We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Fluke Calibration 5611T is an especially versatile PTFE coated thermistor probe. With a PTFE encapsulated tip that is just 3 mm (0.12 in) in diameter and a PTFE coating that makes it resistant to most liquids, the PTFE Probe is handy for measuring in a vast variety of applications, including bio-pharmaceuticals. It's even immersible to nearly 20 feet and flexible enough that you could roll it up into a ball in your hand if you wanted to!
The 5611T's thermistor bead is encapsulated in a Mylar sleeve that is encapsulated inside a PTFE sleeve. The PTFE sleeve is then melted around the PTFE-insulated cable, forming a moisture-proof seal.
Resistance | Nominal 10,000 Ω at 25 °C |
Range | 0 °C to 100 °C |
Calibration | R vs. T table with 0.1 °C increments, interpolation equation furnished |
Calibration Uncertainty | Table and equation are accurate to ± 0.01 °C |
Drift | Better than ± 0.01 °C per year (k=3) |
Repeatability | Better than ± 0.005 °C |
Termination | Specify when ordering. |
Manuals