Technical Specification - Altek 134-2
Accuracy: ±(0.08% of Full Scale + 1 LSD)
Warm up time: 10 seconds to specified accuracy, 2 minutes to maximum accuracy
Temperature effect: ±0.015%/°C based on 23°
Batteries: Two “9V”, (1604) batteries (Alkaline supplied and recommended)
Battery life:
Source and power transmiiter: Nominal 22 hours at 12
mA, 16 hours at 20 mA with 250 ohm load
Read & 2-wire simulator: Nominal 40 hours
Low battery indication: “BAT” indication on the display at 4 hours left
Overvoltage protection: Protected to 120V AC/DC for 30 seconds
Operating temperature range: -5 to +130 °F (-20 to +55°C)
Storage temperature range: -13 to +130°F (-25 to +55°C)
Relative humidity: 10 to 90%, non-condensing for 24 hours from 0 to 35°C
Overall size: 121.9 x 66.0 x 35.8 mm (4.8 x 2.6 x 1.41 inches)
Weight: 0.22 kg (7.8 oz)
Milliamp source
Ranges: 0.00 to 24.00 mA; -25.0 to 125.0 % of 4 to 20 mA
Accuracy: ±(0.08% of 24 mA span + 0.01 mA) = 0.03 mA
Typical drive capability: 700 ohms @ 20.00 mA with fresh batteries, 300 ohms @ 20.00 mA with low batteries
Compliance: nominal 14 VDC @ 20 mA with fresh batteries
Power & measure 2-wire transmitters
Ranges and accuracy: Same as for Milliamp Source
Output current: up to 24.00 mA
Typical drive capability: 700 ohms @ 20.00 mA with fresh batteries, 300 ohms @ 20.00 mA with low batteries
Compliance: nominal 14 VDC @ 20 mA with fresh batteries
2-wire transmitter simulator
Ranges and accuracy: Same as for Milliamp Source
Loop voltage limits: Minimum, 3 VDC; maximum 45 VDC
Overload protection: Current limited to 25 mA nominal
Milliamp read
Ranges and accuracy: Same as for Milliamp Source
Voltage burden: 0.9V at 4 mA, 1.2V at 20 mA, 1.9V at 24 mA excitation current