Temperature range
@ ambient temp. 0°C/32°F:.........10 to 700°C/32 to 1292°F
@ ambient temp. 23°C/73°F:.......33 to 700°C/91 to 1292°F
@ ambient temp. 40°C/104°F:...50 to 700°C/122 to 1292°F
Accuracy (model B & C) with external STS ref. sensor
RTC-700 B & C (33 to 660°C)...................... ±0.11°C/±0.2°F
Relative to reference standard. Specifications by use of the
external JOFRA STS-200 reference sensor
Accuracy with internal reference sensor
RTC-700 A, B & C.....................................*±0.29°C/±0.52°F
* Accuracy 33 to 660ºC is including calibration uncertainty.
Accuracy 660 to 700ºC ± 0.29ºC exclusive calibration uncertainty.
RTC-700 is delivered calibrated in 660 to 700ºC with calibration
uncertainty ±1.4ºC
Stability
@ 33 to 125ºC/91 to 257ºF.................... ±0.008°C/±0.015°F
@ 125 to 425ºC/257 to 797ºF................ ±0.015ºC/±0.027ºF
@ 425 to 700ºC/797 to 1292ºF................ ±0.02ºC/±0.036ºF
Measured after the stability indicator has been on for 15 minutes.
Measuring time is 30 minutes.
Axial homogeneity @ 60 mm
@ temp. range 33 to 100°C..................................... ±0.10°C
@ temp. range 100 to 420°C................................... ±0.25°C
@ temp. range 420 to 700°C................................... ±0.40°C
@ temp. range 91 to 212°F.......................................±0.18°F
@ temp. range 212 to 788°F.....................................±0.45°F
@ temp. range 788 to 1292°F...................................±0.72°F
Resolution (user-selectable)
All temperatures ............................ 1°, 0.1°, 0.01° or 0.001˚
Heating time
33 to 700°C/91 to 1292°F................................... 45 minutes
Cooling time
700 to 100°C/1292 to 212°F............................... 43 minutes
100 to 50°C/212 to 122°F................................... 17 minutes
50 to 33°C/122 to 91°F....................................... 20 minutes
Time to stability (approx.)
RTC-700.............................................................. 10 minutes
Immersion depth
RTC-700........................................................ 200 mm/7.9 in
PHYSICAL SPECIFICATIONS
Weight
Instrument..................................................... 11.3 kg/24.9 lb
Carrying case................................................ 11.0 kg/24.3 lb
Instrument in Carrying case.......................... 24.3 kg/53.6 lb
Instrument in packaging................................ 16.3 kg/36.0 lb
Instrument in Carrying case & packaging..... 27.3 kg/60.2 lb
Dimensions (LxWxH)
Instrument................................................ 362x171x421 mm
Carrying case........................................... 470x350x600 mm
Instr. & Carrying case............................... 470x350x600 mm
Instr. & packaging.................................... 442x251x481 mm
Instr., Carrying case & packaging............ 550x430x660 mm