The SR701 is a general purpose 1000AAC current probe designed for use with DMMs, loggers and other instruments that have a current input. It has a 1000:1 ratio and output of 1mA per amp measured in the jaw. The SR701 has jack output terminations that accept 4mm banana plugs. The rounded jaw design on this probe makes it well-suited for power applications where high accuracy and low phase shift and wide frequency response are important.
MODEL | SR701* |
ELECTRICAL | |
Nominal Range | 1000A |
Measurement Range | 1mA to 1200A |
Transformation Ratio | 1000:1 |
Output Signal | 1mAac/Aac (1Aac @ 1000A) |
Accuracy (1000A Range) 1mA to 100mA |
≤3% of Reading ± 5mA |
0.1 to 1A | ≤2% of Reading ± 3mA |
1 to 10A | <1% of Reading |
10 to 100A | ≤0.5% of Reading |
100 to 1200A | ≤0.3% of Reading |
Phase Shift (1000A Range) 1 to 10A |
≤2° |
10 to 100A | ≤1° |
100 to 1200A | ≤0.7° |
Overload | 1200A for 40 min ON, 20 min OFF |
Frequency Range | 30Hz to 5kHz; current derating above 1kHz using the formula: 1000A x 1F (in kHz) |
Load Impedance | 5Ω max |
Working/Common Mode Voltage | 600Vrms CAT III |
Output Termination | Jack |
MECHANICAL | |
Operating Temperature | -14° to 122°F (-10° to 50°C) |
Storage Temperature | -4° to 158°F (-20° to 70°C) |
Operating Relative Humidity | 0 to 85% @ 35°C |
Jaw Opening | 2.25” (57mm) max |
Maximum Conductor Size | 2.05” (52mm) |
Maximum Bus Bar Size | One 1.95 x 0.19” (50 x 5mm) |
Dimensions | 4.37 x 8.50 x 1.77” (111 x 216 x 45mm) |
Weight | 1.21 lbs (550g) |
Polycarbonate Material | Handles: Polycarbonate + ABS, Gray, UL94 V0. Jaws: Polycarbonate, Red, UL94 V0 |
SAFETY | |
Electrical | EN 61010-2-032 |
UL Approval | Yes – Canada and United States |
Double Insulation | Yes |
CE Mark | Yes |
Note: Reference conditions: 23 ± 3°K, 20 to 75% RH, 48 to 65Hz, external magnetic field <40A/m, no DC component, no external current carrying conductor, test sample centered. Load impedance 5Ω.