Fluke Calibration 1523-P2 Mega-Kit

SKU
1523-P2-156
Manufacturer:
Fluke Calibration
Fluke Calibration

Key Features:

Fluke Calibration 1523 Reference Thermometer

  • High accuracy
  • PRTs: up to ± 0.011 °C
  • Thermocouples: up to ± 0.24 °C
  • Precision thermistors: ± 0.002 °C
  • Smart connectors to load probe information automatically
  • A simple user interface to quickly see trends

 

Fluke Calibration 5628 Reference Grade Secondary PRT Probe

  • Range to 661°C
  • Meets all ITS-90 requirements for resistance ratios
  • Rtp drift < 20 mK after 500 hours at 661°C
  • Calibrated accuracy of ± 0.006 °C at 0 °C
$7,100.00

Secured and trusted checkout with

The Fluke Calibration 1523-P2 includes a 1523 Reference Thermometer, 5628 PRT probe (–200 °C to 661 °C, 25 ohm (1/4 x 12 inch)), 2373-LTC Lemo to TC Adapter, TPAK, and a 1523/1524 carrying case.

Fluke Calibration 1523 Reference Thermometer Specifications

Specifications
Input channels1
Logging25 readings with statistics
Sample interval
(normal)
1 second
Sample interval (fast mode)0.3 seconds (see technical manual for details)
Sensor typesPRTs, RTDs, Thermistors, and Thermocouples
Thermocouple typesC,E,J,K,L,M,N,T,U,B,R,S
Operating temperature-10°C to 60°C (Best accuracy 13°C to 33°C)
Power requirements3 AA alkaline batteries
Size96 x 200 x 47 mm
(3.75 x 7.9 x 1.86 in)
Weight0.65 kg (1.4 lb)
Environmental conditions for best accuracy13°C to 33°C
Millivolt range and accuracy–10 mV to 75 mV ±(0.005% + 5 μV)
Resistance range and accuracy0 Ω to 400 Ω ±(0.004% + 0.002 Ω)
200 Ω to 50 kΩ ±(0.01% + 0.5 Ω)
50 kΩ to 500 kΩ ±(0.03%)
Temperature coefficient, voltage (–10°C to 13°C , +33°C to 60°C)±(0.001%/°C + 1 μV/°C)
Temperature coefficient, resistance (–10°C to 13°C , +33°C to 60°C)0.0008%/°C + 0.0004 Ω (0 Ω to 400 Ω)
0.002%/°C + 0.1 Ω (0 Ω to 50 kΩ)
0.06%/°C + 0.1 Ω (50 kΩ to 500 kΩ)
Excitation current, resistance1 mA (0 Ω to 400 Ω)
10 μA (0 Ω to 50 kΩ)
2 µA (50 kΩ to 500 kΩ)
Thermocouple Equivalent Temperature Accuracies (readout only)
Type B±0.85°C from 600°C to 800°C
±0.68°C from 800°C to 1000°C
±0.57°C from 1000°C to 1800°C
Type C±0.32°C from 100ºC to 550°C
±0.71°C from 550°C to 2300°C
Type E±0.52°C from -200°C to 0°C
±0.22°C from 0°C to 950°C
Type J±0.52°C from -200°C to 0°C
±0.23°C from 0°C to 1200°C
Type K±0.61°C from -200°C to 0°C
±0.24°C from 0°C to 1370°C
Type L±0.36°C from -200°C to 0°C
±0.23°C from 0°C to 1370°C
Type M±0.26°C from -20°C to 0°C
±0.25°C from 0°C to 400°C
±0.22°C from 400°C to 1400°C
Type N±0.72°C from -200°C to 0°C
±0.28°C from 0°C to 1300°C
Type R±1.09°C from -20°C to 0°C
±0.97°C from 0°C to 500°C
±0.49°C from 500°C to 1750°C
Type S±1.05°C from -20°C to 0°C
±0.95°C from 0°C to 500°C
±0.56°C from 500°C to 1750°C
Type T±0.60°C from -200°C to 0°C
±0.25°C from 0°C to 400°C
Type U±0.54°C from -200°C to 0°C
±0.24°C from 0°C to 400°C
Accuracies are based on internal Reference Junction Compensation. Refer to Technical manual for equivalent accuracies with an external reference junction.
Accuracies with Selected Probes (±°C)
-200°C5616-12: 0.014
5615-6: 0.025
5627A-12: 0.027
5610-9: n/a
0°C5616-12: 0.021
5615-6: 0.021
5627A-12: 0.049
5610-9: 0.009
100°C5616-12: 0.027
5615-6: 0.028
5627A-12: 0.065
5610-9: 0.009
300°C5616-12: 0.040
5615-6: 0.043
5627A-12: 0.103
5610-9: n/a
420°C5616-12: 0.050
5615-6: n/a
5627A-12: 0.130
5610-9: n/a
Includes readout accuracy, probe calibration, and probe drift
PRT Equivalent Temperature Accuracy (readout only)
–100°C±0.011
0°C±0.015
100°C±0.019
200°C±0.023
400°C±0.031
600°C±0.039
Thermistor Equivalent Temperature Accuracy (readout only)
0°C±0.002
25°C±0.003
50°C±0.006
75°C±0.014
100°C±0.030

 

Fluke Calibration 5628 Reference Grade Secondary PRT Probe Specifications

TemperatureRange 

–200 °C to 661 °C 

Handle Temp. 

0 °C to 80 °C 

RTPW 

5626: 100 Ω (± 1 Ω)
5628: 25.5 Ω (± 0.5 Ω) 

W(Ga) 

³ 1.11807 

Calibrated Accuracy† (k=2) 

± 0.006 °C at –200 °C
± 0.006 °C at 0 °C
± 0.015 °C at 420 °C
± 0.022 °C at 661 °C 

Stability 

5626: ± 0.003 °C
5628: ± 0.002 °C 

Long-Term Drift (k=2) 

5626:  < 0.006 °C/100 hours at 661 °C
5628: < 0.004 °C/100 hours at 661 °C 

Immersion 

At least 12.7 cm (5 in) recommended 

Sheath 

Inconel™ 600 

Lead Wires 

4-wire Super-Flex PVC, 22 AGW 

Termination 

Gold-plated spade lugs, or specify 

Size 

6.35 mm dia. x 305 mm,  381 mm,  or 508 mm (0.25 x 12, 15, or 20 in) standard, custom lengths available 

Calibration 

Accredited calibrations from Fluke Calibration

†Includes calibration and 100 hr drift

Write Your Own Review