Fluke Calibration 9140 Dry-Well Calibrator

SKU
9140
Manufacturer:
Fluke Calibration
Fluke Calibration

Key Features:

  • Lightweight and very portable
  • Accuracy to +/-0.25 C 
  • RS-232 and Interface-it software included 
  • Easy to recalibrate 
  • The instrument features:
    -Rapid heating and cooling
    -Interchangeable multiple hole aluminum probe sleeves
    -Convenient hand strap
    -RS-232 interface capability
  • Built in programmable features include:
    -Temperature scan rate control
    -Temperature switch hold
    -Eight Setpoint memory
    -Adjustable readout in C or F
Product Options:
$4,176.00

Secured and trusted checkout with

The 9140 has a temperature range of 35 C to 350 C, and it reaches its maximum temperature in 12 minutes. At six pounds, it’s small enough to easily carry in one hand. It’s truly a unique innovation in dry-wells.

The unit has a stability of +/-0.05 C or better and a uniformity of at least 0.4 C in the largest-diameter wells and 0.1 C in the smaller wells. Despite its small size, this unit performs.

Use the display, calibrated to +/-0.5 C, as your reference, or use an external thermometer for maximum calibration accuracy. With three removable inserts to choose from, the 9140 is as versatile as it is fast.

Range 35 °C to 350 °C
(95 °F to 662 °F) 
Accuracy ± 0.5 °C (holes greater than 1/4" [6.35 mm]: ± 1 °C) 
Stability ± 0.03 °C at 50 °C
± 0.05 °C at 350 °C 
Well-to-Well Uniformity ± 0.1 °C with similarly sized wells 
Heating Times 12 minutes from ambient to 350 °C 
Cooling Times 15 minutes from 350 °C to 100 °C 
Stabilization Time 7 minutes 
Immersion Depth 124 mm (4.875") 
Inserts Insert A, B, C, or D included (specify when ordering) 
Outside Insert Dimensions 31.8 mm dia. x 124 mm (1.25 x 4.88 in) 
Computer Interface RS-232 included with free Interface-it software (Model 9930) 
Power 115 V ac (± 10 %), 4.4 A or 230 V ac (± 10 %), 2.2 A, switchable, 50/60 Hz, 500 W 
Size (WxHxD) 152 x 86 x 197 mm
(6 x 3.375 x 7.75 in) 
Weight 2.7 kg (6 lb) 
NIST-Traceable Certificate Data at 50 °C, 100 °C, 150 °C, 200 °C, 250 °C, 300 °C, and 350 °C 
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account