The Druck DPI 705E-IS is the Intrinsically Safe version of the DPI 705E. The DPI 705E-IS is for use in hazardous environments.
Accuracy Specification
Absolute Sensors | Standard Accuracy | High Accuracy | ||
Pressure Range | NLH&R | Total Uncertainty | NLH&R | Total uncertainty |
-10 to 50°C (14 to 122°F) | ||||
(% FS) | (% FS) | (% FS) | (% FS) | |
350 mbar to 1400 bar | 0.08 | 0.1 | 0.04 | 0.075 |
Gauge/Differential Sensors | Standard Accuracy | High Accuracy | ||
Pressure Range | NLH&R | Total Uncertainty | NLH&R | Total Uncertainty |
-10 to 50°C (14 to 122°F) | ||||
(% FS) | (% FS) | (% FS) | (% FS) | |
25 mbar | 0.3 | 0.348 | N/A | N/A |
70 mbar | 0.1 | 0.121 | N/A | N/A |
200 mbar | 0.08 | 0.1 | N/A | N/A |
350 mbar to 700 mbar | 0.08 | 0.1 | 0.04 | 0.05 |
1 bar to 200 bar | 0.08 | 0.1 | 0.04 | 0.05 |
Note: NLH&R Non-linearity, hysteresis and repeatability.
General Specifications | |
Working Pressure | 110% FS (alarm active beyond this range) |
Sealing | IP54 |
Operating temperature | -10 to 50°C (14 to 122°F) |
Storage temperature | -20 to 70°C (-4 to 158°F) |
Humidity | 0 to 90% RH non condensing |
Shock and vibration | MIL-PRF-28800F for Class II equipment. 1 m Drop Tested @ -20°C (-4°F) |
EMC | BS EN 61326-1 |
Electrical safety | BS EN 61010-1 UL 61010-1 |
Pressure safety | Pressure equipment directive class SEP. UL61010 Safe to 2xFS |
Approval | CE marked, RCM |
Size and weight | L 200 mm, W 95 mm, D 43 mm 563 g maximum |
RoHS | Compliant |
Media Compatibility | Media to be compatible with Stainless Steel and Hastelloy |