GW Instek GPT-15004 5KVac - 6KVdc AC/DC/IR/GB Electrical Safety Analyzer

SKU
GPT-15004
Manufacturer:
GW Instek
GW Instek

Key Features:

  • 5KVac AND/OR 6KVdc Test Capacity (short circuit current > 200 mA)
  • Comply with IEC 61010-2-034
  • 7”TFT LCD
  • Manual / Auto Mode
  • RMS Current Measurement
  • Zero Crossing Turn-on Operation
  • Controllable Ramp-up & Ramp-down Time
  • Capacitive Load Testing Capability up to 47μF
  • Statistics & Analysis Function
  • Sweep Function for DUT Characteristic Analysis
  • Convenience Listed AUTO mode easy to read result & judge
  • Internal Storage & USB Storage available
  • Barcode Function available
  • Setting Data Export / Import
  • Rear panel output available
  • Interface : RS-232C, USB host/device, Signal I/O and option GPIB or LAN
  • Universal power input
Product Options:
$5,824.00

Secured and trusted checkout with

The GW Instek GPT-15004 is the first safety analyzer in the world to comply with IEC 61010-2-034. The GPT-15004 features AC/DC withstanding voltage test, insulation resistance test, AC ground bond test and continuity test.

 

Includes:

• Quick Start Guide x1, CD x1(completed user manual)
Power cord x1, Interlock key x1
Remote terminal Cable GHT-119 x1
Test lead GHT-115 x1, GTL-215 x1

AC WITHSTANDING

Output-Voltage Range

0.050kV~5.000kV

Output-Voltage Resolution

1V

Output-Voltage Accuracy

± (1% of setting + 5V) [no load]

Maximum Rated Load

500 VA (5kV/100mA)

Maximum Rated Current

100mA (0.5kV< V ≦5kV)

10mA (0.05kV≦ V ≦0.5kV)

Output-Voltage Waveform

Sine wave

Output-Voltage Frequency

50 Hz / 60 Hz selectable

Voltage Regulation

± (1% + 5V) [maximum rated load → no load]

Voltmeter Accuracy

± (1% of reading + 5V)

Current Measurement Range

1μA~100.0mA

Current Best Resolution

1μA / 10μA / 100μA

Current Measurement Accuracy

±(1.5% of reading + 30μA)

Current Offset

60μA Maximum

Window Comparator Method

Yes

ARC Detect

Yes

RAMP UP (Rise T ime)

0.1s~999.9s

RAMP DOWN (Fall Time)

0.0s~999.9s

TIMER (Test Time)*

OFF, 0.3s~999.9s

TIMER Accuracy

±(100ppm + 20ms)

WAIT TIME

0.0s~999.9s

GND

ON/OFF

DC WITHSTANDING

Output-Voltage Range

0.050kV~6.000kV

Output-Voltage Resolution

1V

Output-Voltage Accuracy

± (1% of setting + 5V) [no load]

Maximum Rated Load

100W (5kV/20mA)

Maximum Rated Current

20mA (0.5kV< V ≦6kV)

2mA (0.05kV≦ V ≦0.5kV)

Voltage Regulation

± (1% + 5V) [maximum rated load → no load]

Voltmeter Accuracy

± (1% of reading + 5V)

Current Measurement Range

1μA~20.00mA

Current Best Resolution

0.1μA /1μA /10μA

Current Measurement Accuracy

±(1.5% of reading + 3μA) when I Reading < 1mA

±(1.5% of reading + 30μA) when I Reading  ≧1mA

Current Offset

5μA Maximum

Window Comparator Method

Yes

ARC Detect

Yes

RAMP UP (Rise T ime)

0.1s~999.9s

RAMP DOWN (Fall Time)

0.0s~999.9s

TIMER (Test Time)*

OFF, 0.3s~999.9s

TIMER Accuracy

±(100ppm + 20ms)

WAIT TIME

0.0s~999.9s

GND

ON/OFF

INSULATION RESISTANCE

Output Voltage

50V~1200V dc

Output-Voltage Resolution

50V

Output-Voltage Accuracy

± (1% of setting + 5V) [no load]

Resistance Measurement

 

Test Voltage

Display Range

Measurement Range / Accuracy

50V≦V≦100V

0.1MΩ~ 10.00GΩ

0.1MΩ~1MΩ:±(5% of reading + 3 count)

1 MΩ~50MΩ:±(5% of reading + 1 count)

51MΩ~2GΩ:±(10% of reading + 1 count)

150V≦V≦450V

0.1MΩ~ 20.00GΩ

500V≦V≦1200V

0.1MΩ~ 50.00GΩ

0.1MΩ~1MΩ:±(5% of reading + 3 count)

1 MΩ~500MΩ:±(5% of reading + 1 count)

501MΩ~9.999GΩ:±(10% of reading + 1 count)

10GΩ~50GΩ:±(20% of reading + 1 count)*

Voltage Regulation

± (1% + 5V) [maximum rated load → no load]

Voltmeter Accuracy

± (1% of reading + 5V)

Short-Circuit Current

10mA max.

Output Impedance

2kΩ

Window Comparator Method

Yes

RAMP UP (Rise T ime)

0.1s~999.9s

RAMP DOWN (Fall Time)

0.0s~999.9s

TIMER (Test Time)*

OFF, 0.3s~999.9s

TIMER Accuracy

±(100ppm + 20ms)

WAIT TIME

0.0s~999.9s

GND

ON/OFF

Ground Bond

Output-Current

03.00A~32.00A ac

Output-Current Resolution

0.01A

Output-Current Accuracy

3A≦ I≦ 8A:±(1% of reading + 0.2A)

8A<I≦ 32A:±(1% of reading + 0.05A)

Test-Voltage

8Vac max (open circuit)

Test-Voltage Frequency

50Hz/60Hz selectable

Ohmmeter Measurement Range

1mW~ 650mW

Ohmmeter Measurement Resolution

0.1mW

Ohmmeter Measurement Accuracy

±(1% of reading + 2 mW)

Window Comparator Method

Yes

TIMER (Test Time)

0.3s~999.9s

TIMER Accuracy

±(100ppm + 20ms)

Test Method

Four Terminal

GND

ON/OFF

Continuity Test

Output-Current

100mA dc (fixed)

Ohmmeter Measurement Range

0.10W~ 70.00W

Ohmmeter Measurement Resolution

0.01W

Ohmmeter Measurement Accuracy

±(10% of reading + 2 W)

Window Comparator Method

Yes

TIMER (Test Time)*

0.3s~999.9s

TIMER Accuracy

±(100ppm + 20ms)

MEMORY

Single Step Memory

MANU:100 blocks

Automatic Testing Memory

AUTO:100 blocks, manu per auto:10

INTERFACE

REMOTE (Front) terminal

Standard

USB host (Front)

Standard

Rear Output

Standard

RS-232C

Standard

USB device

Standard

Signal I/O

Standard

GPIB

Option

LAN

Option

DISPLAY

 

7” color LCD

POWER SOURCE

 

AC 100V~240V ± 10%, 50Hz/60Hz

DIMENSION & WEIGHT

GPT-15001/15002/15003

GPT-15004

380(W) x 148(H) x 492(D) mm; Approx. 17kg

380(W) x 148(H) x 546(D) mm; Approx. 21kg

* When Ground Mode is “ON”, the measurement range is 30GΩ max. and adding 10% error for accuracy.

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as

Now $2,771.00

As low as

Now $3,182.00

As low as

Now $3,891.00