Additel ADT 761A-500 Automated Pressure Calibrator

SKU
ADT761A-500
Manufacturer:
Additel
Additel

Key Features:

  • Automated and self-contained pressure generation and control to 500 psig (35 bar.g)
  • 0.02%FS accuracy
  • Two removable internal pressure modules for multi-range selection
  • Control stability to 0.005%FS
  • Portable, designed for use in the field and in the lab
  • Ability to measure two external pressure modules
  • Wi-Fi, Bluetooth, USB and Ethernet communication
  • Full HART field communicator
  • HART and profibus communication
  • Data logging and task management
  • Patented electric pump technology and improved speed
Product Options:
$22,481.00

Secured and trusted checkout with

Part of the Additel ADT761A series, the ADT 761A-500 generates and controls from vacuum pressures up to 500 psig (35 bar.g). With the built-in barometer both gauge and absolute pressures can be realized. The ADT 761A-500 comes with a CP500 module (-12.5 to 500 psig) for the high range and the low range can be preconfigured based on the variety of modules available down to 10 psig (0.7 bar.g).

 

The ADT161 pressure module is sold separately, Click the Accessories Tab to see the pressure modules

Pressure Specifications
Specification761A-LLP761A-D761A-500761A-1K761A-APXR761A-BP
Max Pressure Range30 inH2O (75 mbar)35 psi (2.5 bar)500 psig (35 bar.g)1,000 pisg (70 bar.g)Dependent on APXR1,200 hPasensor
Min Pressure Range-30 inH2O (-75 mbar)-13.5 psi (-0.95 bar)-13 psi (-0.9 bar)-13 psi (-0.9 bar)Dependent on APXR100 hPasensor
Accuracy(1)0.05%FS0.02%FS0.02%FS0.02%FS0.01%FS0.01%FS
Stability0.005%FS or 0.05 Pa0.005%FS0.005%FS0.005%FS0.003%FS0.02 hPa
Pressure TypeDifferential, GaugeDifferential, GaugeGauge, AbsoluteGauge, AbsoluteGauge, AbsoluteAbsolute
Over Range Indication1.2
Resolution6 digits
Measurement UnitsPa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, inH2O@20°C, inH2O@68°F, kg/m2, mtorr, torr, lb/ft2, tsi, custom
Barometric AccuracyN/AN/A60 Pa60 Pa60 PaN/A
[1] One year accuracy (including 1 year stability). FS specification applies to the span of the module range.
[2] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT761A-500 and ADT761A-1K
* Additel 761A calibrators support 160A intelligent digital pressure modules that are available for gauge, vacuum and absolute pressure from -15 psi to 60,000 psi (-1 bar to 4200 bar).For detailed specs, see the 160A pressure module data sheet.

 

Electrical Specifications
SpecificationRangeResolutionAccuracyNote
mA MeasureAuto-ranging, ±25 or ±50 mA0.1 μA0.008%RD+0.004%FSImpedance<10Ω
V Measure±300 mV1 μV0.008%RD+0.002%FSImpedance >1GΩ
Auto-ranging, ±5, ±12 or ±30 V±5V:20 μV0.008%RD+0.002%FSImpedance >1MΩ
±12V:100 μV
±30V:100 μV
Loop Power Source24 VN/A±0.24 V50 mA (Max Loading)
mA Source0 to 2.5 mA or 2.5 to 25 mA0-2.5 mA: 0.05μA0-2.5 mA: 0.008%RD+0.1 μA20 mA @ 1KΩ
0-25 mA:0.5μA0-25 mA: 0.008%RD+1.0 μA
Power Source16 to 30 V1 V±0.5 V(24 V: ±0.24 V)70 mA (Max Loading)
V Source0 to 16 V250 μV0.008%RD+0.002%FS 
Pressure SwitchMechanical Switch, Live Mechanical Switch, NPN Switch, PNP SwitchN/AN/AResponse time<10 ms. If the switch is live, voltage range will be (3-30) V
Temperature Compensation5°C to 35°C
Temperature CoefficientOutside of 5°C to 35°C: <±0.0005%RD+0.0001%FS/°C
Misuse ProtectionUp to 30 V on any two sockets

 

Internal Module Specification and Compatibility
ModelinH2OmbarMediaAccuracy (%FS)(1)Burst Pressure761A-LLP761A-D761A-500761A-1K
ADT155-20-DP025±0.25±0.62G0.2 (2)100x   
ADT155-10-DP050±0.5±1.25G0.1(3)100x   
ADT155-05-DP1±1±2.5G0.05(4)100x   
ADT155-05-DP2±2±5G0.05(4)100x   
ADT155-05-DP5±5±10G0.05(4)50x   
ADT155-05-DP10±10±25G0.05(4)20x  
ADT155-05-DP20±20±50G0.0520x  
ADT155-05-DP30±30±75G0.0520x  
ADT155-05-DP50±50±125G0.053x   
ADT155-02-DP100±100±250G0.023x   
ADT155-02-DP150±150±350G0.023x   
ADT155-02-DP300±300±700G0.023x   
ADT155-02-DP400-380 to 400-950 to 1KG0.023x   
ADT155-02-DP800-380 to 800-950 to 2KG0.023x   
ADT155-02-DP1K-380 to 1K-950 to 2.5KG0.023x   
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules.
[2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS.
[3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS.
[4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS.
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode.Applicable only for use with the ADT761A-500 and ADT761A-1K.

