Fluke Calibration's line of portable dry block calibrators is impressive. They' are the most compact, lightest and most portable dry-wells in the world. The Model 9100S weighs only 2 pounds 3 ounces.
The Fluke Model 9100S HDRC handheld dry-well is a low cost solution and you can have four of them for the price of one of other brands. Despite its small size (2 1/4 inches high and 5 inches wide) and light weight, it performs absolutely as well as any dry-well in its class in the world.
It's simple and convenient too. Anyone can learn to use one in less than 12 minutes. It has a range to 300 Deg C (572 Deg F) and is perfect for checking RTDs, thermocouples and bi-metal thermometers in the field.
Plug it in, switch it on, set the temperature with the front panel buttons and insert your probe into the properly sized well. Compare the reading of your device to the display temperature, and the difference is the error in your device. The 9100S HDRC has a display resolution of 0.1 degrees, an accuracy of +/-0.5 Deg C between 33 and 200 Deg C and +/-1 Deg C between 200 and 300 Deg C,a stability of +/-0.3 Deg C and a stabilization time of five minutes.
Range | 35 °C to 375 °C (95 °F to 707 °F) |
Accuracy | ± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C |
Stability | ± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C |
Well-to-Well Uniformity | ± 0.2 °C with sensors of similar size at equal depths within wells |
Heating Times | 35 °C to 375 °C: 9.5 minutes |
Stabilization | 5 minutes |
Cooling Times | 375 °C to 100 °C: 14 minutes |
Well Depth | 102 mm (4 in);1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep |
Removable Inserts; | N/A |
Power | 115 V ac (± 10 %), 55–65 Hz, 1.5 A or 230 V ac (± 10 %), 0.8 A, 45–55 Hz, 175 W |
Size (HxWxD) | 57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in) |
Weight | 1 kg (2 lb 3 oz) |
Computer Interface | RS-232 included with free Interface-it software |
Factory Test Report | Data at 50, 100, 150, 200, 250, 300, and 375 °C |