We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
A small housing and multiple mounting options allow the CXLdp to be easily mounted on a DIN rail, wall or in-plenum. With the convenient "Euro" style electrical connector, the CXLdp can be installed into a panel without power interruption. The CXLdp is available in ranges from 0/0.1 through 0/25 In H2O and in accuracies of +/-0.8% and +/-0.4% F.S.. Other standard features include high over-pressure capability, LED status light and a CE Heavy Industrial EMI/RFI immunity rating. With an attractive price, the CXLdp is the ideal pressure transmitter for most building control system applications.
PERFORMANCE SPECIFICATIONS
Ref. Temperature: 70°F ±2°F (21°C ±1°C)
Accuracy Class (Span): ±0.8% or ±0.4%
Includes non-linearity, hysteresis, non-repeatability,
zero offset and span setting errors.
Stability – Max. Change
(Span/year): ≤0.25% ≤0.25%
Standard Ranges (Inches W.C.)
Unidirectional Ranges:
Differential
0/0.10 0/0.75 0/2.50 0/10.00
0/0.25 0/1.00 0/3.00 0/15.00
0/0.50 0/2.00 0/5.00 0/25.00
Bidirectional Ranges:
Compound
±0.10 ±0.50 ±2.00 ±10.00
±0.25 ±1.00 ±5.00 ±15.00
Response Time: 250 msec
ENVIRONMENTAL SPECIFICATIONS
Temperature Limits:
Storage –40 to 180°F
Operating +0 to 160°F
Compensated Range +35 to 130°F
(10-95% R.H. non-condensing)
Temperature Coefficients:
Zero & Span ±0.03% Span/°F
CE Marked: Per DoC
EMC Directive 2004/108/EC
IEC/EN 61326-1:Edition 1.0 Industrial
IEC/EN 61326-2-3:Edition 1.0 Annex BB Industrial
FUNCTIONAL SPECIFICATIONS
Overpressure Limits:
Proof Pressure 15 psid
Burst Pressure 25 psid
Max. static line pressure 25 psi
Mounting Position Effect: ±1%/g (lowest range)
Note: Calibration in vertical position is standard
ELECTRICAL SPECIFICATIONS
Voltage Output: Supply Voltage: Supply Current:
4-20mA (2 wire) 12-36 Vdc 21.5mA
0-5 Vdc (3 wire) 11.5-36 Vdc 4.5mA
0-10 Vdc (3 wire) 14-36 Vdc or 24 Vac 6mA
Output signal is independent of power supply changes
Reverse Wiring Protected
Zero and Span Adjustment:
Externally accessible
Zero: ±5% F.S.
Span: ±5% F.S.
PHYSICAL SPECIFICATIONS
Pressure Connections:
1⁄4˝ brass barbed fittings
1⁄8 NPT Female brass
Electrical Connection: Euro style pluggable terminal
block accepts 12-26 gauge wire
Enclosure: NEMA Type 1 Fire-retardant ABS
(meets UL 94-5VA)
LED visual indicator standard
Weight: Approx. 2.5 oz
Media: Clean, dry and non-corrosive gas
Mounting: Threaded fastener and 35mm DIN rail
mount standard.
Option: 1⁄2˝ plenum/conduit mounting bracket and
cover kit (order part #101A213-01)
XRH: (9 point NIST Calibration Certification)
- Static or Velocity pressure measurement for four stations
- Ducts
- Building Pressure
- Filter Efficiency
- VAV boxes or room pressurization