Crystal Engineering XP2i Digital Pressure Gauge

Generating PDF - Please be Patient

Part Number: XP2i

Manufacturer: Crystal Engineering

Configuration:

XP2i-

Key Features:

  • Now available to 15000 psi
  • Available in three Accuracies (0.1% of Reading, 0.05% of F.S. (S5), or 0.02% of F.S. (S2))
  • Marine grade enclosure resists corrosion
  • Accuracy not affected from -10 to 50°C
  • Replaces 5 or more digital test gauges 
  • Fast pressure safety valve (PSV) mode
  • Define custom pressure units
  • Shift everyday wear to replaceable CPF adapter 
  • Easy to use-No Menu. No Manual? No Problem!

With the optional DataLoggerXP upgrade:

  • Record 32 000 data points
  • Battery-optimizing, Ultra Low Power (ULP) Mode
  • Over an entire year of continuous recording time
  • Now available on the Dual-Line Display XP2i

Availability: In stock

Product Options:
Crystal Engineering Model XP2i Pressure Range Selection
Crystal Engineering Model XP2i Dual Display
Crystal Engineering Model XP2i Pressure Accuracy Selection
Crystal Engineering Model XP2i Adapter Type (for <15K PSI Ranges Only)
Crystal Engineering Model XP2i Connection Location
Crystal Engineering Model XP2i Data-logging
Crystal Engineering Model XP2i Pump System (Optional)
Crystal Engineering Model XP2i Without Liquid (Pump drained)(Optional)
Crystal Engineering Model XP2i Carrying Case ONLY When Ordering Pump System
Crystal Engineering Model XP2i Calibration Options
Crystal Engineering Model XP2i Accessory Options

Regular Price: $1,149.00

Special Price $1,104.19

Product Detail

Details

Crystal Engineering XP2i is an ultra rugged, intrinsically safe, digital test gauge with 0.1% of reading accuracy.

Specifications

Details

ACCURACY

0 to 20% of Range:

± (0.02% of Full Scale)

Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.

20% to 110% of Range:

± (0.1% of Reading)

All models indicate vacuum, but vacuum specification applies to 15 psi, 30 psi, 100psi, and 300psi models only.

Vacuum:

± (0.25% of Full Scale*)

Not recommended for continuous use below -14.5 psi.

 

* Full Scale = -14.5 psig

Refer to XP2i-DP data sheet for gauges that are intended for continuous high vacuum use.

 

Vacuum to Full Scale

S2 Option: ± (0.02% of Full Scale) 15 to 25° C
± (0.05% of Full Scale) -10 to 15° C,
25 to 50° C, typical
S5 Option: ± (0.05% of Full Scale) -10 to 50° C

 

OPERATING TEMPERATURE

Temperature Range:

-10 to 50°C (14 to 122°F)

Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification.

 

DISPLAY

Screen:

5.5 digits

LCD readable in sunlight with bright backlight.

Single Line Display:

16.9 mm (0.67") high

PSV mode is intended for Pressure Safety Valve testing, and requires software to enable.

Dual Line Display:

Main Display: 13.9 mm (0.55")
Secondary Display: 5.3 mm (0.21")

Display Rate:

4 readings/second (standard)
8 readings/second (PSV mode)

 

POWER

The XP2i is Intrinsically Safe only if powered by one of the following battery types.

ATEX/IECEx: Approved Battery Type Ta= Marking
Rayovac Max Plus 815 -20 to 50° C Ex ia IIC T4 Ga
Energizer E91
Energizer EN91 -20 to 50° C Ex ia IIC T3 Ga
Duracell MN1500

 

CSA: Approved Battery Type Ta= Marking
Rayovac Max Plus 815 -20 to 45° C Class I, Division 1, A, B, C, D T4
Energizer E91 -20 to 50° C Class I, Division 1, A, B, C, D T3C
Energizer EN91
Rayovac Max Plus 815

 

 

3 x AA:

1500 hours, typical

Uses 3 alkaline AA (LR6) batteries. Use of backlight and/or RS232 reduces operating time.

 

 

Optional AC power supply available. Do not use AC power supply in hazardous locations.

Ultra Low Power:

> 1 year, typical**

** 20 minute recording interval or greater in LT5 mode.

 

DATA/COMMUNICATION

Digital Interface:

RS232, DB-9F

Free Labview drivers and programming instructions are available for download from website.

