We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
BK Precision’s 9801 has the capacity of a programmable AC power source and a measurement tool all in one piece of equipment. The 9801 provides a maximum of 300 VA through the universal line output terminals on the front and the output connector on the rear. With a programming resolution of 0.1 V, the output can be varied from 0 to 300 V. The 9801 has a range of 45 Hz to 500 HZ in which the output frequency can be adjusted, and a start and stop phase angle from 0 to 360 degrees. Irms, Vrms, frequency, Ipeak, apparent power, power factor (PF), true power and elapsed output time can all be viewed on the expansive VFD display.
BK Precision has made the 9801 a solution for a wide variety of application requirements with functions and modes such as, list mode, power line disturbance (PLD) simulator, and the sweep mode for replication of conventional power grid disturbances and faults. A built-in dimmer mode is even available for testing LEDs and motors.
Model | 9801 | 9803 | 9805 | |
AC Input | ||||
Phase | Single | |||
Voltage | 110 / 220 VAC ± 10% | |||
Frequency | 47 - 63 Hz | |||
Max. Current | 8 A max. | 15 A max. | 30 A max. | |
Power Factor | 0.5 (typical) | 0.7 (typical) | 0.7 (typical) | |
AC Output | ||||
Max. Power | 300 VA | 750 VA | 1500 VA | |
Max. Current (rms) | 0 - 150 V | 3.0 A | 6 A | 12 A |
0 - 300 V | 1.5 A | 3 A | 6 A | |
Max. Current (peak) | 0 - 150 V | 12 A | 24 A | 48 A |
0 - 300 V | 6 A | 12 A | 24 A | |
Crest Factor | ≥4 | |||
Phase | Single | |||
Total Harmonic Distortion (THD) |
≤0.5% at 45 - 500 Hz (Resistive load) | |||
Line Regulation | 0.1% max for a ±10% line change | |||
Load Regulation | ≤0.5% FS (Resistive load) | |||
Response Time | <100 µs | |||
Programming | ||||
Voltage (rms) | Range | 0 -300 V, 150 V / 300 V (Auto) | ||
Resolution | 0.1 V | |||
Accuracy | ±(0.2% + 0.6 V) | |||
Frequency | Range | 45 - 500 Hz | ||
Resolution | 0.1 Hz at 45 - 99.9 Hz 1 Hz at 100 - 500 Hz | |||
Accuracy | ±0.1 Hz (100 Hz) ±1 Hz (100 - 500 Hz) | |||
Phase Angle | Range | 0 - 360° | ||
Resolution | 0.1° | |||
Accuracy | ±1° (45 - 65 Hz) |
Measurements | ||||
Voltage (rms) | Range | 0 - 300 V | ||
Resolution | 0.1 V | |||
Accuracy | ±(0.2% + 0.6 V) | |||
Current (rms) |
Range* | Low: 120.0 mA / Mid: 1.200 A / High: 3.00 A | Low: 120.0 mA / Mid: 1.200 A / High: 6.00 A | Low: 120.0 mA / Mid: 1.200 A / High: 12.00 A |
Resolution | Low: 0.1 mA / Mid:1 mA / High: 10 mA | |||
Accuracy | Low: ±(0.2% + 0.4 mA) / Mid: ±(0.2% + 4 mA) / High: ±(0.2% + 20 mA) | |||
Current (peak) | Range | 0 - 12 A | 0 - 24 A | 0 - 48 A |
Resolution | 0.01 A | |||
Accuracy | ±(1% + 120 mA) | |||
True Power (watts) | Resolution | Low: 0.01 W / Mid:0.1 W / High: 1 W | ||
Accuracy (47 - 65 Hz) | Low: ±(0.2% + 0.05 W) / Mid: ±(0.2% + 0.5 W) / High: ±(0.2% + 2 W) | |||
Frequency | Range | 45 - 500 Hz | ||
Resolution | ±0.1 Hz (45 - 99.9 Hz), ±1 Hz (100 - 500 Hz) | |||
Accuracy | ±0.1 Hz | |||
Power Factor | Range | 0.000 - 1.000 | ||
Resolution | 0.001 | |||
Accuracy | True Power (w) / Apparent Power (VA) | |||
Apparent Power (VA) | Resolution | Low: 0.01 VA / Mid:0.1 VA / High: 1 VA | ||
Accuracy | Voltage (rms) x Current (rms) | |||
Temperature Coefficient (typical) |
±0.04% per ºC | |||
General | ||||
Memory | 10 Locations | |||
External BNC I/O | Trigger input, sync output, output status, output indicator / control | |||
Interface | LAN, USB, RS232 | LAN, USB, RS232, & GPIB | ||
Operating Temperature | 32 °F to 104 °F (0 °C to 40 °C) 20 - 80% R.H. | |||
Storage Temperature | -4 °F to 158 °F (-20 °C to 70 °C) ≤ 85% R.H. | |||
Environmental conditions | For indoor use only, max humidity 80%, no condensation | |||
Dimensions (W x H x D) | 8.45" x 3.47" x 17.83" (214.5 x 88.2 x 453.5 mm) |
17.3” x 5.2” x 21.1” (439 x 131.4 x 535.7 mm) | ||
Weight | 20.94 lb (9.5 kg) | 88.2 lb (40 kg) | 115 lb (52.16 kg) | |
Two-Year Warranty | ||||
Standard Accessories | AC Power cord (9801 only), unterminated power cord with input connector (9803 & 9805 only), rackmount ears & handles (9803 & 9805 only), instruction manual, test report & certificate of calibration | |||
Optional Accessories | IT-E151 rack mount kit (9801 only) |
Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes
over an ambient temperature range of 23 °C ± 5 °C.
* The current range switches from low to mid range or mid to high range when Ipeak > 300
% of the present range.
When Ipeak is <80 % of the high range, the current range switches from high to mid range.
When Ipeak is <20 % of the mid range, the current range switches from mid to low range.