20.0mA | ||
Current Best Resolution | 0.01mA / 0.1mA | ||
Current Measurement Accuracy | ±(2.0% of reading + 5 counts) when HISET < 1.00mA ±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA | ||
Current Judgment Accuracy | ±(3.0% of setting + 5 counts) when HISET < 1.00mA ±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA | ||
Window Comparator Method | Yes | ||
ARC Detect | Yes | ||
RAMP (Ramp-Up Time) | 0.1s fixed | ||
TIMER (Test Time) | OFF, 1s~180s | ||
GND | ON | ||
DC WITHSTANDING (GPT-9602 & GPT-9603) | |||
Output-Voltage Range | 0.10kV~ 6.00kV dc | ||
Output-Voltage Resolution | 10V | ||
Output-Voltage Accuracy | ± (1.5% of setting + 2 counts) with no load | ||
Maximum Rated Load | 25W (5kV/5mA) | ||
Maximum Rated Current | 6mA (0.5kV < V ≦ 6kV) 2mA (0.1kV ≦ V≦ 0.5kV) | ||
Voltage regulation | ± (1.5% + 2 counts) [full load → no load] | ||
Voltmeter Accuracy | ± (1.5% of reading + 2 counts) | ||
Current Measurement Range | 0.01mA~6.00mA | ||
Current Best Resolution | 0.01mA | ||
Current Measurement Accuracy | ±(2.0% of reading + 5 counts) when HISET < 1.00mA ±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA | ||
Current Judgment Accuracy | ±(3.0% of setting + 5 counts) when HISET < 1.00mA ±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA | ||
Window Comparator Method | Yes | ||
ARC Detect | Yes | ||
RAMP (Ramp-Up Time) | 0.1s fixed | ||
TIMER (Test Time) | OFF, 1s~180s | ||
GND | ON | ||
INSULATION RESISITANCE (GPT-9603 & GPT-9612) | |||
Output-Voltage | 50V, 100V, 250V, 500V, 1000V dc | ||
Output-Voltage Accuracy | ± (3.0% of setting + 1 count) [no load] | ||
Resistance Measurement Range | 1MW~ 2000MW | ||
| Test voltage | Measurable Range | Accuracy |
| 50V/100V/250V | 1~ 50MW 51~ 2000MW | ±(5% of reading + 2MW) ±(10% of reading + 2MW) |
| 500V/1000V | 1~ 500MW 501~ 2000MW | ±(5% of reading + 2MW) ±(10% of reading + 2MW) |
Window Comparator Method | Yes | ||
Output Impedance | 600kW | ||
RAMP (Ramp-Up Time) | 0.1s fixed | ||
TIMER (Test Time) | OFF, 1s~180s | ||
GND | OFF (fixed) | ||
TEST MODE* | |||
SINGLE | ACW, DCW, IR | ||
AUTO | AC-IR, IR-AC, DC-IR, IR-DC | ||
INTERFACE | |||
Remote Terminal (Front) | Standard | ||
SIGNAL IO | Standard | ||
General | |||
DISPLY | 240 x 48 Ice Blue dot matrix LCD | ||
POWER SOURCE | AC 100V-120V/220V-240V ±10%, 50/60Hz | ||
POWER COMSUMPTION | 400VA max. | ||
DIMENSIONS & WEIGHT | 330(W) x 148(H) x 385(D) mm approx.9kg(Max) |
* The available “Test Mode” depends on selected model