 

 

Do not use RS32 in hazardous locations.

 

DATALOGGING

Capacity:

32 000 data points

DataLoggerXP Firmware Update may be purchased at time of order, or in an Upgrade Package later.

Storage Type:

Non-volatile flash memory

Software available for download from our website.

Minimum Interval:

1 second

Compatible w/ 32 & 64 bit Windows 7, and Vista.

Maximum Interval:

18 hours

Produces csv files, or used Excel template files (samples included) to automatically format and graph data.

Min. Event Duration:

250 mS***

*** Logging Type - Average with Peaks.

 

„„ENCLOSURE

Weight:

562 g (19.8 oz)

Weight includes batteries.

Rating:

IP66 and IP67

Submersible to 1 m for 30 minutes [IEC 60529].

Housing:

Diecast aluminum

Nickel plating over low copper, marine grade aluminum.

Keypad and Labels:

UV Resistant Polyester

LCD protected from impact damage by 1.5 mm (0.06") thick polycarbonate lens under polyester window.

 

 

Skydrol compatible.

 

SENSOR

Wetted Materials:

316 stainless steel

All welded, with a permanently filled diaphragm seal.

 

and Viton internal o-ring (finger tight)

Metal to metal cone seal; O-ring can be removed if necessary.

Diaphragm Seal Fluid:

Silicone Oil

Optional connection on back.

Connection:

Crystal CPF Female

1/4" male NPT adapter included unless BSP or M20 is specified.

 

 

1/4" medium pressure tube system compatible with HiP LM4 and LF4 Series, Autoclave Engr SF250CX Male and Female Series.

 

STORAGE TEMPERATURE

Temperature Range:

-40 to 75°C (-40 to 167°F)

Batteries should be removed if stored for more than one month.

 

SPECIAL FEATURES

The following requires the use of our free ConfigXP software.

User Defined Unit:

Allows the user to define and display any pressure units not included, or to use the gauge to display force, level, or other pressure related parameters. Reduce or increase displayed resolution.

PSV Mode:

Intended for pressure safety valve / burst disk testing. 8 readings per second. Captures point at which valve opens, as well as reseat pressure.

Averaging:

Smooths unsteady readings (1 to 10 samples).

Remove:

Unwanted pressure units from Units button.

Disable:

Peak pressure button.

Limit or Disable:

Zero button.

Password Protect:

Changes to configuration or calibration factor(s).

Save:

Configuration to a file.

Copy:

Configuration to other gauges.

 

 


Optional Pressure Pump Systems


A includes System A (pneumatic) 0 to 30 psi

AH includes System A (pneumatic) 0 to 580 psi

includes System B (pneumatic) -25 inHg to 30 psi

BH includes System B (pneumatic) -25 inHg to 580 psi

C includes System C (oil) 0 to 3,000 psi

CH includes System C (oil) 0 to 5,000 psi

DO includes System D (oil) 0 to 5,000 psi

DW includes System D (water) 0 to 5,000 psi

EO includes System E (oil) 0 to 10,000 psi

FOV includes System F (oil) 0 to 15,000 psi

FWV includes System F (water) 0 to 15,000 psi

GO includes System G (oil) 0 to 15,000 psi

GW includes System G (water) 0 to 15,000 psi

HO includes System H (oil/pneumatic) -25 inHg to 580 psi and 0 to 5,000 psi

Data Sheet Resources

Reviews

Write Your Own Review

You're reviewing: Crystal Engineering XP2i Digital Pressure Gauge

How do you rate this product? *

  1 star 2 stars 3 stars 4 stars 5 stars
Price

Applications

applications

*Not recommended for oxygen service

  • Field gauge calibration
  • Permanent outdoor installation
  • Burst disc testing
  • Pressure safety valve (PSV) testing
  • Hydraulic torque wrench indication
  • Torque datalogging
  • Pressure regulator testing
  • Pipeline static pressure testing
  • Chart recorder replacement
  • Wellhead monitoring
  • Filter performance testing
  • Hydrostatic leak testing
  • Accessories

    Accessories

    No accessories available

    Videos

    Products In This Series

    As low as

    Now $2,441.00