Additel ADT 672 Digital Pressure Calibrator (0.1% FS Accuracy) - DISCONTINUED

SKU
ADT672-10
Manufacturer:
Additel
Additel

Key Features:

This Product is DISCONTINUED
Please see the Additel ADT673-10 as a Replacement

  • Pressure ranges to 40,000 psi (2,800 bar)
  • Pressure Measurement Uncertainty:
    ADT 672-02: 0.02% FS
    ADT 672-05: 0.05% FS
  • Measure mA with 0.02%RD + 0.9 μA accuracy
    Measure V with 0.02%RD + 0.9 mV accuracy
  • Power transmitters during test using 24V loop supply
  • Pressure Switch Test
  • Hart Communication capability
  • Advanced temperature compensation
  • Dual readout
  • Min/Max/Hold to capture changing measurements
  • Data Logging
  • Large, Easy to read display with 6-digit resolution
  • Back lighted display
  • Rechargeable battery or AC adapter
  • NIST traceable calibration with data (included)
Call For Price

Secured and trusted checkout with

An essential asset for gauges, switches and transmitters over a large pressure range, the ADT672-10 is part of Additel’s 672 series of precision pressure calibrators.

Configured to reach 0.1% F.S accuracy up to 40,000 psi, each silicon pressure sensor is aged, tested and screened before assembly to guarantee quality performance. The calibrator can read the current or mV generated by a transducer, while the built-in pressure sensor enables for the accurate measurement of pressure. Even more, an excitation voltage to power sensors or transmitters can be supplied during calibration.

Its progressive microprocessor technology, portable design and quality pressure sensors, all make the ADT672-10 one of Additel’s leading precision pressure calibrators.

Accuracy

ADT672-02: 0.02% of full scale
ADT672-05: 0.05% of full scale

>15,000 (1,000 bar): 0.1% FS

Gauge Types

Gauge Pressure
Compound Pressure
Absolute Pressure
Differential Pressure

Display

Description: Dual-line 6 full digit FSTN LCD with LED backlight
Display rate: 3.5 readings per second (Default setting)
Numerical display height: 16.5mm (0.65")

Pressure UnitsPa, kPa, MPa, psi, bar, mbar, inH2O 4°C, mmH2O 4°C, inHg 0°C, mmHg 0°C, kgf/cm2
EnvironmentalCompensated Temperature: 14°F to 122°F (-10°C to 50°C)
Operating Temperature*: 14°F to 122°F (-10°C to 50°C)
*0.025% FS accuracy guarenteed only over the ambient temperature range of 68°F to 79°F (20°C to 26°C)
Storage Temperature: -4°F to 158°F (-20°C to 70°C)
Humidity: <95%
Pressure Connections1/4NPT male, 1/4BSP male, M20X1.5 male  (≤15,000 psi)
1/4HP female or 1/4HP male (>15,000 psi)
*1/4HP female: Autoclave F-250-C, 9/16" - 18 UNF-2B
*1/4HP male: Autoclave M-250-C, 9/16" - 18 UNF-2A
0.236 inch (Ø6mm) test hose (for differential pressure)
Other connections available per request
Electrical Connection0.156 inch (Ø4mm) sockets
Over Pressure Warning 120%
 Data LoggerStorage capacity: 30 files, 40 records per file
Mode: manual and automatic
Hourly-record: record the data every hour
Interval-record: set by user
 PowerBattery: Rechargeable Li-ion polymer battery
Li-Battery working time: 40 hours
Recharge time: 4 hours
External power: 110V/220V power adapter (DC10V)
 EnclosureCase material: Aluminum alloy
Wetted parts:  316L SS
Dimension: Ø120mm X 46mm depth X 184mm height
Weight: 0.7kg
European Compliance: CE marked
 CommunicationRS232 (DB9/F, environmentally sealed)
Baud rate: 1200,2400,4800,9600
Data length: 8 bits
Stop bit: 2 bits
Address: from 1 to 112
Warranty1 year
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account
  • Gauge calibration
  • Precision pressure measurement
  • Pressure transmitters calibration
  • Pressure switch testing
  • Safety relief valve resting
  • Pressure regulator testing
  • Hydraulic leak testing