 

Internal Module Specification and Compatibility (Gauge Pressure)
Modelpsigbar.gMediaAccuracy (%FS)(1)Burst Pressure761A-LLP761A-D761A-500761A-1K
ADT155-02-CP10±10±0.7G0.02(5)3x  
ADT155-02-CP15-13.5 to 15-0.95 to 1G0.02(5)3x  
ADT155-02-CP30-13.5 to 30-0.95 to 2G0.02(5)3x 
ADT155-02-CP35-13.5 to 35-0.95 to 2.5G0.02(5)3x 
ADT155-02-CP50-13.5 to 50-0.90 to 3.5G0.02(5)3x  
ADT155-02-CP100-13.5 to 100-0.95 to 7G0.02(5)3x  
ADT155-02-CP150-13.5 to 150-0.95 to 10G0.02(5)3x  
ADT155-02-CP200-13.5 to 200-0.95 to 14G0.02(5)3x  
ADT155-02-CP300-13.5 to 300-0.95 to 20G0.02(5)3x  
ADT155-02-CP500-13.5 to 500-0.95 to 35G0.02(5)3x  
ADT155-02-CP600-13.5 to 600-0.95 to 40G0.02(5)3x   
ADT155-02-CP1K-13.5 to 1K-0.95 to 70G0.02(5)3x   
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules.
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode.Applicable only for use with the ADT761A-500 and ADT761A-1K.

 

General Specifications
SpecificationDescription
User InterfaceColor touch screen and/or keypad operation
 Four total: one electrical, high or low internal
Channelspressure module, two external pressure modules
Enclosure IP RatingIP31
BatteryRechargeable Li-Ion battery, typically 16 hours of operation, recharges in less than 5 hours.
PowerRechargeable Li-Ion battery, external power 110/220 V power adapter 27 V
Display7” TFT touch screen 800 x 480 color
CommunicationsUSB, LAN, Bluetooth, Wi-Fi
Weight<17.52 lb (7.95 KG)
Size11.77 x 7.60 x 7.56 in (299 x 193 x 192 mm)
CertificationISO17025 accredited certificate of calibration with NIST-traceable data
Data Storage6 GB
Data LoggingUp to 1,000,000 readings (data and time stamped)
HART CommunicatorRead, configure and calibrate HART devices - DD files updated periodically
Task documentationUp to 1000 tasks can be stored with data
Automation FunctionsSwitch test, auto step, leak test
User Interface LocalizationEnglish, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian
Pump life>100,000 cycles
Environmental SpecificationsOperating temperature: 32 °F to 122 °F( 0°C to 50°C)
Compensated temperature: 32 °F to 122 °F( 0°C to 50°C)
Storage temperature: -20°C to 60°C ( -4 °F to 120 °F)
Humidity: <90%, non-condensing
Vibration and ShockVibration: 4g (20 to 2,000 Hz)
Shock: 8g, 1 meter drop test
ComplianceCE
Software CompatibilityACal, Additel Land and Additel Link for access via mobile application
Warranty1 year

 

Pressure Range
TypeModelPressure RangeHigh-Range Pressure ModuleLow-Range Pressure Module
RangeAccuracyRangeAccuracy
Low/Differential PressureADT761A-LLP-75 to 75 mbar (-30 to 30 inH2O)DP30: -75 to 75 mbar (-30 to 30 inH2O)0.05%FSUser selectable from DP20 to DP025See Internal Module Table
ADT761A-D-0.95 to 2.5 bar (-13.5 to 35 psi)CP35: -0.95 to 2.5 bar (-13.5 to 35 psi)0.02%FSUser selectable from DP10 to CP30See Internal Module Table
Standard Accuracy Gauge/Absolute PressureADT761A-500-0.9 to 35 bar (-13 to 500 psi)CP500: -0.9 to 35 bar (-13 to 500 psi)0.02%FSUser selectable from CP10 to CP300See Internal Module Table
ADT761A-1K-0.9 to 70 bar (-13 to 1000 psi)CP1K: -0.9 to 70 bar (-13 to 1000 psi)0.02%FSUser selectable from CP30 to CP600See Internal Module Table
Precision Accuracy Gauge/Absolute PressureADT761A-30R-0.95 to 1 bar (-13.5 to 15 psi)AP30R: -0.95 to 1 bar (-13.5 to 15 psi)0.01%FSSingle sensor only
ADT761A-50R-0.95 to 2.5 bar (-13.5 to 35 psi)AP50R: -0.95 to 2.5 bar (-13.5 to 35 psi)0.01%FSSingle sensor only
ADT761A-100R-0.95 to 7 bar (-13.5 to 100 psi)AP100R: -0.95 to 7 bar (-13.5 to 100 psi)0.01%FSSingle sensor only
ADT761A-300R-0.9 to 21 bar (-13 to 300 psi)AP300R: -0.9 to 21 bar (-13 to 300 psi)0.01%FSSingle sensor only
ADT761A-500R-0.9 to 35 bar (-13 to 500 psi)AP500R: -0.9 to 35 bar (-13 to 500 psi)0.01%FSSingle sensor only
ADT761A-1KR-0.9 to 70 bar (-13 to 1000 psi)AP1KR: -0.9 to 70 bar (-13 to 1000 psi)0.01%FSSingle sensor only