Fluke Calibration 9143 Field Metrology Well

SKU
9143
Manufacturer:
Fluke Calibration
Fluke Calibration

Key Features:

  • Lightweight, portable, and fast
  • Cool to –25 °C in 15 minutes and heat to 660°C in 15 minutes
  • Built-in two-channel readout for PRT, RTD, thermocouple, 4-20 mA current
  • True reference thermometry with accuracy to ±0.01 °C
  • On-board automation and documentation
  • Metrology performance in accuracy, stability, uniformity, and loading
  • Product Options:
    $5,861.00

    Secured and trusted checkout with

    The new 914X Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance.

    Field Metrology Wells are packed with functionality and are remarkably easy to use. They are lightweight, small, and quick to reach temperature set points, yet they are stable, uniform, and accurate. These industrial temperature loop calibrators are perfect for performing transmitter loop calibrations, comparison calibrations, or simple checks of thermocouple sensors. With the “process” option, there is no need to carry additional tools into the field. This optional built-in two-channel readout reads resistance, voltage, and 4–20 mA current with 24 volt loop power. It also has on-board automation and documentation. Combined, the three models (9142, 9143, and 9144—each with a “process” option) cover the wide range of –25 °C to 660 °C.

     

    Temperature Range at 23 C 33 C to 350 C(91 F to 662 F)
    Display Accuracy +/- 0.2 C Full Range 
    Stability +/- 0.02 C at 33 C
    +/- 0.02 C at 200 C
    +/- 0.03 C at 350 C
    Axial Uniformity at 40 mm (1.6 in) +/- 0.04 C at 33 C
    +/- 0.1 C at 200 C
    +/- 0.2 C at 350 C
    Radial Uniformity +/- 0.01 C at 33 C
    +/- 0.015 C at 200 C
    +/- 0.02 C at 350 C
    Loading Effect (with a 6.35 mm reference
    probe and three 6.35 mm probes) 
    +/- 0.015 C Full Range
    Hysteresis 0.03
    Operating Conditions 0 C to 50 C, 0 % to 90 % RH (non-condensing) 
    Environmental Conditions (for all specifications
    except temperature range)  
    13 C to 33 C
    Immersion (Well) Depth 150 mm (5.9 in) 
    Insert OD 25.3 mm (1.00 in)
    Heating Time 5 min: 33 C to 350 C
    Cooling Time 32 min: 350 C to 33 C
    14 min: 350 C to 100 C
    Resolution 0.01 Deg 
    Display LCD, C or F user-selectable 
    Size (H x W x D) 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in) 
    Weight 7.3 kg (16 lb)
    Power Requirements 100 V to 115 V (+/- 10 %), 50/60 Hz, 1400 W 230 V (+/- 10 %), 50/60 Hz,
    Computer Interface RS-232 and 9930 Interface-it control software included 

     9143 Includes:

    • Users Manual
    • Technical Manual on CD
    • Serial Cable
    • 9930, Interface-itSoftware
    • Report of Calibration
    • Insert Removal Tool

    9143-_-P Includes:

    • Users Manual
    • Technical Manual on CD
    • Serial Cable
    • 9930, Interface-it Software
    • Report of Calibration
    • Test leads
    • Insert Removal Tool
    • 6-pin DIN Connector for Reference Probe

     